1、Measurement Methods During network deployment, mainte- nance, and trouble shooting phases, insertion loss can be measured by dis- connecting the antenna and connecting an enclosed short at the end of the transmission line. If a Tower Mounted Amplifier (TMA) is used in the transmis- sion feed line sy
2、stem, it is best to remove the TMA and antenna from the system configuration to perform an insertion loss measurement. It is best to always dis- connect the cable at the same location so the measured data can be compared to the historical data for accuracy and repeatability. Using Site Master, cable
3、 insertion loss can be measured in CABLE LOSS or RETURN LOSS mode. In Cable Loss mode, Site Master automatically consid- ers the signal traveling in both directions thus making the measurement easier for the user in the field. The following section explains the pro- cedure to measure insertion loss
4、in cable loss mode and return loss mode. The measurement setup and equipment required is the same for both modes. Insertion Loss Measurement Methods APPLICATION NOTE Introduction Transmission feed line system performance plays an important role in wireless network coverage. Insertion loss measuremen
5、t is one of the critical measurements used to analyze transmission feed line installation and performance quality. This application note explains how Site Master is used to measure cable insertion loss with different test methods and how to predict the maxi- mum allowable cable insertion loss throug
6、h manual calculations. In wireless communication systems, the transmit and receive antennas are connected to the radio through coaxial cable and/or waveguide transmission lines (Figure 1). Insertion loss measures the energy absorbed by the transmission line in the direction of the signal path in dB/
7、meter or dB/feet. Transmission line losses are dependent on cable type, operating frequency and the length of the cable run. Insertion loss of a cable varies with frequency; the higher the frequency, the greater the loss. Insertion loss measurements help troubleshoot the network by verifying the cab
8、le installation and cable performance. High insertion loss in the feedline or jumpers can contribute to poor system performance and loss of coverage. Measuring insertion loss using Site Master assures accurate and repeatable measurements. SITE MASTER Verify Cable Insertion Loss From Ground Level Fig
9、ure 1. Typical transmission line system Jumper Connectors Main Feeder Antenna Jumper Antenna RadioFans.CN 收音机爱 好者资料库 2 Measuring Insertion Loss Using CABLE LOSS MODE Application Note Required Equipment Site Master Model S11xx, S33xx, or S251x Precision Open/Short, Anritsu 22N50 or Precision Open/Sho
10、rt/Load, Anritsu OSLN50LF Precision Load, Anritsu SM/PL Test Port Extension Cable, Anritsu 15NNF50-1.5C Optional 510-90 Adapter, DC to 7.5 GHz, 50 , 7/16(f)-N(m) Figure 3. Insertion loss measurement setup when TMA is in line. Step 4. Connect the phase stable test port extension cable to the RF port.
11、 Step 5. Calibrate the Site Master at the end of the phase stable test port extension cable. (See the section entitled “Calibrating Site Master with the Phase Stable Test Port Extension Cable” for details.) Step 6. Disconnect the antenna and connect an enclosed precision short at the end of the tran
12、smission line. Step 7. Connect the other end of the transmission line to the phase stable cable of the Site Master. A trace will be displayed on the screen when the Site Master is in continuous sweep mode. Step 8. Press the AMPLITUDEkey and set the TOP and BOTTOM values of the display. In Figure 4,
13、the TOP is set to 2 dB, and the BOTTOM is set to 5 dB. Step 9. Press the MARKERkey. Step 10.Set M1to MARKER TO PEAK. Step 11.Set M2to MARKER TO VALLEY. Step 12.Calculate the insertion loss by averaging M1 (MARKER TO PEAK) and M2 (MARKER TO VALLEY) values as follows: Step 13. Press SAVE DISPLAYand na
14、me the trace using the soft keys and press ENTER.Saving the display is recommended for historical documentation. Saved traces can be used in the future to compare, check or verify changes in the trans- mission feedline performance. Insertion Loss = (M1 dB + M2 dB) 2 = 3.23 + 3.95 = 3.59 dB 2 Procedu
15、re Step 1. Power On the Site Master and press the MODEkey. Step 2. Select FREQ-CABLE LOSSusing the Up/Down arrow key and press ENTER. Step 3. Set the start and stop frequen- cies F1and F2. As an example, F1=750 MHz and F2=850 MHz for a typical cellular frequency band. Insertion Loss Measurement Setu
16、p The insertion loss measurement set up for a typical transmission feed line system is shown in Figure 2. Remove the antenna and connect an enclosed precision “short” at the end of the transmission line. If a Tower Mounted Amplifier (TMA) is in the transmission feed line system, remove the TMA and a
17、ntenna and connect an enclosed short at the end of the transmission line. Insertion loss measurement for a transmission feed line system with a tower mounted amplifier is shown in Figure 3. Figure 2. An insertion loss measurement setup after antenna is removed. Site Master S251C MODEFREQ/DISTAMPLITU
18、DESWEEP 1 56 ENTER 8 MARKER 7 LIMIT 3 START CAL 4 AUTO SCALE SAVE SETUP RECALL SETUP 9 SAVE DISPLAY ESCAPE CLEAR 0 RECALL DISPLAY SYS PRINT . ON OFF RUN HOLD +/- 2 Test Port CableJumperMain Feeder Antenna Jumper Precision Short Antenna Removed Site Master S251C MODEFREQ/DISTAMPLITUDESWEEP 1 56 ENTER
19、 8 MARKER 7 LIMIT 3 START CAL 4 AUTO SCALE SAVE SETUP RECALL SETUP 9 SAVE DISPLAY ESCAPE CLEAR 0 RECALL DISPLAY SYS PRINT . ON OFF RUN HOLD +/- 2 Test Port Cable JumperMain Feeder TMA Jumper Antenna Jumper Precision Short Antenna Removed TMA Removed Tx/Rx RadioFans.CN 收音机爱 好者资料库 In Cable Loss mode,
20、the Site Master automatically considers the signal path in both directions when calculating the cable insertion loss. As such, Cable Loss mode is recommended when making cable insertion loss measurements. 3 Step 7. Connect the other end of the transmission line to the phase stable cable of the Site
21、Master. A trace will be displayed on the screen when the Site Master is in continuous sweep mode. Step 8. Press the AMPLITUDEkey and set the TOP and BOTTOM values of the display. In Figure 5, the TOP is set to 4 dB, and the BOTTOM is set to 10 dB. Step 9. Press the MARKERkey. Step 10.Set M1to MARKER
22、 TO PEAK. Step 11.Set M2to MARKER TO VALLEY. Step 12.Calculate the insertion loss by averaging M1 (MARKER TO PEAK) and M2 (MARKER TO VALLEY) values and dividing by two as follows: Step 13.Press SAVE DISPLAYand name the trace using the soft keys and press ENTER. Calibrating Site Master with the Phase
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- Anritsu 11410-00276 电路图 11410 00276
收音机爱好者资料库所有资源均是用户自行上传分享,仅供网友学习交流,未经上传用户书面授权,请勿作他用。



Marshall-9100-62-02-Schematic电路原理图.pdf
SEIBT 326GW 电路原理图.jpg