Biddle DLRO600_UG 电路图.pdf
《Biddle DLRO600_UG 电路图.pdf》由会员分享,可在线阅读,更多相关《Biddle DLRO600_UG 电路图.pdf(15页珍藏版)》请在收音机爱好者资料库上搜索。
1、M DLRO600 High Current, Low Resistance Ohmmeter USER MANUAL DLRO600_UG_en_V04.qxp 11/11/08 1:27 pm Page 1 RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS THESE SAFETY WARNINGS MUST BE READ AND UNDERSTOOD BEFORE THE INSTRUMENT IS USED. IDLRO600 is supplied without a plug terminating the supply lead. An appr
2、opriate plug must be fitted prior to use. The instrument can draw 25 amps when running at full current output so please ensure an appropriate plug is used. IThe DLRO600 must be grounded (earthed) when in use. If the ground / earth in the socket outlet being used to energise the DLRO600 cannot be rel
3、ied upon, the instrument must be connected to a reliable earth / ground via the large green and yellow terminal on the left hand side of the instrument. If no ground / earth is present the instrument will refuse to start a test. ICircuits must be de-energised before testing. The DLRO600 must only be
4、 used on dead systems. If the item under test has a voltage upon it greater than 10 volts ac peak or dc with reference to ground the DLRO600 will indicate the presence of the voltage and prevent any tests being carried out. ITesting inductive circuits can be hazardous: The DLRO600 is a high power in
5、strument, with an output power measured in kilowatts, designed for testing resistive loads. It must NOT be used to test inductive loads. IDuring use the DLRO600, its current leads and the sample being tested can become hot. This is normal and is caused by the passage of high currents. Take care when
6、 touching the instrument, current leads, clips and the test sample. IThis product is not intrinsically safe. Do not use in an explosive atmosphere. IEnsure that the power switch is always accessible so that power can be disconnected in an emergency. IIf this equipment is used in a manner not specifi
7、ed by the manufacturer, the protection provided by the equipment may be impaired. 2 DLRO600_UG_en_V04.qxp 11/11/08 1:27 pm Page 2 RadioFans.CN 收音机爱 好者资料库 SYMBOLS USED IN THIS DOCUMENT AND ON THE INSTRUMENT uProtective Conductor Terminal sHot Terminals GRefer to the User Manual for more information 1
8、Power On 0Power Off cEquipment complies with relevant EU Directives CONTENTS 3 SAFETY WARNINGS2 GENERAL DESCRIPTION4 OPERATION5 Main Menu Screen5 TEST Menu6 OPTIONS Menu7 CURRENT Menu9 Warning Messages9 Entering Notes In The Memo Screen9 Test Result Storage10 Real Time downloading of test data.10 TE
9、ST TECHNIQUES TEST, OPTIONS and “CURRENT. Below this are details of the test current selected (SET CURRENT), maximum current to be allowed (CURRENT LIMIT), the number of stored tests and the current date and time. To perform a measurement is straightforward. 1. Use the Left and Right arrows of the t
10、wo-axis cursor control to highlight the TEST menu. Use the down arrow to select the required test type. Press Enter. 2. Use the Left and Right arrows of the two-axis cursor control to highlight the CURRENT menu. Use the down arrow to select SET and press Enter. Type in the desired current using the
11、numeric keys on the keypad. When finished press Enter. If the current you have set is greater than the current limit, DLRO600 will beep and clear the SET CURRENT field. Re-enter the desired current, if necessary increasing the current limit to a value equal to or greater than the desired level. (See
12、 separate topic for setting the current limit). 3. Connect the sample to be tested to the instrument terminals and press the TEST button. The different test types have slightly different connection requirements, which are detailed, in the sections below. Please note that if your DLRO600 is inadequat
13、ely grounded / earthed it will OPERATION 5 Terminate the supply lead with a plug suitable for your circumstances. Plug the instrument supply lead into a socket outlet and switch the instrument on using the On/Off switch situated on the left side of the instrument. The POWER lamp will illuminate, the
14、 instrument will display the firmware version number and perform a calibration check. If this is successful the display will show CALIBRATED. If the check fails the display will show NOT CALIBRATED”. After a slight pause the display will change to the Main Menu screen. DLRO 600 V1.00 COPYRIGHT MEGGE
15、R LIMITED 2002 INITIALISING HARDWARE CALIBRATED TEST OPTIONS CURRENT SET CURRENT450 CURRENT LIMIT 500 23 TESTS STORED 21/03/02 15:13 DLRO600_UG_en_V04.qxp 11/11/08 1:27 pm Page 5 6 OPERATION not be possible to start a test. TEST Menu The TEST menu allows the selection of one of three test modes - No
16、rmal, Automatic or Continuous. Only one mode can be active at a time and the active mode is displayed below the heading TEST when you return to the MAIN MENU screen. Normal Mode Normal mode makes a single measurement of the resistance of the sample under test. Please note that in this mode both the
17、current and voltage leads must be connected across the test sample before the Test button is pressed. When the TEST button is pressed DLRO600 will check for good contact in the P circuit. If the P circuit continuity is faulty the display will show POTENTIAL FAIL. Press the ENTER key to remove this m
18、essage and adjust the contact of the P probes to ensure good contact is achieved. Press the TEST button again. P circuit continuity will be checked and if satisfactory DLRO600 will proceed to carry out a test. After a short period of time, if there is inadequate contact in the C circuit the instrume
19、nt will display CURRENT LOOP FAIL. Press Enter to remove this message, make good the connection and press the TEST button to start the test. A sequence of bars will appear across the display while the test is being completed. To make another measurement, ensure the test leads are connected and press
20、 the Test button. NORMAL mode is abbreviated to NORM under the TEST heading in the main menu screen. Continuous Mode Continuous mode requires the current leads and the voltage leads to be securely connected to the item under test before pressing the Test button. The DLRO600 checks voltage and curren
21、t loop continuity as in Normal Mode and, if adequate, passes a continuous d.c. current, repeating resistance measurements at approximately 2-second intervals until the test is terminated by pressing the Test button (or until the DLRO600 overheats if particularly high currents are used). CONTINUOUS m
22、ode is abbreviated to CONT under the TEST heading in the main menu screen. Overheating When operating continuously at 600 A, overheating will occur after approximately 3 minutes (assuming a starting temperature of 20C). A message HOT will then appear on the display until the internal parts have cool
23、ed. When the HOT message is on the screen, test current is turned off and all front panel controls are disabled. Operating at reduced current will considerably extend the operating time. Automatic Mode In Automatic mode, only connect the current leads to the sample to be measured. Press the TEST but
24、ton. The red status lamp next to the TEST button will flash signifying that DLRO600 is armed and will remain in a state of readiness until the P circuit is completed. At this time the light will illuminate continuously and the test will be carried out. The progress of the test is shown by a series o