Philips_PM6507_Transistor_Curve-Tracer 电路图.pdf
《Philips_PM6507_Transistor_Curve-Tracer 电路图.pdf》由会员分享,可在线阅读,更多相关《Philips_PM6507_Transistor_Curve-Tracer 电路图.pdf(102页珍藏版)》请在收音机爱好者资料库上搜索。
1、HILIPS RadioFans.CN 收音机爱 好者资料库 ttC-, iy I . A A A - PHILIPS TRANSISTOR CURVE-TRACER PM 6507 9452 065 07. t 9499 520 04511 1/469/1/04 Manual RadioFans.CN 收音机爱 好者资料库 2 CONTENTS GENERAL I. INTRODUCTION 7 II. WORKING PRINCIPLE 7 III. BLOCK DIAGRAM 8 1. Control section 8 2. Display section i2 IV. TECHNIC
2、AL DATA 13 V. ACCESSORIES 15 DIRECTIONS FOR USE VI. INSTALLATION 17 A. Adjustment to the local mains voltage 17 B. Earthing 17 C. Connection to the mains 17 D. Rack mounting version 18 VII. OPERATING INSTRUCTIONS 19 A. Survey of the controls 19 B. Hints for testing semi-conductors 19 C. Functions of
3、 the controls 19 D. Operation 23 1. Basic setting 23 2. Switching-on the apparatus 23 3. Connection of asemi-conductor 23 4. Adjusting the display section 23 5. Setting for testing semi-conductors with 3 or 4 terminals 23 6. Setting for breakdown tests of diodes and other elements with 2 connections
4、 26 7. Setting for external control or programming 27 SERVICE DATA VIII. CIRCUIT DESCRIPTION 37 A. Supply section 37 ,_, 1. Mains supply 37 2. Collector voltage supply 37 3. H.T. supply 37 4. Direct voltage source +31.51 +300 x-30 V 37 B. Step generator 39 1. Phase shifter, amplifier and rectifier,
5、pulse shaper 39 2. Duty cycle 39 , 3. Ring counter 39 4. Amplifier 43 5. Power amplifier 44 C. Display section 44 1. Circuit for I, and V, l; measurements 45 2. Circuit for Vi;l.; measurements 46 3. Circuit for breakdown voltage and leakage current measurements 47 ,_ 4. Difference amplifier for X an
6、d Y-deflection 48 5. Output amplifiers for X and Y-deflection 48 6. Brightness control 48 IX. DISASSEMBLY 49 A. Cabinet 49 B. C.R.T. 49 RadioFans.CN 收音机爱 好者资料库 3 C. Removing the subunits D. Removing components X. MAINTENANCE XI. SURVEY OF ADJUSTING ELEMENTS XII. CHECKING AND ADJUSTING A. Direct volt
7、ages -x-30 V and 30 V B. EHT unit 4.5 kV C. 3 kV voltage for testing diodes D. Step generator E. Collector voltage supply F. Monitor XIII. THE SHIPMENT OF THE INSTRUMENT 52 57 61 62 63 63 64 64 64 64 65 67 XIV. LISTS OF PARTS 68 4 LIST OF FIGURES 1 2 3 4 5 Basic construction Block diagram Accessorie
8、s Rear side of the instrument Making the instrument suitable fora 19 rack 6 10 15 17 18 6 7 Basic considerations for testing semi-conductors 19 8 9 Front view with controls and sockets 20 10 Trigger point 20 11 Effect of the duty-cycle control 24 12 Curve display with current limiting 24 13 Breakdow
9、n characteristics 25 14 Compensation for the rion-linear part of the base-emitter diode characteristic 25 15 I-V aow m s r u d W Na a e m a _ _+_ _- E Fig. 2. Block diagram 12 whereby 3 voltage measuring ranges are available which can be reduced by a factor 10 with push- button x.1, as well as sever
10、al switchable ranges ._ for uA and nA measurements. Push-button Read connects the H.T. circuit only when it is closed manually or by means of a pro- tection device. 2. Voltage source for fourth electrode Connection socket S on the transistor holder, apart from earthing the transistor case (e.g. with
11、 ._, H.F. transistors), may also serve for connecting e.g. a second gate in case of field effect transis- tors. For this a variable direct voltage of max. 10 V can be applied (Shield Voltage), whose polarity can be selected by means of a switch. 2. Display section The display section of the curve tr
12、acer does not affect . the working of the other sections. After having passed the relevant circuits and after being amplified the vol- tage or current changes occuring in the transistor are displayed on the c.r.t. Two step-switches serve for range selection and mode selection for the horizontal and
13、vertical deflection. The horizontal deflection comprises 12 collector vol- tage ranges, 4 base voltage measuring ranges, one step range position, 3 ranges for diode test voltages and position EXT for external control. By means of the step switch for the vertical deflection the collector current rang
14、e is selected in 18 steps, s whereby the sensitivity of 10 steps can be increased by a factor 103 for measuring diode leakage currents. Moreover, the base voltage measuring sensitivity can be adjusted in 4 steps by means of this switch and an additional position is available for external control. By
15、 means of a slide switch all vertical deflection ranges can be extended to either side by a factor 0.5 and 5. To enable proper curve display in case of an inverted signal, a switch has been provided by means of which the polarity of the horizontal and vertical deflection voltages can be reversed. Th
16、is switch also serves for switching on the calibration voltage (CAL). The voltages present on the semi-conductor (or the current relations) are not influenced by the display section. Therefore the operating voltages for the semiconduc- tor under test are applied to the differential amplifiers via hi
17、gh-ohmic input stages. The output amplifiers supply the power required for the magnetic deflection, whereby the trace can be shifted in the horizontal and the vertical direction by means of the combined coarse and fine controls. The acceleration voltage for the picture tube is delive- red by a 4.5 k
18、V generator. In order to keep the brightness of the picture constant at different sweep speeds a brightness modulator has been provided which increases the intensity at the higher speeds. The brightness of the picture is also influenced by control Duty Cycle. 13 IV. Technical data Tolerances: Numeri
19、cal values with statement of tole- rances are guaranteed by the factory. Data without tolerances serve merely for information and represent the properties of an average apparatus. Step generator _, Step current generator: , Current steps (constant current values) Number of curves Stability of the st
20、ep spacings Stability of the current generator Step voltage generator: Voltage steps mV/step max. internal resistance 2: (current per step 200 mA) Switchable source resistors Duty cycle Zero adjustment of the step voltage 1, 2, 5A etc .200 mA/step (17 steps plus a 0 -step) All current steps can be d
21、oubled for a brief period 7 (8 with datum line) adjustable from 2 . . 8 Between the first and seventh step within an arbitrary range 1 %; with respect to each other: in all ranges 3 A load variation of 0.1 V at 1 V causes an error of 1 % in all ranges 0 20 50 100 200 500 1000 0.1 0.1 0.25 .5 1 2.5 5
22、 All voltage steps can be doubled for a brief period steps: 0, 1, 2, 5 4 etc . . 10 MQ (23 steps) at 100 pulses/sec. 250 sus . . . 6 ms (2.5%.60%and 100% at 200 pulses/sec. the same pulse times apply, however, twice the duty cycle in percentages. Overlap from 4 ms. Can be shifted 100 % (7 steps) in
23、the positive or negative direction (at 200 mA/step and all mV/step ranges max. 12 V at 30 %duty cycle). Collector supply Rectified sine-wave voltage, continuously adjustable in 4 ranges Voltage range Max. current Internal resistance Ri (at 220 V) 0 . . . 10 VP _ 20 A 0.3 2 0. . 50Vp_, 4 A 3.8 S2 0.1
24、00VP_u 2 A 16 S? 0.500 V, _ u 0.2 A 400 52 Polarity Overload protection Current limitation Connection sockets BU2B Diode voltage source Voltage range Maximum load current Short circuit current ., Internal resistance Voltage sourcefor 4th electrode Reversible by means of button PNP-NPN By means of th