1、 Fluke 123/124/125 Industrial ScopeMeter Service Manual 4822 872 05398 March 2007 2007 Fluke Corporation, All rights reserved. All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 Limited Warranty adjust the trigger level and verify that the signal will be triggere
2、d now. The trigger level is indicated by the trigger icon (). 6. Set the 5500A to source a 25 MHz (Fluke 123) or 40 MHz (Fluke 124/125) leveled sine wave of 400 mV peak-to-peak. 7. Adjust the amplitude of the sine wave 1.5 divisions on the test tool display. 8. Verify that the signal is well trigger
3、ed. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level and verify that the signal will be triggered now. 9. Set the 5500A to source a 40 MHz (Fluke 123) or 60 MHz (Fluke 124/125) leveled sine wave of 1.8V peak-to-peak. 10. Adjust the amplitude
4、 of the sine wave to exactly 4 divisions on the test tool display. 11. Verify that the signal is well triggered. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level and verify that the signal will be triggered now. Performance Verification 3.5
5、Input A and Input B Tests 3 3-13 12. When you are finished, set the 5500A to Standby. 3.5.8 Input A and B Trigger Level and Trigger Slope Test Proceed as follows: 1. Connect the test tool to the 5500A as shown in Figure 3-5. ST8001.WMF Figure 3-5. Test Tool Input A-B to 5500A Normal Output 2. Select
6、 the following test tool setup: Turn Input B on ( if not already on). Using change the sensitivity to select manual sensitivity ranging, and lock the Input A and Input B sensitivity on 1V/div. Move the Input A and Input B ground level (indicated by zero icon ) to the center grid line. Proceed as fol
7、lows: Press to enable the arrow keys for moving the Input A ground level. Press to enable the arrow keys for moving the Input B ground level. Using the keys move the ground level. Using change the time base to select manual time base ranging, and lock the time base on 10 ms/div. Fluke 123/124: press
8、 / to open the SCOPE INPUTS menu. Fluke 123/124: press to open the TRIGGER menu, and choose: INPUT: ? A | SCREEN UPDATE: ? FREE RUN | AUTO RANGE: ? 15Hz Fluke 125: press to open MENU. Fluke 125: press to open the TRIGGER menu, and choose : INPUT: ? A | UPDATE: ? FREE RUN | AUTO RANGE: ? 15Hz Press t
9、o enable the arrow keys for Trigger Level and Slope adjustment. Fluke 123/124/125 Service Manual 3-14 Using select positive slope triggering (trigger icon ). Using set the trigger level to +2 divisions from the screen center. For positive slope triggering, the trigger level is the top of the trigger
10、 icon (). Fluke 123/124: press / to open the SCOPE INPUTS menu. Fluke 123/124: press to open the SCOPE OPTIONS menu, and choose : SCOPE MODE: ? SINGLE SHOT | WAVEFORM MODE: ? NORMAL Fluke 125: press to open MENU. Fluke 125: press to open the TRIGGER menu, and choose : UPDATE: ? SINGLE Fluke 125: pre
11、ss to open the SMOOTH menu, and choose : WAVEFORM: ? NORMAL 3. Set the 5500A to source 0.4V DC and to OPR. 4. Verify that no trace is shown on the test tool display, and that the status line at the display bottom shows Wait:A . If the display shows the traces and status Hold:A , then press to re-arm
12、 the test tool for a trigger. 5. Increase the 5500A voltage slowly in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 6. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at
13、 step 3. 7. Set the 5500A to Standby. 8. Press to clear the display. 9. Press to enable the arrow keys for Trigger Level and Slope adjustment. 10. Using select negative slope triggering ( ). 11. Using set the trigger level to +2 divisions from the screen center. For negative slope triggering, the tr
14、igger level is the bottom of the trigger icon (). 12. Set the 5500A to source +3V DC and to OPR. 13. Verify that no trace is shown on the test tool display, and that the status line at the display bottom shows Wait:A. If the display shows the traces and status Hold:A, then press to re-arm the test t
15、ool for a trigger. 14. Decrease the 5500A voltage slowly in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 15. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at step 12.
16、 16. Set the 5500A to Standby. 17. Press to clear the display. Performance Verification 3.5 Input A and Input B Tests 3 3-15 18. Select the following test tool setup: Fluke 123/124: press / to open the SCOPE INPUTS menu. Fluke 123/124: press to open the TRIGGER menu, and choose: INPUT: ? A | SCREEN
17、UPDATE: ? FREE RUN | AUTO RANGE: ? 15Hz Fluke 125: press to open MENU. Fluke 125: press to open the TRIGGER menu, and choose : INPUT: ? A | UPDATE: ? FREE RUN | AUTO RANGE: ? 15Hz Press to enable the arrow keys for Trigger Level and Slope adjustment. Using select positive slope triggering (trigger i
18、con ). Using set the trigger level to +2 divisions from the screen center. For positive slope triggering, the trigger level is the top of the trigger icon (). 19. Set the 5500A to source 0.4V DC and to OPR. 20. Verify that no trace is shown on the test tool display, and that the status line at the d
19、isplay bottom shows Wait:B . If the display shows the traces and status Hold:B , then press to re-arm the test tool for a trigger. 21. Increase the 5500A voltage slowly in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 22. Verify that the
20、 5500A voltage is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at step 19. 23. Set the 5500A to Standby. 24. Press to clear the display. 25. Press to enable the arrow keys for Trigger Level and Slope adjustment. 26. Using select negative slope triggering ( ). 27
21、. Using set the trigger level to +2 divisions from the screen center. For negative slope triggering, the trigger level is the bottom of the trigger icon (). 28. Set the 5500A to source +3V DC and to OPR. 29. Verify that no trace is shown on the test tool display, and that the status line at the disp
22、lay bottom shows Wait:B. If the display shows the traces and status Hold:B, then press to re-arm the test tool for a trigger. 30. Decrease the 5500A voltage in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 31. Verify that the 5500A volta
23、ge is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at step 28. 32. When you are finished, set the 5500A to Standby. Fluke 123/124/125 Service Manual 3-16 3.5.9 Input A and B DC Voltage Accuracy Test WARNING Dangerous voltages will be present on the calibration s
24、ource and connecting cables during the following steps. Ensure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Proceed as follows: 1. Connect the test tool to the 5500A as for the previous test (see Figure 3-5). 2. Select the following te
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- FLUKE 123_125_smeng0000 电路图 123 _125_smeng0000
收音机爱好者资料库所有资源均是用户自行上传分享,仅供网友学习交流,未经上传用户书面授权,请勿作他用。



Marshall-9100-62-02-Schematic电路原理图.pdf