FLUKE 87_89iv 电路图.pdf
《FLUKE 87_89iv 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 87_89iv 电路图.pdf(38页珍藏版)》请在收音机爱好者资料库上搜索。
1、 87 a Caution identifies conditions and actions that may damage the Meter or the test instruments. International Symbols International symbols used on the Meter and in this manual are explained in Table 1. Table 1. International Symbols Symbol Meaning Symbol Meaning Alternating current Earth ground
2、Direct current Fuse Alternating or direct current Conforms to European Union directives Refer to the manual. Important information. Conforms to relevant Canadian Standards Association directives Inspected and licensed by TV Product Services. Double insulated Battery Underwriters Laboratories, Inc. S
3、afety Information 3 Safety Information Warnings and Precautions To avoid possible electric shock or personal injury, follow these guidelines: Do not use the Meter if it is damaged. Before you use the Meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulat
4、ion surrounding the connectors. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads before you use the Meter. Do not use the Meter if it operates abnormally. Protection may be impaired. When in doubt, have the Meter serviced
5、. Do not operate the Meter around explosive gas, vapor, or dust. Do not apply more than the rated voltage, as marked on the Meter, between terminals or between any terminal and earth ground. Before use, verify the Meters operation by measuring a known voltage. When measuring current, turn off circui
6、t power before connecting the Meter in the circuit. Remember to place the Meter in series with the circuit. When servicing the Meter, use only specified replacement parts. Use caution when working above 30 V ac rms, 42 V peak, or 60 V dc. Such voltages pose a shock hazard. Avoid working alone. When
7、using the probes, keep your fingers behind the finger guards on the probes. Connect the common test lead before you connect the live test lead. When you disconnect test leads, disconnect the live test lead first. Remove test leads from the Meter before you open the battery door. Do not operate the M
8、eter with the battery door or portions of the cover removed or loosened. To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the low battery indicator () appears. Use only type AA batteries, properly installed in the Meter case, t
9、o power the Meter. Caution To avoid possible damage to the Meter or to the equipment under test, follow these guidelines: Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance. Use the proper terminals, function, and range fo
10、r your measurements. Before measuring current, check the Meters fuses and turn power OFF to the circuit before connecting the Meter to the circuit. 87 updates 4/second. Analog: 51 segments, updates 40/second. Temperature Operating: -20 C to +55 C; Storage: -40 C to +60 C Altitude Operating: 2000 m;
11、Storage: 10,000 m Temperature Coefficient 0.05 x (specified accuracy) / C (28 C) Electromagnetic Compatibility (EMC) Susceptibility and Emissions: Commercial Limits per EN61326-1 Relative Humidity 0 % to 90 % (0 C to 28 C); 0 % to 70 % (28 C to 55 C) Battery Type 4 AA Alkaline, NEDA 15A or LR6 Batte
12、ry Life 72 hrs typical (with backlight off) Shock Vibration Per MIL-T-PRF 28800 for Class II instruments Size 3.94 in x 8 in x 1.97 in (10.0 cm x 20.3 cm x 5.0 cm) (Not Including Accessory Mount) Weight 1.2 lbs (545 g) Case Sealing IP-42 per IEC 529, Section 3 Warranty Lifetime Calibration Interval
13、1 year Compliances Complies with ANSI/ISA S82.01-94, CSA 22.2 No. 1010.1- 92 to 1000 V Overvoltage Category III, Pollution Degree 2.* Certifications (listed and pending) CSA per standard CSA/CAN C22.2 No. 1010.1-92 UL per standard UL3111-1 (pending) TV per standard EN 61010 Part 1-1993 (pending) *CA
14、T III: OVERVOLTAGE (Installation) Category III, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Examples include electricity meter and primary over-current protection equ
15、ipment. 87 2.5 ms for single events. Frequency Counter Sensitivity Minimum Sensitivity (RMS Sine Wave) 1 Input Range 40 Hz to 20 kHz 2 15 to 40 Hz 3 20 to 500 kHz 3 Approximate Trigger Levels (DC Voltage Function) 50 mV (dB only) 15 mV to 3 mV 15 mV 15 mV 25 mV 500 mV 50 mV to 30 mV 50 mV 30 mV 35 m
16、V 6 mV 3000 mV 1500 mV to 300 mV 2000 mV 2000 mV 170 mV 6 mV 5 V 1.5 V to 0.3 V 2 V 2.2 V 1.7 V 0.25 V 50 V 15 V to 3 V 15 V 5 V 3.5 V 2.5 V 500 V 50 V to 20 V 50 V 50 V 35 V 25 V 1000 V 250 V to 100 V 250 V 50 V 35 V 25 V 1. Maximum input for specified accuracy = 10 x Range or 1000 V, ac-coupled on
17、ly. Accuracy: add 1 count 2. Sensitivity improves linearly from 40 Hz to 20 kHz. 3. Useable at reduced sensitivity to 0.5 Hz and 999.99 kHz. 87 & 89 Series IV Service Manual 10 Burden Voltage (A, mA, A) FunctionRangeBurden Voltage (typical) mA - A500.00 A102 V / A 5,000 A102 V / A 50.000 mA1.8 mV /
18、mA 400.00 mA1.8 mV / mA A5.0000 A0.04 V / A 10.000 A0.04 V / A Input Characteristics FunctionInput Impedance (Nominal) Volts, mV10 M, 100 dB -dc, 50 Hz, or 60 Hz 0.1%90 dB at 50 Hz, or 60 Hz 0.1% AC Volts, mV 90 dB dc to 60 Hz Full-Scale Voltage Open Circuit Test VoltageTo 5 M30 M + nS Ohms 5 V500 m
19、V3.1 V Diode Test 5 V3.1000 V Typical Short-Circuit Current 500 5 k50 k500 k5 M30 M Ohms100 A100 A10 A1 A0.1 A0.1 A Diode Test0.8 mA typical Required Equipment 11 Required Equipment Required equipment is listed in Table 2. If the recommended models are not available, equipment with equivalent specif
20、ications may be used. Repairs or servicing should be performed only by qualified personnel. Table 2. Required Equipment EquipmentRequired CharacteristicsRecommended Model CalibratorAC Voltage Range: 0-1000 Vac Accuracy: + 0.1 % Frequency Range: 20 Hz 100 kHz Accuracy: +3 % DC Voltage Range: 0-1000 V
21、dc Accuracy: + 0.006 % Current Range: 0-10A Accuracy: AC mA - +0.18 % AC A - + 0.38 % Frequency range: 1 kHz 20 kHz DC mA - + 0.06 % DC A - + 0.125 % Frequency Source: 0.5-20 kHz Accuracy: + 0.00125 % Amplitude: 1 Volt Accuracy: + 5 % Ohms Range: 0 - 50 M Accuracy: + 0.0125 % Capacitance Range: 110.
22、0 nF & 11.00 uF Accuracy: +0.25% Temperature Range: -10 to 350 C Accuracy: 0.3% Fluke 5520A Thermocouple Adapter Accessory K-typeFluke 80AK K-type Thermocouple K-type, mini-plug on both endsN/A Infrared (IR) AdapterFluke P/N 690518 87/89 IV Service Software Fluke P/N 676152 87 & 89 Series IV Service
23、 Manual 12 Basic Maintenance Warning To avoid electrical shock or personal injury, remove the test leads and any input signals before opening the case or replacing the battery or fuses. To prevent damage or injury, install ONLY batteries or fuses specified for this product. Opening the Meter Case Ca
24、ution To avoid unintended circuit shorting, always place the uncovered Meter assembly on a protective surface. When the case of the Meter is open, circuit connections are exposed. To open the Meter case, do the following: 1. Disconnect test leads from any live source, turn the rotary switch to OFF,