1、DCM300E Earth Leakage Clamp Meter Safety Warnings Do not take measurements beyond maximum selected range. Take extreme care and keep hands behind the hand guard when taking measurements above 30 V RMS. Do not operate the instrument with the battery cover removed. Do not operate the instrument if any
2、 part of it is damaged. Do not operate the instrument in environments subject to high temperature, damp, humidity or excessive vibration. Safety Warnings must be read and understood before the instrument is used, and observed during use. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND
3、COMPETENT PERSONS The DCM300Eis a rugged lightweight pocket size clamp meter designed to measure a.c. earth leakage currents. This enables earth leakage faults to be detected and located without having to isolate and disconnect circuit wiring. Additionally, the DCM300E measures a.c. circuit currents
4、 up to 300 Amps. Powered by two LR44 or SR44 cells, the instrument design takes full advantage of microprocessor technology and features a clear 312 digit LCD combining digital and bar- graph analogue readings. The two position mode switch provides 4 ranges; 30 mA, 300 mA, 30 A and 300 A, with minim
5、um resolution of 0,01 mA on the 30 mA range. To conserve battery power, Auto shut off operates after a period of 10 minutes of inactivity by the instrument. The instrument can be switched back on by selecting Off and then On again. Transducer Jaws Jaw Guard Tactile Handguard Jaw Opening Lever 312Dig
6、it LCD Display On / Off SwitchMode Selector Switch (mA / A) Data Hold Switch Measurement of Leakage Current Earth Conductors 1)Set the range selector switch to the 30 /300 mA position. 2)Set the Power switch to On. 3)Ensure that the Data hold switch is Off.is not displayed. 4)Clamp the jaws around t
7、he earth conductor. 5)If necessary, press the Data hold switch. is displayed. 6)Take the reading from the Digital and/or Bar-graph display. Single phase or 3 phase conductors 1)Set the range selector switch to the 30 /300 mA position. 2)Set the Power switch to On. 3)Ensure that the Data hold switch
8、is Off.is not displayed. 4)Clamp the jaws around the two single phase conductors, or the 3 conductors in the case of 3 phase. 5)If necessary, press the Data hold switch.is displayed. 6)Take the reading from the Digital and/or Bar-graph display. Features and Controls General Description Wrist Strap p
9、oint Measurement of Line Current Battery Replacement 1)Set the range selector switch to the 30 /300 mA position. 2)Set the Power switch to On. 3)Ensure that the Data hold switch is Off.is not displayed. 4)Clamp the jaws around one conductor of the circuit under test. 5)If necessary, press the Data h
10、old switch. is displayed. 6)Take the reading from the Digital and/or Bar-graph display. 1)When the symbol appears on the display, the two 1,5 V button cells must be replaced. 2)Switch the DCM300E Off. 3) Loosen the small cross head screw securing the battery cover, and remove the cover. Take care no
11、t to lose the small screw. 4)Remove both exhausted cells and carefully fi t two new cells into the recess. Position both cells +ve side up. 5)Replace the battery compartment cover and re- secure with the small cross head screw. Battery Cover Screw The correct size screw must be used and fi tted to e
12、nsure battery cover security, and safe operation. RadioFans.CN 收音机爱 好者资料库 Specifi cation Display:312 digit L.C.D. Maximum Indication:3200. Over Range Indication: Data Hold Indication: Low Battery Indication:2.5 V 2,7 V Auto Shut-Off:Operates after approximately 10 minutes of instrument inactivity. A