1、Withstand Testers MHP1 to the left provides the unlatched mode, to the right provides the latched mode. Gently pull to remove the link from the terminals, and re-position as required. 6. Interface and Interlock terminals (under cover plate) To gain access to the interface terminals, remove the IEC m
2、ains connector from the instrument. The screw holding the cover in position is then removed and the cover can be swung up out of the way, exposing the interface terminals. The terminals are not suitable for solid core wires. Multi-stranded conductors, 22-16 AWG (1mm2- 1,5 mm2) should be used. The us
3、e of wire ferrules is recommended to eliminate stray strands of wire; recommended insulation strip length is 8 mm. The terminal screws should be tightened securely. On completion, the cover plate should be swung back into position and secured by replacing and tightening the cover plate retaining scr
4、ew. Terminal connections and fuses The interface terminals are in three groups:- i)Relay Terminals The top three terminals are for applications that require an external test status indication. Test Failed and Test OK can be signalled. The maximum current that can be switched is 8 Amp at 250 Volts. 7
5、 When the Failed condition is active, the Common terminal is shorted to the Normally Open (N/O) terminal. When the Failedcondition is active the Common terminal and the Normally Closed (N/C) terminal are open circuited by the instrument. To gain access to the relay terminals, the IEC line connector
6、must be removed from the instrument. The screw holding the cover in position is then removed and the cover can be swung up out of the way, exposing the interface terminals. ii)HV On Terminals Some regulations call for various signalling lamps to indicate hazardous and potential Risk situations. The
7、terminals shown allow for the connection of a Line powered High Voltage On indicator or beacon. See Accessories for Warning beacon. A maximum current of 12Amp can be drawn from these terminals. HV On terminal connections 9 Normal Open Common Phase Neutral Alarm Sounder Common Phase Neutral Normal Cl
8、osed Indicator Lamp Cable gland Cable gland Start mode linkStart mode link HV On HV On Cable gland Start mode link Line voltage at up to Amp 1 /2 iii)HV Interlock Terminals The bottom four terminals enable the HV supply to be safety interlocked. The instrument is supplied without the interlock termi
9、nals connected. The HV interlock terminals must be connected before the instrument will generate High Voltage. The Users HV safety interlock system must conform to National and Local safety codes and be designed in such a way that any contact with live parts is prevented. This is normally achieved b
10、y a solid barrier that must be removed before access can be obtained to the hazardous area. The action of removing the barrier forces the interlock switches to the open position thus interrupting the supply voltage for the HV generation circuit. The interlock connections operate at Line (mains) volt
11、age. All wiring connected to these terminals must rated for Line (mains) use and capable of carrying a current of 3 amps. Interlock connection example 7. Cable entry glands When wiring the safety interlocks, the wires to the terminals must pass through the cable entry glands and then be connected to
12、 the appropriate terminals. Cables of 1 8” to 1 4” (3 mm to 6.5 mm) outer insulation diameter can be used. The glands must be sufficiently tightened, after wiring, so that flexing the external cord does not cause any movement of the wires inside the panel. On completion, the cover plate should be sw
13、ung back into position and secured by replacing and tightening the cover plate retaining screw. Features, Controls and Connections 10 P2 - R Guard Op Optional Ready to ? Test Indicator P1 - S P1 - R P2 - S Cable gland Cable gland ? Line voltage ? on terminals Start mode link Guard in Open Position I
14、nterlock switch (user supplied) Testing Precautions High voltage testing is potentially lethal. Care must be taken at all times to ensure that risks to safety are eliminated. The instruments must only be used by suitably trained and competent persons. The instrument User is responsible for their own
15、 safety and that of others who may come into contact with dangerous voltages as a result of a test. For safety, the instrument must be double grounded (earthed) to known good ground (earth) point. Firstly by the line powercord, and secondly from the rear protective ground (earth) terminal. During th
16、e installation process both ground (earth) connections should be checked to prove their integrity. See Before each testing sequence. Before operating the Line On / Off switch, ensure that the voltage control is set to minimum. If this is not done a voltage transient could occur at the output termina
17、l which could cause damage to the item under test or endanger the user. When making connections to the instrument, and to the device to be tested, ensure the instrument is in a safe condition, with all safety interlocks open. If any part of the instrument is damaged, it should not be used, but retur
18、ned to the manufacturer or an approved organization for repair. Charged Circuit Discharge For d.c. tests and when testing capacitors or equipment containing capacitors, extra care must be taken by the User to ensure that the item under test is fully discharged before handling. The instrument provide
19、s a discharge path which will discharge 0,5 F in 10 seconds. Even a small deflection of the voltmeter may represent a hazardous voltage. Warning:- Care must be taken to prevent capacitive circuits becoming disconnected during a test, leaving the circuit in a charged state. Automatic discharge must n
20、ot be relied upon as an alternative to normal safe working practice. Dielectric Absorption Some capacitive items exhibit an effect called Dielectric Absorption. The dielectric can absorb charge during the test. After the item is discharged, the absorbed charge is released back onto the capacitive pl
21、ates effectively raising the voltage of the item, possibly to dangerous levels. Safe working practice dictates that tested items should be firmly shorted out with a shorting link, after discharge, until required for use. The instrument is supplied with a 3 wire power cord. The instrument must be use
22、d with a grounded (earthed) supply, and all 3 wires must be connected. The colour code of the cord is :- U.S.A.U.K. / International Ground (Earth)GreenYellow/Green NeutralWhite Blue Line (Phase)Black Brown If using a fused plug, a 3 Amp fuse to BS1362 should be fitted. Operation 11 Before each testi
23、ng sequence:- 1. Decide upon the Start mode required, and connect the Start mode link in the appropriate position. 2. Check the condition of the HV probe and test lead (or LV probe) to ensure that they are in good order, clean, with no broken or damaged insulation. A daily Continuity test is recomme
24、nded. To check the HV probe and test lead (or LV probe):- i)Plug the HV probe in the a.c. HV receptacle. ii)Ensure that the Low Return lead (or LV probe) is firmly secured to the Low return terminal. iii)Set the voltage control to minimum. iv)Set the trip mode to TRIP and set the trip current limit