1、M AVO300 Series Digital Multimeters USER MANUAL MANUAL DE LUTILISATEUR BEDIENUNGSANLEITUNG MANUAL DEL USUARIO RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS The following safety information must be observed to insure maximum personal safety during the operation at this meter: IDo not use in wet environmen
2、ts IMeasurements beyond the maximum selected range must not be attempted. IExtreme care must be taken when measuring above 50 V , especially on live bus-bars. ITo measure voltage, the instrument must not be switched to a current or resistance range, or to the diode check or buzzer position. ICircuit
3、s must be de-energised and isolated before carrying out resistance tests. IThe rotary selector switch must only be turned after removing test connections. IAll external voltages must be disconnected from the instrument before removing the battery. ITest leads and prods must be in good order, clean,
4、and with no broken or cracked insulation. IUK Safety Authorities recommend the use of fused test leads when measuring voltage on high energy systems. IReplacement fuses must be of the correct type and rating. IThe instrument must not be used if any part of it is damaged. ICheck for correct instrumen
5、t operation by testing a known voltage before and after use. Do not use if misleading results are obtained. IWarnings and precautions must be read and understood before an instrument is used. They must be observed during the operation of this instrument. 2 Users of this equipment and/or their employ
6、ers are reminded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments show that the risk is
7、 significant then the use of fused test leads may be appropriate. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFans.CN 收音机爱 好者资料库 Symbols used on the instrument are: FCaution: risk of electric shock GCaution: refer to accompanying notes tEquipment protected t
8、hroughout by Double Insulation (Class II) cEquipment complies with current EU directives. Equipment complies with C tick requirements CONTENTS 3 Introduction4 Symbols and annuniciators 4 Operation4 Mode button4 Hold button4 Backlight button (AVO310 only)4 Max/Min button (AVO310 only)5 DC Voltage mea
9、surements5 AC Voltage measurements5 DC Current measurement5 AC Current measurement5 Resistance measurement6 Continuity check6 Diode test6 Specifications7 Battery and fuse replacement10 Battery installation10 Replacing the fuses10 Accessories11 Repairs and warranty12 MANUAL DE LUTILISATEUR13 BEDIENUN
10、GSANLEITUNG27 MANUAL DEL USUARIO41 N13117 RadioFans.CN 收音机爱 好者资料库 To turn on the instrument turn the range knob from the OFF position to any measurement range. Note: For best battery life ALWAYS turn the function switch to the OFF position when the meter is not in use. These meters have an Auto OFF
11、that automatically shuts the meter OFF if left switched on. AVO300:15 minutes, AVO310: 30 minutes. Note: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. T
12、he reading will stabilize and give a proper measurement when connected to a circuit. MODE button To select AC or DC measurement when in Amps, mA or uA ranges. HOLD button The HOLD function allows the meter to freeze a measurement for later reference. 1. Press the HOLD button to freeze the reading on
13、 the indicator. The HOLD message will be appear in the display. 2. Press the HOLD button again to return to normal operation. BACKLIGHT button (AVO310 only) 1. Press the BACKLIGHT button to switch on the display light. 2. Press the BACKLIGHT button again to exit the light mode. INTRODUCTIONOPERATION
14、 4 Thank you for purchasing one of the Megger AVO300 series digital multimeters. For your own safety and to get the maximum benefit from your instrument, please ensure that you read and understand the following safety warnings and instructions before attempting to use the instrument. This user guide
15、 describes the operation and functions of the following AVO300 series of digital multimeters. AVO300 AVO310 SYMBOLS AND ANNUNCIATORS Continuity BATLow Battery Diode test HOLDData Hold AUTOAutoRanging AC Alternating Current or Voltage DCDirect Current or Voltage MAX/MINStores the highest or lowest me
16、asurement Backlight VVolts A, mA, uA Current range decimal point, value and symbol (AC, V , etc.). DC CURRENT MEASUREMENT CAUTION: Do not make current measurements on the 10A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads. 1. Insert the bla
17、ck test lead plug into the negative (COM) jack. 2. For current measurements up to 4000 A DC, set the function switch to the uA position and insert the red test lead plug into the (A) jack. 3. For current measurements up to 400 mA DC, set the function switch to the mA range and insert the red test le
18、ad plug into the (mA) jack. 4. For current measurements up to 10 A DC, set the function switch to the A position and insert the red test lead plug into the 10 A jack. 5. Press the AC/DC button until DC appears in the display. 6. Remove power from the circuit under test, then open up the circuit at t
19、he point where you wish to measure current. 7. Connect the black test probe to the negative side of the circuit. Connect the red test probe to the positive side of the circuit. 8. Apply power to the circuit. 9. Read the current in the display. The display will indicate the proper decimal point, valu
20、e and symbol. 10. Remove power from circuit before disconnecting test leads 5 MAX/MIN button (AVO310 only) The MAX/MIN function allows the meter to capture the highest or lowest measurement for later reference. 1. Press the MAX/MIN button to begin measurement. The indicator MAX or MIN will appear in
21、 the display. 2. If the MAX MIN messages are flashing, the instrument is in MAX/MIN mode but not recording, press the MAX/MIN button to select a mode. 3. To return to normal AUTO measurement mode, hold down the MAX/MIN button for 2 seconds. DC VOLTAGE MEASUREMENTS 1. Set the function switch to the V
22、 DC position (mV will appear in the display). 2. Insert the black test lead plug into the negative (COM) jack and the red test lead plug into the positive (V) jack. 3. Connect the test probe tips to the circuit under test. Be sure to observe the correct polarity (red lead to positive, black lead to
23、negative). 4. Read the voltage in the display. The display will indicate the proper decimal point and value. If the polarity is reversed, the display will show (-) minus before the value. AC VOLTAGE MEASUREMENTS 1. Set the function switch to the V AC position. 2. Insert the black test lead plug into
24、 the negative (COM) jack and the red test lead plug into the positive (V) jack. 3. Connect the test probe tips to the circuit under test. 4. Read the voltage in the display. The display will indicate the proper 6 AC CURRENT MEASUREMENT CAUTION: Do not make current measurements on the 10 A scale for