1、 PN 200409 September 1996 Rev. 1, 2/98 1996, 1998 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies. 1 36 AC/DC True Rms Clamp Meter Service Information WWarning Do not service the Meter unless you are qualified to do so. The service
2、 information provided in this document is for the use of qualified personnel only. To avoid electrical shock or personal injury: Always remove the test leads and input signals from the Meter before opening the case. Use caution when working with voltages above 60V dc or 30V ac. WWarning To avoid fal
3、se readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears. Y Caution The Meter is protected throughout by double insulation or reinforced insulation. When servicing the Meter, use only specified replacement parts. Cautio
4、n This Meter contains parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. Introduction This Service Information document provides the following service information for the 36 AC/DC True Rms Clamp Meter (hereafter referred to as the Mete
5、r): Product specifications Required equipment Performance test procedure and test values Preparing for calibration Calibration procedure and adjustments AC amps test Replacement parts For operating information, refer to the Instruction Sheet that came with the Meter. RadioFans.CN 收音机爱 好者资料库 36 AC/DC
6、 True Rms Clamp Meter 2 Service The Meter is warranted to be free from defects in material and workmanship for one year. The warranty does not cover abuse, batteries or fuses, or damage caused by using the Meter outside its rated specifications or operating environment. The complete warranty stateme
7、nt is in the Meters Instruction Sheet. To contact Fluke, call one of the following telephone numbers: USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65-*-276-6196 Anywhere in the world: +1-425-356-5500 Or, visit Flukes Web site at . To orde
8、r replacement parts, call 1-800-526-4731. Outside U.S.A. contact your nearest service center. Service, repairs, and replacement parts are warranted for 90 days. Product Specifications Measurement Accuracy The accuracy of the Meter is specified for one year after calibration, at 18C to 28C (64F to 82
9、F), with relative humidity to 90%. The Meter is true rms sensing and calibrated to the rms value of a sine wave input. General Specifications General specifications are shown in Table 1. Table 1. General Specifications Maximum voltage and frequency between any terminal and earth ground 600V, 400 Hz
10、Operating Temperature10C to +50C (14F to 122F) Operating Altitude2,000 meters (6,562 ft) Storage Altitude10,000 meters (32,808 ft) Storage Temperature20C to +60C indefinite (4F to 140F) Temperature Coefficient0.05 x (specified accuracy)/C (28C; 82F) Electromagnetic CompatibilityIf RF field 1.6 V/m,
11、total accuracy = specified accuracy; If RF field = 3 V/m, total accuracy = specified accuracy + 3.6% of range Operating HumidityTo 30C (85F), 95% relative humidity; To 40C (104F), 75% relative humidity; To 50C (122F), 45% relative humidity Battery Type9V, NEDA 1604A or IEC 6LR61 Battery LifeAlkaline
12、, 60 hours continuous Jaw SizeOne 3 cm (1.2 inch) conductor, or two 2.46 cm (0.97 inch) conductors Size (HxWxL)4.5 cm x 9.5 cm x 25 cm (1.77 x 3.75 x 9.8 inches) Safety UL 3111, , TV, Complies with IEC 1010-1 for use in 600V, CAT III locations (industrial) RadioFans.CN 收音机爱 好者资料库 Required Equipment
13、3 Table 1. General Specifications (cont) Case ProtectionIP 30 per IEC 529 MAX Hold FeatureAccuracy: add 10 digits to ac and dc specification; Acquisition time: 250 ms (ac); 10 ms (dc) ContinuityBeeper on at 30; Beeper off at 70 Rms Settling TimeThe rms converter has a longer settling time for displa
14、ys reading 100 counts. The display may read up to 5 digits with no voltage or current inputs. True Rms SensingTrue rms sensing provides more accurate readings on circuits containing harmonics or distorted waveforms. An instrument with true rms reading indicates the true heat-providing current, which
15、 may cause overheated conductors, connections, breakers, or transformers. Functional Specifications Functional specifications are shown in Table 2. Table 2. Functional Specifications FunctionRangeAccuracy* N 1000A 1000A 200A 0-600A, 1.9% Rdg + 4 dgts 600-1000A, 3% Rdg + 3 dgts 0-200A, 1.9% Rdg + 7 d
16、gts F600V 200V 1.0% Rdg + 3 dgts O (45-400 Hz) 600A 200A 10-600A, 2% Rdg + 3 dgts 0.5-200A, 2% Rdg + 5 dgts G (45-400 Hz) 600V 200V 1.5% Rdg + 3 dgts J2001.2% Rdg + 5 dgts Crest Factor Overload Protectionm N1000A 1000A 200A NA 1000A F600V 200V NA1200V rms continuous O (45-400 Hz) 600A 200A 2 CF 3, a
17、dd 2% Rdg 600A continuous (CF 3) G (45-400 Hz) 600V 200V 3 (peak voltage 933V)1200V rms J200NA660V rms continuous * See electromagnetic compatibility specification under General Specifications. RadioFans.CN 收音机爱 好者资料库 36 AC/DC True Rms Clamp Meter 4 Required Equipment Equipment required for the perf
18、ormance test and calibration procedure is shown in Table 3. Table 3. Required Equipment EquipmentRecommended Model AC CalibratorFluke 5700A or 5500A Digital Multimeter (DMM)Any Fluke model Transconductance AmplifierFluke 5220A Resistor, MF, 31.6, 1%Fluke P/N 296673 Resistor, MF, 34.8, 1%Fluke P/N 34
19、3897 Resistor, MF, 61.9, 1%Fluke P/N 305961 50-Turn Current CoilFluke 5500A/Coil Performance Test PWarning To avoid shock, do not execute the performance tests unless the Meter is fully assembled. The performance test verifies that the Meter is working properly and performs to specification. To test
20、 each of the Meters functions and operating ranges, do the following: 1. Connect the source to the Meters V and COM input jacks. 2. Referring to Table 4, put the Meter in the function and range for Function Test number 1 (i.e., the 200V dc function). 3. Apply the first indicated output (i.e., 20.0V
21、dc) from the source. 4. The reading on the Meter display should be within the low and high values shown (i.e., 19.5 and 20.5, respectively). 5. Repeat steps 1-4 for each function and range in Table 4. (Tests 9-15 require the 50-Turn Current Coil; tests 11-15 require the 50-Turn Current Coil and Tran
22、sconductance Amplifier.) If the Meter fails to perform within the low-high range indicated for each test in Table 4, the Meter needs to be calibrated and adjusted, or requires some repair. Table 4. Performance Test Values Function Test Meter RangeSourceOutputMeasuredLowHigh 1. DC Volts200V5700A or 5
23、500A20.0V dc20.0V19.5V20.5V 100.0V dc100.0V98.7V101.3V 200.0V dc200.0V202.3V197.7V 2. AC Volts200V5700A or 5500A200V at 50, 60, apply 200V dc from the Calibrator. If the Meter display does not read between 199.5 and 200.5, adjust R39 until it does. 6. Set the output of the Calibrator to zero; then d
24、isconnect the Calibrator from the Meter. 7. Put the case top back on the case bottom, making sure that the rotary switch shaft is correctly seated in the rotary switch knob. 8. Turn the Meter off. 9. Proceed to calibrate the AC Voltage function. Calibration and Adjustment 7 R46 AC/DC Rotary Switch R