1、M MIT510 5kV Digital Insulation Tester USER MANUAL RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS IThe circuit under test must be switched off, de-energised, isolated and checked to be safe before insulation test connections are made. Make sure the circuit is not re-energised whilst the instrument is conn
2、ected. ICircuit connections must not be touched during an insulation test. IAfter completing a test, capacitive circuits must be completely discharged before disconnecting the test leads. ITurning the instrument off, and manually discharging circuits on completion of tests is recommended as standard
3、 procedure, before touching the test leads. IIt is rare, but in certain circumstances, breakdown of the circuit under test may cause a loss of display, presence of HV on the test leads will still be shown by a flashing red LED. If this happens, stop the test immediately by pressing the TEST or Power
4、 ON/OFF button then ensure the circuit is manually discharged before touching any connections. ITest leads, including crocodile clips, must be in good order, clean and with no broken or cracked insulation. IThe instrument should not be used if any part of it is damaged. IWater on the surface of the
5、instrument may lead to erroneous results. It is recommended that the instrument is not used in wet weather conditions. IThis instrument is not intrinsically safe and must not be used in hazardous atmospheres. 2 NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFan
6、s.CN 收音机爱 好者资料库 Symbols used on the instrument are: FCaution: risk of electric shock GCaution: refer to accompanying notes tEquipment protected throughout by Double Insulation (Class II) cEquipment complies with current EU directives. CONTENTS 3 Description4 Power lead and charging the batteries4 Qu
7、ick start5 Instrument controls and indicators6 Power On/Off button6 VL and VM buttons6 Test button6 /I button6 Bbutton7 “HV On” LED7 “Charging indicator” LED7 Test terminals7 The guard terminal7 Factory programming socket7 Battery bar graph8 Voltage at terminals8 Timer8 Digital display8 Analogue dis
8、play8 Capacitance display9 Time constant9 Specification9 Accessories11 Repair and Warranty12 RadioFans.CN 收音机爱 好者资料库 If the power lead supplied is not suitable for your type of mains outlet, do not use an adaptor. Always use a power lead fitted with the correct plug. NB: A plug severed from the powe
9、r lead is a hazard if plugged into a live socket. Severed plugs must be destroyed immediately. The instrument is fitted with a two-pin IEC60320 power inlet. Most power leads are made with three-core cable so the ground connection will not be used. Power lead cable colour coding: InternationalUSA Ear
10、th/GroundYellow/GreenGreen NeutralBlueWhite Live (Line)BrownBlack If using a fused plug ensure it is fitted with a 3 Amp fuse. The instrument can be powered from 95 - 240 V rms, 10% 50/60Hz. With power connected the battery will charge as long as a test is not in progress. The Power On/Off button ha
11、s a green LED which illuminates when mains power is present. For best battery lifetime, charge the battery after each use. A completely exhausted battery will take 14 hours to recharge. INTRODUCTIONPOWER LEAD AND CHARGING 4 The MIT510 is a compact microprocessor controlled high voltage d.c. insulati