1、1090.5255.541E-2 Instruments Division Service Kit for SML SML-Z2 1090.5203.02 Printed in the Federal Republic of Germany RadioFans.CN 收音机爱 好者资料库 1090.5255.542E-2 Contents 1Service program .8 1.1Overview.Fehler! Textmarke nicht definiert. 1.2Installation.4 1.3Program Description.4 1.3.1Selection of t
2、he Interface.5 1.3.2Main View.5 1.3.3Module Test.5 1.3.4Calibration .5 1.3.5Direct Mode.6 1.3.6Help.6 2External Level Calibration.7 2.1Overview.8 2.2Installation.9 2.3Preparations.10 2.4Program Description.10 2.4.1Entry of IEC/IEEE Bus Address .10 2.4.2Calibration .10 2.4.3Log File.11 2.4.3.1Header.
3、11 2.4.3.2Detector Linearity.11 2.4.3.3Level Correction Data.11 2.4.3.4IQ Temperature Correction Data.11 RadioFans.CN 收音机爱 好者资料库 SMLService Program 1090.5255.543E-2 1 Service program 1.1 Overview The SML Service Program is used to support troubleshooting and module testing. It enables operation of t
4、he internal control bits and reading out of the available diagnostic points. In addition, the program can be used to transfer basic remote-control commands. The program runs under Windows (95, 98, NT and 2000). The instrument can be remote-controlled via IEC bus or RS232 interface. External level ca
5、libra- tion is possible in conjunction with a power meter of the type Rohde this description refers to this status. To use the level correction program, you need a PC with Windows 95, 98, NT or 2000, an IEC/IEEE bus interface, the program smllevco_3.12 and Power Meter R&S NRVD, stock number 0857.800
6、8.02, with a Z51 sensor. No other power meters are currently supported and no alternatives are planned. External Level CalibrationSML 1090.5255.548E-2 2.2Installation To install the program under Windows (95, 98, NT or 2000), insert installation diskette 1 into the diskette drive and call the file s
7、etup.exe. Alternatively, you can start setup.exe from the directory that contains all files of the distribution kit. Follow the instructions of the installation routine. The service program requires the CVI runtime engine. This engine will be installed if not already present on the system. SMLExtern
8、al Level Calibration 1090.5255.549E-2 2.3Preparations Connect the IEC/IEEE buses of the PC, power meter and the R&S SML with one another. Connect sensor Z51 directly to the RF output of the R&S SML. Switch on the R&S SML, power meter and PC. Make sure that the ambient temperature is between +20 and
9、+25C. This temperature should not vary more than 2 to 3 K during the measurement time (which can be up to 25 minutes depending on whether the R&S SML01/02/03 or R&S SMV is used). Let the R&S SML warm up for at least 30 to 45 minutes before you start a calibration. Note: The service program easily al
10、lows you to check the temperature of all modules. Do not start calibration until these temperatures have stabilized (fluctuations must be 1 K). Warning: An external level correction will destroy the current correction data memory of the R&S SML. External Level CalibrationSML 1090.5255.5410E-2 2.4Pro
11、gram Description To start the program, click LEVEL CALIBRATION in the Start menu with the left mouse button. (The path within the start menu will depend on the specifications provided during installation.) Alterna- tively, you can call smllevco_3.12.exe from the program directory. 2.4.1Entry of IEC/