1、G Safety Warnings and Precautions These must be read and understood before the instrument is used. They must be observed during use. IThe circuit under test must be switched off, de-energised and isolated before test connections are made when carrying out insulation and continuity tests. ICircuit co
2、nnections and exposed metalwork of an installation or equipment under test must not be touched. Remember remote conductors may be charged to the test voltage. IThe live circuit warning and automatic discharge are additional safety features which may fail and therefore safe working procedures/practic
3、es must be observed. IThe voltage function will only work if the instrument is functional and switched on. IAfter insulation tests, capacitive circuits must be allowed to discharge before disconnecting test leads. IThe instrument, test-leads, probes and crocodile clips must be in good order, clean a
4、nd with no broken or cracked insulation. IEnsure that hands remain behind guards of probes/clips when testing. INational Safety Authorities may recommend the use of fused test leads when measuring voltage on high-energy systems. IReplacement fuses must be of the correct type and rating. Failure to o
5、bserve this will result in a safety hazard. IThe rear cover must be in place whilst conducting tests. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. M MIT200 Series Insulation and Continuity Testers User guide Users of this equipment and/or their employers are remin
6、ded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments show that the risk is significant
7、then the use of fused test leads may be appropriate. Symbols used on the instrument: G Caution: refer to accompanying notes t Equipment protected throughout by Double Insulation (Class II) c Equipment complies with relevant EU Directives Equipment complies with C tick requirements Do not dispose of
8、in the normal waste stream Maximum input voltage 600 V rms Application BS EN 61010 defines measurement categories from I to IV relating transient over voltages and the location within the electrical installation. This instrument is designed for use at Category III (Building installation/distribution
9、 board level) up to 600 V (maximum) rms to earth and 600 V (maximum) rms between terminals. It may also be used at Category IV (Primary supply level) on 300 V rms phase to earth systems. Display layout Operation All circuits must be isolated before an insulation or continuity test. Automatic Voltage
10、 detection The instrument will automatically display voltage between 25 V and 600 Vac or dc when connected to the circuit under test. Remove the source of any voltage before continuing to test. Test lead check 1.Before each use of the instrument, inspect the test leads, prods and crocodile clips to
11、confirm that their condition is good, with no damaged or broken insulation. 2.Check continuity of the test leads by firmly shorting the leads together and read the test lead resistance measurement directly from the display. Backlight operation 1.Turn the instrument backlight on by selecting the posi
12、tion. Backlight stays on for 1 minute. 2.When the backlight is activated, select the desired test position. 3.Change range or press TEST to re-initialise backlight for another minute. Test lead null To remove test lead resistance before continuity tests: 1.Switch the instrument to 2.Short test leads
13、 together, wait for a stable reading and press the test button on the instrument. The display should read 0.00 and display the z symbol. Lead null value is automatically stored until reset. To remove, press the TEST button again with leads open circuit. Continuity Measurement 1.Set range switch to .
14、 2.Connect the red and black test leads to the circuit to be tested. The instrument will automatically display continuity resistance when connected (up to 99.9 ). NOTE: If “100.0 “ is displayed the resistance of the circuit is over-range. Continuity Buzzer 1.Set the instrument to range. 2.Connect th
15、e test leads to the circuit to be tested. 3.The buzzer will sound automatically when a circuit is made (2 V is detected the test will be inhibited. Insulation testing M all instruments 1.Set range switch to 250 V, 500 V or 1 kV as available. 2.Press and hold down the TEST button to start test. The d
16、isplay will settle, and show the insulation test resistance in M. 3.If a circuit voltage of 25 V is detected the instrument will give a warning Beep but allow testing to continue. If a circuit voltage 50 V is detected, further testing will be prevented. Warning: After testing ensure the display retu
17、rns to Zero (circuit is discharged) before removing test leads. Error messages: 100.0 Continuity over-range 1000 M Insulation test result is over-range UNC Uncalibrated E17Instrument not configured to a specific type. Battery replacement Low battery voltage is indicated by the symbol in the display.
18、 To maintain performance the batteries should be replaced at this point. Uses 6 x 1.5 V AA (LR6) alkaline dry cells or NiMH rechargeable batteries. To replace batteries switch off instrument and disconnect test leads from the circuit under test. Remove the rear battery cover and replace the batterie
19、s. Do not mix old and new batteries. Warning: OBSERVE CORRECT POLARITY. Incorrect battery polarity could cause batteries to leak and damage the instrument. Fuse replacement A blown fuse is indicated by the symbol in the display. To replace the fuse remove battery cover and replace fuse with type: 50
20、0 mA (F) HBC 50 kA 600 V . Cleaning: Wipe disconnected instrument with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA). Specification Insulation ranges Test voltage accuracy:-0% +25% over full operating temperature Nominal Test Voltage:1000 V , 500 V , 250 V (d.c.) Measuring Range
21、:10 k - 1000 M on all ranges Short Circuit Current: 2 mA +0 mA -50% Test Current on Load:1 mA at min. pass values of insulation (as specified in BS 7671, EN61557, HD 384 and IEC 364) Accuracy (at 20 C)3%, 2 digits (or 30% of reading 200 k to 10 M) Continuity ranges Measuring Range:0,01 - 100 (0 -50
22、on analogue scale) Open Circuit Voltage: 5 V 1 V Short Circuit Current:205 mA, 5 mA (0 to 10 ) (20 mA up to 100 ) Accuracy (at 20 C):3% of reading 2 digits up to 10 5% of reading 2 digits up to 100 M 30 of reading 2 digits up to 1000 M Null Offset Adjust: 0 9.99 Continuity Buzzer:Operates at 5 Respo
23、nse time 25 V a.c. or d.c. applied display will operate as a voltmeter Test inhibit:50 V AutoPower Down: Operates after 10 minutes if left in standby mode. Environmental Operating Range:-10C to +55C Operating Humidity:93% R.H. at +40C max. Storage Range:-25C to +65C Dimensions:195 mm x 98 mm x 40 mm
24、 Weight:550 gms Battery:6 x 1,5 V cells IEC LR6 type (AA alkaine) or NiMH rechargeable Battery life:3000 consecutive tests (5 sec/test) on any test using 2Ah batteries Fuses Terminals:500 mA (F) 600 V , 32 x 6 ceramic HBC 50 kA minimum Safety Meets the requirements of BS EN61010-1, Cat III 600 V pha