1、5kV Digital Insulation Testers MEGGER BM11D MEGGER BM21 User Guide Guide de lutilisateur Gebrauchsanleitung Gua del usuario MEGGER RadioFans.CN 收音机爱 好者资料库 Note Users of this equipment and/or their employers are reminded that Health and Safety Legislation requires them to carry out valid risk assessm
2、ents of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as from inadvertent short circuits. Where the assessments show that the risk is significant then the use of test leads constructed in accordance with the HSE guidance note GS38 Ele
3、ctrical Test Equipment for use by Electricians should be used. 1 RadioFans.CN 收音机爱 好者资料库 Safety Warnings3 General Description4 Features and Controls5 - 6 Operation Instrument Controls7 - 8 Features8 Testing Procedures9 Battery9 - 10 Guard Terminal10 -11 Condition and Warning Indicators11 -12 Applica
4、tion Notes Preventive Maintenance13 Insulation Testing Techniques14 Short Time or Spot Test15 Polarization Index (PI) Test15 Stress Conditions16 Specification17 -19 Accessories20 Repair and Warranty21 Guide Utilisateur22 - 37 Betriebsanleitung38 - 53 Gua del Usuario54 - 69 Contents Symbols used on t
5、he instrument are: Caution: Refer to accompanying notes. Risk of electric shock. Equipment protected throughout by Double Insulation (Class II). Equipment complies with EU Directives 2 RadioFans.CN 收音机爱 好者资料库 3 SAFETY WARNINGS #The circuit under test must be switched off, de-energised and isolated b
6、efore insulation test connections are made. #Circuit connections must not be touched during an insulation test. #After insulation tests, capacitive circuits must be discharged before disconnecting the test leads. #Turning the instrument off, and manually discharging of circuits on completion of test
7、s is recommended as standard procedure, before touching any connections. #In certain circumstances, break-down of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner, possibly causing a loss of display while the circuit remains energised. In this event, it
8、 is even more important that the ON/OFF key is pressed, and the circuit manually discharged before touching any connections. #Test leads including crocodile clips must be in good order, clean and with no broken or cracked insulation. #Replacement fuses must be of the correct type and rating. #The in
9、strument should not be used if any part of it is damaged. #If water is present in the charger recess, the instrument must not be used. It is recommended that the instrument is not used in wet weather conditions. #See Battery Charging Power Cord on page 4. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUI
10、TABLY TRAINED AND COMPETENT PERSONS. The BM11D and BM21 are compact, battery powered, high voltage d.c. insulation testers with resistance measurement capabilities to around 1 T. The BM11D has a digital full scale of 500 G whilst the BM21 can measure up to 5 T. The BM21 can be set to display leakage
11、 current instead of resistance and this mode of operation also allows higher resistances to be measured, possibly up to 500 T. The instruments are microprocessor controlled and feature a large, clear LCD combining digital and analogue readings of insulation resistance. Both instruments have test vol
12、tage positions of 500 V, 1000 V, 2500 V and 5000 V. The BM21 has an additional range with a variable output voltage of 25 V to 5000 V in 25 V steps. An integral timer is started automatically on commencement of the test, and displays elapsed minutes and seconds since the start of the test. The BM21
13、timer can be used to set the duration of test, and will automatically stop the output voltage when the time has expired. At the end of a test, capacitive loads are discharged automatically and the decaying voltage is displayed as the value returns to zero. The test leads should not be disconnected u
14、ntil the item has been discharged fully. Similarly, during a test the test leads should not be removed, and the item under test should not be touched. A flashing LED, and flashing high voltage symbols on the LCD remind the user that hazardous voltage is present. A battery symbol is displayed continu
15、ously showing the condition of the battery. Power is obtained from internal mains rechargeable batteries. To recharge the batteries requires an input voltage within the range 95-265 V a.c., 50-60 Hz. The power input and fuses are located in a splashproof recess on the front panel. Battery Charging P
16、ower cord: If the power cord plug is not suitable for your type of socket outlets (receptacles), do not use an adaptor. You should use a suitable alternative power cord, or if necessary change the plug by cutting the cord and fitting a suitable plug. The colour code of the cord is: Earth (Ground)Yel
17、low/Green NeutralBlue Phase (Line)Brown If using a fused plug, a 3 Amp fuse to BS 1362 should be fitted. Note: A plug severed from the power cord should be destroyed, as a plug with bare conductors is hazardous in a live socket outlet (receptacle). 4 General Description 5 Features and Controls BM11D
18、 Front Panel Layout IL M 1000V 00:30 1T 100G 10G 1G100M 10M 1M 0.1M 0 L M +G FS1FS2 100mA(T) 250V IEC127/1 95-265V a.c. 50-60Hz 24VA 400mA 100mA 250V IEC127/1 + 10-15 V d.c. G 12 V Charger Input Mains Socket Sliding Cover (open) Terminal Voltage Cursor Cursor TimerSelector Keys On / Off Button Fuse
19、FS2 Fuse FS1 Warnings Battery Symbol Test Button HV Warning LED 6 BM21 Front Panel Layout IL ML M I R 1000V 00:30 1T 100G 10G 1G100M 10M 1M 0.1M 0 L M G On / Off Button Sliding Cover (closed) Terminal Voltage Battery Symbol Cursor Range Keys Selector Keys Warnings Timer Test Button HV Warning LED IN
20、STRUMENT CONTROLS On/Off Button. Pressing the On/Off button starts the instrument and initiates a self check and calibration. All LCD segments appear for 5 seconds, then the word CAL appears for about 5 seconds. When the word CAL disappears, the instrument is ready for use, and is in standby mode. T
21、o switch the instrument off, press the on/off button once. (Auto switch off will occur after 10 minutes of non-use). Selector Keys The test voltage is selected by moving the cursor. When the test button is pressed the instrument will operate at the voltage indicated. BM21: In addition to the fixed t
22、est voltages the cursor can be used to select the variable voltage and test duration positions. Range Keys (BM21 only) The range keys have two purposes. The first purpose is to select the measuring range, whether insulation resistance, leakage current or burn mode. The second function is indicated b
23、y two arrows which appear above the keys when either the variable voltage or test duration positions are selected. Using the range keys in this mode will raise and lower the applicable parameter. The change in voltage is shown at the top of the display, and the change in test duration is indicated i
24、n the bottom left hand corner of the display. Test Button Pressing the large TEST button for more than one second will initiate a test at the selected voltage, simultaneously starting the timer. Pressing the button again will terminate the test and connect the internal discharge resistors across the