1、TB 9-6625-2271-50 CHANGE 1 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AUDIO ANALYZER BOONTON, MODEL 1120-S/10 and 1121 Headquarters, Department of the Army, Washington, DC 8 July 2004 Distribution Statement A: Approved for public release; distribution is unlimited. TB 9-6625
2、-2271-50, 18 July 2003, is changed as follows: 1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page. Remove Pages Insert Pages 1 and 2 1 and 2 3 and 4 3 and 4 9 and 10 9 and 10 15 and 16 15 and 16 2. File this c
3、hange sheet in front of the publication for reference purposes. Distribution: To be distributed in accordance with Std. IDS No. RLC-1500, 2 January 2003, requirements for calibration procedure TB 9-6625-2271-50. By Order of the Secretary of the Army: PETER J. SCHOOMAKER General, United States Army C
4、hief of Staff Official: JOEL B. HUDSON Administrative Assistant to the Secretary of the Army 0413104 RadioFans.CN 收音机爱 好者资料库 _ This bulletin supercedes TB 9-6625-2271-50, dated 21 August 2000. *TB 9-6625-2271-50 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AUDIO ANALYZER BOONT
5、ON, MODEL 1120-S/10 and 1121 Headquarters, Department of the Army, Washington, DC 18 July 2003 Approved for public release; distribution is unlimited REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS You can improve this manual. If you find any mistakes or if you know of a way to improve these proce
6、dures, please let us know. Mail your letter or DA Form 2028 (Recommended Changes to Publications and Blank Forms) directly to: Commander, U.S. Army Aviation and Missile Command, ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A reply will be furnished to you. You may also provide DA Form 202
7、8 information to AMCOM via e-mail, fax, or the World Wide Web. Our fax number is DSN 788-6546 or Commercial 256-842-6546. Our e-mail address is: 2028redstone.army.mil. Instructions for sending an electronic 2028 may be found at the back of this manual. For the World Wide Web, use https:/amcom2028.re
8、dstone.army.mil. Paragraph Page SECTION I. IDENTIFICATION AND DESCRIPTION Test instrument identification. 1 2 Forms, records, and reports . 2 2 Calibration description. 3 2 II. EQUIPMENT REQUIREMENTS Equipment required. 4 4 Accessories required. 5 4 III. CALIBRATION PROCESS Preliminary instructions.
9、 6 5 Equipment setup and internal calibration. 7 5 Dc measurement accuracy . 8 7 Ac measurement accuracy. 9 8 Frequency measurement. 10 10 Filters frequency accuracy . 11 11 Distortion measurement accuracy. 12 12 Output frequency accuracy . 13 14 Output level. 14 16 Output distortion. 15 19 SINAD me
10、asurement accuracy . 16 20 Final procedure. 17 22 RadioFans.CN 收音机爱 好者资料库 TB 9-6625-2271-50 CHANGE 1 2 SECTION I IDENTIFICATION AND DESCRIPTION 1. Test Instrument Identification. This bulletin provides instructions for the calibration of Audio Analyzer, Boonton, Model 1120-S/10 and 1121. The manufac
11、turers manuals were used as the prime data source in compiling these instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin. a. Model Variations. Variations among models are described within the text. b. Time and Technique. The time requ
12、ired for this calibration is approximately 6 hours, using the dc and low frequency technique. 2. Forms, Records, and Reports a. Forms, records, and reports required for calibration personnel at all levels are prescribed by TB 750-25. b. Adjustments to be reported are designated (R) at the end of a s
13、entence in which they appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those adjustments made and designated with (R). 3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1. Table
14、 1. Calibration Description Test instrument parameters Performance specifications Dc level measurement Range: 3, 30, 300 V FS Accuracy: +1% or 3 counts whichever is greater Ac level measurement Range: 3, 30, 300 mV and 3, 30, 300 V Range: 50 Hz to 50 kHz (1 mV to 300 V) Accuracy: +1% Range: 20 to 50
15、 Hz and 50 to 100 kHz (1 mV to 300 V) Accuracy: +2% Range: 10 to 20 Hz1 (1 mV to 300 V) Accuracy: +3% Range: 10 Hz to 100 kHz1 (.3 mV to 300 V) Accuracy: +4 Flatness Range: 1 mV to 300V Range: 50 Hz to 50 kHz (1 mV to 300 V) Accuracy: +0.5% Range: 20 to 50 Hz and 50 kHz to 100 kHz (1 mV to 300 V) Ac
16、curacy: +1.0% Range: 10 to 20 Hz1 (1 mV to 300 V) Accuracy: +2.0% See footnotes at end of table. RadioFans.CN 收音机爱 好者资料库 TB 9-6625-2271-50 CHANGE 1 3 Table 1. Calibration Description Continued Frequency counter Range: 20 Hz to 500 kHz Sensitivity: 3 mV Accuracy: +1 ppm/yr +1 count Audio filters Rang
17、e: 30 kHz low-pass Accuracy: +2 kHz Range: 80 kHz low-pass Accuracy: +4 kHz Range: 220 kHz low-pass Accuracy: +20 kHz Range: 400 Hz high-pass Accuracy: +40 Hz Distortion measurement Range: 10 Hz to 20 kHz Accuracy: +1 dB Range: 20 to 100 kHz Accuracy: +2 dB Input voltage range: 10 mV to 300 V SINAD
18、measurement Frequency range: 10 Hz to 100 kHz Input voltage range: 10 mV to 300 V Display range: -120.00 to 0.00 dB2 Accuracy: +1 dB, 20 Hz to 20 kHz +2 dB, 10 Hz to 100 kHz Output frequency Range: 20 Hz to 140 kHz Accuracy: +10 ppm +1 count +10 ppm +Time base error +1 count( Boonton 1121) Output le
19、vel Range: 10 Hz to 50 kHz, 0.6 mV to 16 V Accuracy: +0.5% of setting +0.05% of range Range: 50 to 100 kHz, 0.6 mV to 16 V Accuracy: +1.0% of setting +0.05% of range Range: 100 to 140 kHz, 0.6 mV to 16 V Accuracy: +1.5% of setting +0.1% of range Flatness Ref 1 kHz (.30 mV to 8 V into 50) Accuracy: +
20、0.5%; 10 Hz1 to 50 kHz Output distortion Range: 10 Hz to 20 kHz, 80 kHz bandwidth Accuracy: 0.01% (-80 dB) Range: 20 to 50 kHz, 220 kHz bandwidth Accuracy: 0.02% (-74 dB) Range: 50 to 100 kHz, 500 kHz bandwidth Accuracy: 0.056% (-65 dB) Range: 100 to 140 kHz, 500 kHz bandwidth Accuracy: 0.1% (-60 dB
21、) 1Not checked below 20 Hz. 2Not verified below -30 dB. TB 9-6625-2271-50 4 SECTION II EQUIPMENT REQUIREMENTS 4. Equipment Required. Table 2 identifies the specific equipment to be used in this calibration procedure. This equipment is issued with Secondary Reference Calibration Standards Set NSN 493
22、1-00-621-7878. Alternate items may be used by the calibrating activity. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed
23、in table 2 provide a four-to-one ratio between the standard and TI. 5. Accessories Required. The accessories required for the calibration are common usage accessories, issued as indicated in paragraph 4 above and are not listed in this calibration procedure. Table 2. Minimum Specifications of Equipm