1、 Function and Arbitrary/Function Generators Model 4084AWG, 4086AWG, 4084, 4085, 4086, 4087 INSTRUCTION MANUAL 4080 Series RadioFans.CN 收音机爱 好者资料库 4080 series Arbitrary/Function Generator with Counter Instruction Manual 2 Safety Summary The following safety precautions apply to both operating and mai
2、ntenance personnel and must be observed during all phases of operation, service, and repair of this instrument. Before applying power, follow the installation instructions and become familiar with the operating instructions for this instrument. Failure to comply with these precautions or with specif
3、ic warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. B When the start frequency is higher than stop frequency, the frequency sweep decreases gradually from the start frequency (high frequency) to stop frequency (low frequency). 40
4、80 series Arbitrary/Function Generator with Counter Instruction Manual 25 The frequency range from start frequency to stop frequency is 1Hz Fmax in linear sweep mode and 1mHz Fmax in log sweep mode. (Fmax see specification section for main waveform) Sweep time “TIME” The time needed to complete one
5、sweep from start to stop frequency is called sweep time. The sweep time range is 1ms 800s in linear sweep mode or 100ms 800s in LOG sweep mode. Trigger mode “TRIG”: A sweep can be triggered internally or externally. The corresponding display values are “1: INT” and “2: EXT”. The default value is INT
6、ernal trigger. In internal trigger mode, the instrument will continuously sweeps from the start frequency to the stop frequency according to the Sweep parameters. An external trigger signal can be generated in one of two ways. a) Press the 【output】 key to trigger a single sweep. The signal frequency
7、 will in/decrease from start frequency to stop frequency, then sweep will stop. b) Apply an input trigger signal to the “EXT Trig” terminal on the rear panel. A rising edge will trigger a single sweep. In external trigger mode, symbols “Trig” and “Ext” are displayed in the status field. Start and st
8、op of sweep The sweep will start automatically once sweep mode is selected. If you dont want to output the sweep signal, simply press the 【output】key to disable the signal output (LED above Output button is OFF). Once all the parameters are set, you can make the sweep signal available at the output
9、by pressing the Output button again. In external trigger mode, the 【Output】key functions only as a single pulse trigger key, the Output On/Off functionality is now disabled and the output LED will always be lit. Sweep configuration example Sweep parameters are as followed: Frequency range 100Hz200kH
10、z, sweep time 10s, linear sweep, internal trigger mode. Set the carrier frequency parameters Press【sweep】 Press【menu】 Select “MODE”, wait, then enter【1】 【N】(linear sweep) Press【menu】to select “START F”, press【1】 【0】 【0】 【Hz】 Press【menu】to select “STOP F”, press【2】 【0】 【0】 【kHz】 Press【menu】to select
11、“TIME, press【1】 【0】 【s】 Press【menu】 to select “TRIG”, press【1】 【N】 Hint: The frequency displayed immediately after pressing Sweep is the start frequency. Any changes to that value will automatically update parameter “START F” 4080 series Arbitrary/Function Generator with Counter Instruction Manual 2
12、6 FM DEVIA FM FREQ FM WAVE FM SOURCE 4.6.2 FM modulation A modulated waveform consists of a carrier waveform and a modulating waveform. In FM, the frequency of the carrier is varied by the amplitude of the modulating waveform. The function generator will accept an internal or external FM modulating
13、signal. Press the 【FM】key to enter into FM mode. The display will indicate the carrier frequency and the “FM” annunciator will be lit. The frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in the previous section. Only the main functions sine or square wave c
14、an be selected as a carrier signal. Press the 【menu】 key repeatedly to cycle through the FM modulation parameters listed below. FM DEVIA Peak frequency deviation FM FREQ: Frequency of modulating signal FM WAVE: Waveform of modulating signal FM SOURCE Toggle between internal or external modulating si
15、gnal Each time you press the menu key, the parameter will flash for 1 second, followed by the value of that parameter. FM modulation parameters can be entered directly or via the knob. Once the parameter is set, press 【menu】 to advance to the next parameter. Peak frequency deviation ”FM DEVIA” Expla
16、nation: The variation in frequency of the modulating waveform from the carrier frequency (center frequency). Range of deviation: For internal FM modulation, the maximum value should not exceed 50% of the carrier frequency. In external FM mode, the maximum deviation is 10% of the carrier frequency. A
17、dditionally, the frequency deviation plus carrier frequency should not exceed the maximum operating frequency of the instrument. (Fc + Fd AM FREQ AM WAVEAM SOURCE modulating signal source “external” is selected, the symbol “Ext” is displayed and parameters FM DEVIA, FM FREQ and FM WAVE do not apply
18、(disabled) Start and stop of FM modulation An FM signal is generated once FM function mode is selected. The instrument will automatically output a signal according to the preset parameters. If you do not want to output the FM signal, simply press the 【output】 key to disable the signal output (LED ab
19、ove Output button is OFF). Once all the parameters are set, you can turn make the FM signal available at the output by pressing the Output button again. FM example: Example configuration: Carrier signal is square, frequency is 1MHz, amplitude is 2V, modulating signal is generated internally. Carrier
20、 waveform is sine (No. 1), Frequency is 5kHz. Peak frequency deviation is 200kHz. Press【FM】 Press【frequency】then【1】 【MHz】 (set carrier frequency) Press【amplitude】, then【2】 【V】 (set carrier amplitude) Press【Shift】and 【square】 (set carrier waveform) Press【menu】, select “FM DEVIA”, enter【2】 【0】 【0】 【kH
21、z】 (set FM deviation) Press【menu】, select “FM FREQ”, press【5】 【kHz】 (set FM frequency) Press【menu】, select “FM WAVE, press【1】 【N】 (set FM waveform as sine) Press【menu】, select “FM SOURCE”, press【1】 【N】 (set FM source as internal) 4.6.3 AM modulation AM stands for amplitude modulation. Press 【AM】 to
22、enable AM modulation. The carrier frequency and the AM annunciator is displayed. Frequency, amplitude, waveform and DC offset of the carrier signal can be set as described in the previous section of this chapter. The parameters carry over from the parameter settings of the main waveforms sine and sq
23、uare. In AM only sine and square waves can be selected for carrier. Press the 【menu】 key repeatedly to cycle through the AM modulation parameters listed below. AM LEVEL: Modulation depth AM FREQ: Frequency of modulating signal AM WAVE: Waveform of modulating signal AM SOURCE: internal or external mo
24、dulating signal In AM mode, to ensure normal signal output at 100% modulation depth, the instrument reduces the peak-to-peak value of the carrier by 50%. Only sine and square waves can be selected as AM carrier. 4080 series Arbitrary/Function Generator with Counter Instruction Manual 28 Modulation d