Giga-tronics Incorporated 4650 Norris Canyon Road San Ramon, California USA 94583 Phone: 925.328.4650 Fax: 925.328.4700 INSTRUMENT NOTEINSTRUMENT NOTE Calibration and Zeroing for High Power Sensors with Removable Attenuators When using a Giga-tronics high power sensor with external attenuator, Giga-tronics power meters automatically recognize the sensor type and compensates for the attenuation factor. However, when performing the front panel calibration, be sure to remove the attenuator before connecting to the calibrator port. During measurement, you do not need to enter an offset factor to account for the attenuator loss. The sensor frequency calibration factors automatically correct for the combined frequency response of the sensor and attenuator. Because the sensor and attenuator are a matched set, the serial numbers of the sensor and attenuator are identical. Do not use attenuators from other high power sensors. 1. Remove the high power attenuator from the sensor. 2. Connect the sensor to be calibrated from Channel A or B to the Calibrator Output (See Figure 1). 3. Press the CAL/ZERO hardkey. The meter will automatically zero and calibrate the sensor. 4. Reconnect the high power attenuator to the sensor (See Figure 2). NOTE: There are alignment marks (arrows) on the sensor and attenuator. To reduce measurement uncertainty, align the arrows when reconnecting the attenuator to the sensor. High Power Sensors Figure 1: Setup for Sensor Calibration Figure 2: High Power Sensor Connection AttenuatorSensor RadioFans.CN 收音机爱 好者资料库