1、M MIT300 Series Insulation and Continuity testers USER MANUAL MIT300_UG_en_V03 21/09/2004 2:50 pm Page 1 RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS ISafety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use. IThe circuit under test must
2、 be switched off, de-energised and isolated before test connections are made when carrying out insulation and continuity tests. ICircuit connections and exposed metalwork of an installation or equipment under test must not be touched. IThe live circuit warning and automatic discharge are additional
3、safety features which may fail and therefore safe working practices must be observed. IThe voltage function will only work if the instrument is functional and switched on. IAfter insulation tests, capacitive circuits must be allowed to discharge before disconnecting test leads. IThe instrument shoul
4、d not be used if any part of it is damaged. ITest leads, probes and crocodile clips must be in good order, clean and with no broken or cracked insulation. IEnsure that hands remain behind guards of probes/clips when testing. INational Safety Authorities may recommend the use of fused test leads when
5、 measuring voltage on high-energy systems. IReplacement fuses must be of the correct type and rating. Failure to fit the correctly rated fuse may result in a safety hazard and may cause damage to the instrument in the event of an overload. IThe battery cover must be in place whilst conducting tests.
6、 2 Users of this equipment and/or their employers are reminded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuit
7、s. Where the assessments show that the risk is significant then the use of fused test leads may be appropriate. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. MIT300_UG_en_V03 21/09/2004 2:50 pm Page 2 RadioFans.CN 收音机爱 好者资料库 Insulation testing15 Test Lock15 SP3 Swi
8、tched (optional accessory for MIT320)16 Test result storage (MIT330 only)16 Replacing batteries and fuses18 Low battery warning symbol18 To replace batteries18 Fuse replacement (user replaceable)19 Preventive maintenance19 Technical specification20 Basic and service errors22 Accessories and equipmen
9、t23 Repair and Warranty24 Symbols used on the instrument are: FCaution: risk of electric shock GCaution: refer to accompanying notes tEquipment protected throughout by Double Insulation (Class II) cEquipment complies with current EU directives. Equipment complies with C tick requirements CONTENTS 3
10、Safety warnings4 Introduction2 General description4 Unpacking the carton5 LCD Display6 Front panel (digital instruments)6 Connection panel7 Lid open/closure7 Preparations for use (all instruments)8 Batteries8 Preliminary test lead check8 General operating instructions8 General functions8 Test lead c
11、onnection.9 Continuity measurement (MIT300, 310, 320, 330)10 Test lead Null10 Continuity test (MIT300, 310,320, 330)10 Continuity Buzzer (MIT300, 310, 320, 300)11 Continuity buzzer mode:11 Setup mode (MIT320/330 only)11 Insulation testing - General12 Insulation test12 Test lock13 Insulation test (M
12、alarm ON/OFF (MIT320 only) 13 M threshold adjustment (MIT320/330 only)13 Insulation test lock disable (MIT320/330 only)13 AC / DC Voltage measurement14 Analogue Instrument MIT310A14 Test lead Null 14 Continuity test 15 Continuity buzzer test15 N13117 MIT300_UG_en_V03 21/09/2004 2:50 pm Page 3 RadioF
13、ans.CN 收音机爱 好者资料库 The MIT300 series variants have the following features: MIT300 MIT310 MIT310A MIT320 MIT330 Insulation testing 1 kV Insulation rangeIIII 500 V Insulation rangeIIIII 250 V Insulation rangeIIIII Insulation test full scale999 M999 M999 M999 M 999 M Insulation limit alarm 0.01 M to 1 G
14、II Continuity Voltmeter range (600 V digit)IIII Continuity buzzerIIIII Adjustable buzzer threshold 1 to 20 II Audible buzzer disableII Lead null to 9.0 (0.5 for MIT310A)II0.5II Stored test lead nullIIIII Resistance 10 to 1M resistance rangeII INTRODUCTIONGENERAL DESCRIPTION 4 Thank you for purchasin
15、g the Megger Insulation Tester. For your own safety and to get the maximum benefit from your instrument, please ensure that you read and understand the following safety warnings and instructions before attempting to use the instruments. This user guide describes the operation and functions of the fo
16、llowing: MIT300 Series of Insulation and Continuity Testers: MIT300 MIT310 MIT310A MIT320 MIT330 MIT300_UG_en_V03 21/09/2004 2:50 pm Page 4 Unpacking the carton Unpack the carton contents carefully. There are important documents that you should keep for future reference. Please complete the pre-paid
17、 Warranty Card and return it to Megger Limited as soon as possible to help us reduce any delays in supporting you, should the need arise. Carton contents 1 x MIT300 series insulation tester 1 xRed/Black test lead set with clips 8 xAA (LR6) batteries fitted 1 xTest lead case 1 xWarranty card 1 x Cert
18、ificate of test 1 xCalibration certificate 1 x Owners CD manual 1 x Safety warnings 1 xPrinted Quick Start guide 1xUSB lead 5 MIT300 MIT310 MIT310A MIT320 MIT330 Features Locking test buttonIIIII Backlight and rotary switch illuminationsII Voltage warningI Default voltmeterIIII Auto-power-down with
19、defeatIIII Switched test probe connectorII Switched test probeII Results StorageI DownloadingI USBI MIT300_UG_en_V03 21/09/2004 2:50 pm Page 5 6 LCD DISPLAYFRONT PANEL (DIGITAL INSTRUMENTS) MIT300, 310 and 320 MI MIT310A Voltage warning LED Test Lock LED Test lead null Battery check Battery conditio
20、n indication Moving coil Analogue display indication MIT300_UG_en_V03 21/09/2004 2:50 pm Page 6 Memory control keys: STORE:Store initiates the storing of a test result. LAST/NEXT:Selects the type of location; ie Job, distribution board, circuit, phase etc. ESC:Aborts a save at any time. OK:Final ope
21、ration to save the result. Selects the job, db, circuit number; ie 01,02,03 L etc M Lid open/closure 1. Open lid by lifting up front panel tab (1). 2. Fold-away underneath instrument (2 allow the display to settle. The reading will remain displayed for a few seconds after the push button is released
22、. 4. Release the TEST button before removing the test leads (to enable the instrument to discharge the circuit under test). If the display shows VOLTS, wait. 5. On completion of testing switch to the OFF position. Alternatively auto shut-off operates after a period of inactivity. For MIT310A see not
23、es on Insulation Testing page 15 Note: To prevent accidental application of 1 kV, a warning display will flash 1000 V prior to the start of the first 1000 V test. Note: 50Hz noise from the circuit under test below 100 uA or 25 V is ignored and will not affect test accuracy. Higher levels of noise wi
24、ll cause the default voltmeter to indicate. Above 50 V (or 200 uA) testing is inhibited. Test lock The insulation test can be locked on by pressing the lock Lwhilst holding down the TEST button. To disable lock press the TEST or LOCK button. INSULATION TESTING - GENERAL MIT300_UG_en_V03 21/09/2004 2