1、 51, 52, 53, 54 Series II Thermometer Service Manual PN 1276123 October 1999 1999 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 LIMITED WARRANTY a Caution statement identifies conditions and actions that
2、 may damage the thermometer or the test instruments. Thermometer Precautions and Safety Information 3 Safety Information WWarning To avoid electrical shock or personal injury, follow these guidelines: Before using the thermometer, inspect the case. Do not use the thermometer if it appears damaged. L
3、ook for cracks or missing plastic. Pay particular attention to the insulation around the connectors. Disconnect the thermocouple(s) from the thermometer before opening the case. Replace the batteries as soon as the low battery indicator (B) appears. The possibility of false readings can lead to pers
4、onal injury. Do not use the thermometer if it operates abnormally. Protection may be impaired. When in doubt, have the thermometer serviced. Do not operate the thermometer around explosive gas, vapor, or dust. Do not apply more than the rated voltage, as marked on the thermometer, between the thermo
5、couple(s), or between any thermocouple and earth ground. Models 52 and 54: Measurement errors may occur if voltages on the measurement surfaces result in potentials greater than 1 V between the two thermocouples. When potential differences are anticipated between the thermocouples, use electrically
6、insulated thermocouples. When servicing the thermometer, use only specified replacement parts. Do not use the thermometer with any part of the case or cover removed. Caution To avoid damaging the thermometer or the equipment under test: Use the proper thermocouples, function, and range for your ther
7、mometer. Do not attempt to recharge the batteries. To prevent explosion, do not dispose of batteries in fire. Follow local laws or regulations when disposing of batteries. Match the + and polarities of the batteries with the battery case markings. 51, 52, 53, 54 Series II Service Manual 4 Specificat
8、ions General Specifications Weight280 g (10 oz) Dimensions (without holster) 2.8 cm 7.8 cm 16.2 cm (1.1 in 3 in 6.4 in) Battery3 AA batteries CertificationP, ) (TV pending) SafetyCSA C22.2 No. 1010.1 1992 EN 61010 Amendments 1, 2 CAT IOVERVOLTAGE (Installation) CATEGORY I, Pollution Degree 2 per IEC
9、1010-1* * Refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level. Environmental Operating Temperature10 oC to +50 oC
10、 (+14 oF to +122 oF) Storage Temperature40 oC to +60 oC (40 oF to +140 oF) Humidity Non Condensing ( +10 oC) ( +50 oF) 95% RH (+10 oC to +30 oC) (+50 oF to +86 oF) 75% RH (+30 oC to +40 oC) (+86 oF to +104 oF) 45% RH (+40 oC to +50 oC) (+104 oF to +122 oF) (Without Condensation) Electrical Measureme
11、nt Range J-type: 210 oC to +1200 oC (346 oF to +2192 oF) K-type: 200 oC to +1372 oC (328 oF to +2501 oF) T-type: 250 oC to +400 oC (418 oF to +752 oF) E-type: 150 oC to +1000 oC (238 oF to +1832 oF) N-type: 200 oC to +1300 oC (328 oF to +2372 oF) R- and S-type: 0 oC to +1767 oC (+32 oF to +3212 oF)
12、Display Resolution 0.1 oC / oF / K 1000o 1.0 oC / oF / K 1000o Measurement Accuracy J-, K-, T-, E-, and N-type: 0.05 % of reading + 0.3 oC (0.5 oF) below 100 oC (148 oF): add 0.15 % of reading for J, K, E, and N; and 0.45 % of reading for T R- and S-type: 0.05 % of reading + 0.4 oC (+0.7 oF) Tempera
13、ture Coefficient 0.01 % of reading + 0.03 oC per oC (0.05 oF per oF) outside the specified +18 oC to +28 oC (+64 oF to +82 oF) range below 100 oC (148 oF): add 0.04 % of reading for J-, K-, E-, and N-type; and 0.08 % of reading for T-type Electromagnetic Compatibility Susceptibility: 2 oC ( 3.6 oF)
14、for 80 MHz to 200 MHz in 1.5 v/m field, for 200 MHz to 1000 MHz in 3 V/m field. Emissions: Commercial limits per EN50081-1 Maximum Differential Common Mode Voltage1 V (Maximum voltage difference between T1 and T2) Temperature ScaleITS-90 Applicable StandardsNIST-175 Accuracy is specified for ambient
15、 temperatures between +18 oC (+64 oF) and +28 oC (+82 oF) for a period of 1 year. The above specifications do not include thermocouple error. Thermometer Performance Test and Calibration Required Test Equipment 5 Performance Test and Calibration Required Test Equipment Refer to Table 2 for a list of
16、 test equipment used in the performance test and calibration procedures. Table 2. Recommended Test Equipment EquipmentRequired CharacteristicsRecommended Model DC Voltage Calibrator Output Voltage: 0-1V Accuracy: 0.0125 % Resolution: 0.1 VFluke 5700A or equivalent Thermocouple connector For K-type t
17、hermocouple Fluke 80CK-M Fluke P/N 779942 Thermocouple connectorCopperFluke P/N 601747 K-type Thermocouple wireStandardOmega Dewar Flask/Cap1- pint, or larger capacityThermos Mercury Thermometer (either C or F) 0.02 C Resolution 0.05 F Resolution Princo Model ASTM56C Princo Model ASTM56F DB9 Adapter
18、 ConnectorFluke P/N 804549 IR AdapterFluke P/N 690518 PC with Microsoft Windows installed Performance Tests Performance tests are recommended for incoming inspection, periodic maintenance, and for checking the thermometers specifications. If the thermometer fails any part of the performance tests, c
19、alibration and/or repair is indicated. Before performing these tests: 1. Allow the thermometer to stabilize to room temperature, +23 oC 5 oC (+73 oF 9 oF). 2. Check the batteries and replace them if necessary. Display Test Turn the thermometer on while pressing the h button. All display segments are
20、 displayed until the h button is released. Confirm that all the display segments are showing as in Figure 1. 51, 52, 53, 54 Series II Service Manual 6 52 51 54 53 zb03f.eps Figure 1. Display Segments Keypad Test Check that all keypad switches are functional by pressing each key and confirming the pr
21、oper response on the display. Wide Range Performance Test Procedure This test verifies the instrument specifications. If the instrument fails this test, perform the calibration procedure. If the unit continues to fail, it should be taken to a Fluke service center. Ice Bath Construction To begin the
22、performance test, first construct an ice bath as follows: 1. Prepare a Dewar Flask by drilling two holes in its cap to accept the Mercury thermometer and thermocouple wires, or use a standard laboratory cork. 2. Fill the Dewar Flask with shaved or crushed ice made from distilled water. 3. Fill the D
23、ewar Flask with enough distilled water so that the ice turns to slush, but do not add enough water to float the ice. Note As the ice melts, siphon off the excess water and add more ice. Allow approximately 5 to 10 minutes for the water to drop back to the freezing point. Thermometer Performance Test
24、s 7 4. Replace the Dewar Flask cap or cork, and insert the thermocouple wires and Mercury thermometer as shown in Figure 2. Insert the Mercury thermometer to the same depth as the thermocouple wire. 5700A Calibrator Ice bath K-Type Thermocouple wire Copper wire 51, 52, 53, or 54 Series II Thermomete