1、T.O. 33K1-4-1968-1 15 OCTOBER 1991 TECHNICAL MANUAL CALIBRATION PROCEDURE FOR HIGH POTENTIAL TESTER 4045AI (ASSOCIATED RESEARCH) Distribution Statement - Distribution authorized to U.S. Government agencies and their contractors for official use or for administrative or operational purposes only, 15
2、October 1991. Other requests for this document shall be referred to AFMETCAL Detachment 1/MLLW, 813 Irving-Wick Dr W, Suite 4M, Heath, OH 43056-6116. Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent disclosure of the contents or reconstruction of the
3、document. Published under Authority of the Secretary of the Air Force RadioFans.CN 收音机爱 好者资料库 T.O. 33K1-4-1986-1 1 HIGH POTENTIAL TESTER 4045AI (ASSOCIATED RESEARCH) 1 CALIBRATION DESCRIPTION: Table 1. Test Instrument (TI)PerformanceTest CharacteristicsSpecificationsMethod DC VoltageRange: 0 to 5 kV
4、Measured with a DMM and Voltage Divider Accuracy: 2.0% FS AC VoltageRange: 0 to 5 kVMeasured with a DMM Accuracy: 2.0% FS 2 EQUIPMENT REQUIREMENTS: Minimum UseCalibrationSub- NounSpecificationsEquipmentItem 2.1 DC VOLTMETERRange: 0 to 10 VDCFluke 8506A Accuracy: 0.1% 2.2 DIGITAL MULTIMETERRange: 0 t
5、o 1100 VACFluke 8600A Accuracy: 0.1% 2.3 VOLTAGE DIVIDERRange: 0 to 5 kVDCFluke 80E-10 Accuracy: 0.1% 2.4 RESISTORRange: 3 M, 1/2 wattAs Avail. Accuracy: N/A 3 PRELIMINARY OPERATIONS: 3.1 Review and become familiar with entire procedure before beginning calibration process. RadioFans.CN 收音机爱 好者资料库 T
6、.O. 33K1-4-1968-1 2 WARNING Unless otherwise designated, and prior to beginning the Calibration Process, ensure that all test equipment voltage and/or current outputs are set to zero (0) or turned off, where applicable. Ensure that all equipment switches are set to the proper position before making
7、connections or applying power. 3.2 Ensure that TI is properly grounded, connect all case grounds to facility ground. WARNING This procedure involves voltages that are hazardous to personnel. All necessary precautions during the conduct of these tests must be observed. 3.3 Connect TI and test equipme