1、 43 Power Quality Analyzer Service Manual 4822 872 05377 September 1998, Rev. 3, 07/00 1998, 2000 Fluke Corporation, All rights reserved. Printed in the Netherlands All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 SERVICE CENTERS To locate an authorized service
2、 center, visit us on the World Wide Web: or call Fluke using any of the phone numbers listed below: +1-888-993-5853 in U.S.A. and Canada +31-402-678-200 in Europe +1-425-356-5500 from other countries RadioFans.CN 收音机爱 好者资料库 i Table of Contents ChapterTitlePage 1Safety Instructions. 1-1 1.1 Introduct
3、ion. 1-3 1.2 Safety Precautions. 1-3 1.3 Caution and Warning Statements. 1-3 1.4 Symbols. 1-3 1.5 Impaired Safety. 1-4 1.6 General Safety Information. 1-4 2Characteristics. 2-1 2.1 Introduction. 2-3 2.2 Safety Specifications. 2-3 2.3 Function Specifications . 2-4 2.3.1 Electrical functions. 2-4 2.3.
4、2 Scope. 2-5 2.3.3 Meter. 2-7 2.3.4 Record. 2-7 2.4 Miscellaneous. 2-8 2.5 Current Probe. 2-8 2.6 Environmental Conditions. 2-9 2.7 Electromagnetic Immunity. 2-10 3Circuit Descriptions. 3-1 3.1 Introduction. 3-3 3.2 Block Diagram. 3-3 3.2.1 Input 1 - Input 2 Measurement Circuits . 3-3 3.2.2 Trigger
5、Circuit. 3-5 3.2.3 Digital Circuit. 3-5 3.2.4 Power Circuit. 3-6 3.2.5 Start-up Sequence, Operating Modes. 3-7 3.3 Detailed Circuit Descriptions. 3-9 3.3.1 Power Circuit. 3-9 3.3.2 Input 1 - Input 2 Measurement Circuits . 3-14 3.3.3 Trigger Circuit. 3-20 RadioFans.CN 收音机爱 好者资料库 43 Service Manual ii
6、3.3.4 Digital Circuit. 3-24 4Performance Verification . 4-1 4.1 Introduction. 4-3 4.2 Equipment Required For Verification. 4-3 4.3 How To Verify. 4-3 4.4 Display and Backlight Test. 4-4 4.5 Input 1 and Input 2 Tests in the SCOPE MODE. 4-5 4.5.1 Input 1 Trigger Sensitivity Test. 4-6 4.5.2 Input 1 Fre
7、quency Response Upper Transition Point Test. 4-7 4.5.3 Input 1 Frequency Measurement Accuracy Test. 4-7 4.5.4 Input 2 Frequency Measurement Accuracy Test. 4-9 4.5.5 Input 2 Trigger Level and Trigger Slope Test . 4-10 4.5.6 Input 2 Trigger sensitivity Test. 4-12 4.5.7 Input 1 Trigger Level and Trigge
8、r Slope Test . 4-13 4.5.8 Input 1 and 2 DC Voltage Accuracy Test. 4-15 4.5.9 Input 1 and 2 AC Voltage Accuracy Test. 4-17 4.5.10 Input 1 and 2 AC Input Coupling Test. 4-19 4.5.11 Input 1 and 2 Volts Peak Measurements Test. 4-20 4.5.12 Input 1 and 2 Phase Measurements Test. 4-21 4.5.13 Input 1 and 2
9、High Voltage AC 50 V; 500 V; 1250 V minimum step.1 mV; 10 mV; 100 mV; 1 V (1 % + 10 counts) Current ranges (auto). 50.00 A to 500.0 kA, 1250 kA minimum step. 10 mA; 0.1 A; 1 A; 100 A, 1 kA (1 % + 10 counts) Frequency range. 10.0 Hz to 15.0 kHz 40.0 to 70.0 Hz.(0.5 % + 2 counts) resolution. 0.1 Hz; 1
10、 Hz; 10 Hz; 100 Hz Power Readings.Watt, VA, VAR, PF, DPF, Hz Watt, VA, VAR ranges (auto) .250 W to 250 MW, 625 MW, 1.56 GW when selected: total ( %r):.(2 % + 6 counts) when selected: fundamental ( %f):.(4 % + 4 counts) DPF .0.00 to 1.00 0.00 to 0.25.not specified 0.25 to 0.90. 0.04 0.90 to 1.00. 0.0
11、3 Characteristics 2.3 Function Specifications2 2-5 PF. 0.00 to 1.00 0.04 Frequency range.10.0 Hz to 15.0 kHz 40.0 to 70.0 Hz.(0.5 % + 2 counts) Harmonics Number of harmonics . DC.21, DC.33, DC.51 Readings / Cursor readings Vrms. fund. (3 % + 2 counts).31 st (5 % + 3 counts).51st (15 % + 5 counts) Ir
12、ms. fund. (3 % + 2 counts).31 st (5 % + 3 counts).51st (15 % + 5 counts) Watt. fund. (5 % + 10 counts).31 st (10 % + 10 counts)51st (30 % + 5 counts) Frequency of fundamental. 0.25 Hz Phase.fund. 3 . 51 st 15 K-factor (in Amp and Watt). 10 % Sags so only one of these functions needs to be verified.
13、4.2 Equipment Required For Verification The primary source instrument used in the verification procedures is the Fluke 5500A. If a 5500A is not available, you can substitute another calibrator as long as it meets the minimum test requirements. Fluke 5500A Multi Product Calibrator, including 5500A-SC
14、 Oscilloscope Calibration Option. Stackable Test Leads (4x), supplied with the 5500A. 50 Coax Cables (2x), Fluke PM9091 (1.5m) or PM9092 (0.5m). 50 feed through terminations (2x), Fluke PM9585. Fluke BB120 Shielded Banana to Female BNC adapters (2x), supplied with the Fluke 43. Dual Banana Plug to F
15、emale BNC Adapter (1x), Fluke PM9081/001. Dual Banana Jack to Male BNC Adapter (1x), Fluke PM9082/001. 4.3 How To Verify Verification procedures for the display function and measure functions follow. For each procedure the test requirements are listed. If the result of the test does not meet the req
16、uirements, the test tool should be recalibrated or repaired if necessary. Follow these general instructions for all tests: For all tests, power the test tool with the PM8907 power adapter. The battery pack must be installed. Allow the 5500A to satisfy its specified warm-up period. For each test poin