1、M DLRO 200 High Current, Low Resistance Ohmmeter USER MANUAL DLRO200_UG_En_V01 22/04/2004 4:26 pm Page 1 RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS THESE SAFETY WARNINGS MUST BE READ AND UNDERSTOOD BEFORE THE INSTRUMENT IS USED. IDLRO200 is supplied without a plug terminating the supply lead. An appro
2、priate plug must be fitted prior to use. IThe DLRO200 must be grounded (earthed) when in use. If the ground / earth in the socket outlet being used to energise the DLRO200 cannot be relied upon, the instrument must be connected to a reliable earth / ground via the large green and yellow terminal on
3、the left hand side of the instrument. If no ground / earth is present the instrument will refuse to start a test. ICircuits must be de-energised before testing. The DLRO200 must only be used on dead systems. If the item under test has a voltage upon it greater than 10 volts ac peak or dc with refere
4、nce to ground the DLRO200 will indicate the presence of the voltage and prevent any tests being carried out. ITesting inductive circuits can be hazardous: The DLRO200 is a high power instrument, designed for testing resistive loads. It must NOT be used to test inductive loads. IDuring use the DLRO20
5、0, its current leads and the sample being tested can become hot. This is normal and is caused by the passage of high currents. Take care when touching the instrument, current leads, clips and the test sample. IThis product is not intrinsically safe. Do not use in an explosive atmosphere. IEnsure tha
6、t the power switch is always accessible so that power can be disconnected in an emergency. IIf this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 2 DLRO200_UG_En_V01 22/04/2004 4:26 pm Page 2 RadioFans.CN 收音机爱 好者资料库 SYMBOLS
7、 USED IN THIS DOCUMENT AND ON THE INSTRUMENT uProtective Conductor Terminal sHot Terminals GRefer to the User Manual for more information 1Power On 0Power Off cEquipment complies with relevant EU Directives CONTENTS 3 SAFETY WARNINGS2 GENERAL DESCRIPTION4 OPERATION5 Main Menu Screen5 TEST Menu6 OPTI
8、ONS Menu7 CURRENT Menu9 Warning Messages9 Entering Notes In The Memo Screen9 Test Result Storage10 Real Time downloading of test data.10 TEST TECHNIQUES TEST, OPTIONS and “CURRENT. Below this are details of the test current selected (SET CURRENT), maximum current to be allowed (CURRENT LIMIT), the n
9、umber of stored tests and the current date and time. To perform a measurement is straightforward. 1. Use the Left and Right arrows of the two-axis cursor control to highlight the TEST menu. Use the down arrow to select the required test type. Press Enter. 2. Use the Left and Right arrows of the two-
10、axis cursor control to highlight the CURRENT menu. Use the down arrow to select SET and press Enter. Type in the desired current using the numeric keys on the keypad. When finished press Enter. If the current you have set is greater than the current limit, DLRO200 will beep and clear the SET CURRENT
11、 field. Re-enter the desired current, if necessary increasing the current limit to a value equal to or greater than the desired level. (See separate topic for setting the current limit). 3. Connect the sample to be tested to the instrument terminals and press the TEST button. The different test type
12、s have slightly different connection requirements, which are detailed, in the sections below. Please note that if your DLRO200 is inadequately grounded / earthed it OPERATION 5 Terminate the supply lead with a plug suitable for your circumstances. Plug the instrument supply lead into a socket outlet
13、 and switch the instrument on using the On/Off switch situated on the left side of the instrument. The POWER lamp will illuminate, the instrument will display the firmware version number and perform a calibration check. If this is successful the display will show CALIBRATED. If the check fails the d
14、isplay will show NOT CALIBRATED”. After a slight pause the display will change to the Main Menu screen. DLRO 200 V1.00 COPYRIGHT MEGGER LIMITED 2002 INITIALISING HARDWARE CALIBRATED TEST OPTIONS CURRENT SET CURRENT150 CURRENT LIMIT 200 23 TESTS STORED 21/03/02 15:13 200 200 DLRO200_UG_En_V01 22/04/2
15、004 4:26 pm Page 5 6 OPERATION will not be possible to start a test. TEST Menu The TEST menu allows the selection of one of three test modes - Normal, Automatic or Continuous. Only one mode can be active at a time and the active mode is displayed below the heading TEST when you return to the MAIN ME
16、NU screen. Normal Mode Normal mode makes a single measurement of the resistance of the sample under test. Please note that in this mode both the current and voltage leads must be connected across the test sample before the Test button is pressed. When the TEST button is pressed DLRO200 will check fo
17、r good contact in the P circuit. If the P circuit continuity is faulty the display will show POTENTIAL FAIL. Press the ENTER key to remove this message and adjust the contact of the P probes to ensure good contact is achieved. Press the TEST button again. P circuit continuity will be checked and if
18、satisfactory DLRO200 will proceed to carry out a test. After a short period of time, if there is inadequate contact in the C circuit the instrument will display CURRENT LOOP FAIL. Press Enter to remove this message, make good the connection and press the TEST button to start the test. A sequence of
19、bars will appear across the display while the test is being completed. To make another measurement, ensure the test leads are connected and press the Test button. NORMAL mode is abbreviated to NORM under the TEST heading in the main menu screen. Continuous Mode Continuous mode requires the current l
20、eads and the voltage leads to be securely connected to the item under test before pressing the Test button. The DLRO200 checks voltage and current loop continuity as in Normal Mode and, if adequate, passes a continuous d.c. current, repeating resistance measurements at approximately 2-second interva
21、ls until the test is terminated by pressing the Test button (or until the DLRO200 overheats if particularly high currents are used). CONTINUOUS mode is abbreviated to CONT under the TEST heading in the main menu screen. Overheating When operating continuously at 200 A, overheating will occur after a
22、pproximately 15 minutes (assuming a starting temperature of 20C). A message HOT will then appear on the display until the internal parts have cooled. When the HOT message is on the screen, test current is turned off and all front panel controls are disabled. Operating at reduced current will conside
23、rably extend the operating time. Automatic Mode In Automatic mode, only connect the current leads to the sample to be measured. Press the TEST button. The red status lamp next to the TEST button will flash signifying that DLRO200 is armed and will remain in a state of readiness until the P circuit i
24、s completed. At this time the light will illuminate continuously and the test will be carried out. The progress of the test is shown by a series of bars progressing across the display. To make another measurement it is necessary to break P probe contact with the test sample and remake contact. For e