1、M S1-1052 10kV Digital Insulation Tester USER MANUAL RadioFans.CN 收音机爱 好者资料库 GSAFETY WARNINGS Safety Warning must be observed during use. IThe circuit under test must be switched off, de-energised, isolated and checked to be safe before insulation test connections are made. Make sure the circuit is
2、not re-energised whilst the instrument is connected. ICircuit connections must not be touched during an insulation test. IAfter completing a test, capacitive circuits must be completely discharged before disconnecting the test leads. Capacitive charges can be lethal. ITested items should be firmly s
3、horted out with a shorting link, after discharge, until required for use. This is to guard against any stored dielectric absorption charge subsequently being released thereby raising the voltage to potentially dangerous levels. IThe voltage indicator and automatic discharge features should be regard
4、ed as additional safety features and not a substitute for normal safe working practice. IIt is rare, but in certain circumstances, breakdown of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner, possibly causing a loss of display while the circuit remain
5、s energised. In this event, the unit must be turned off and the circuit discharged manually. ITest leads, including crocodile clips, must be in good order, clean and with no broken or cracked insulation. IThe instrument should not be used if any part of it is damaged. IWater on the surface of the in
6、strument may lead to erroneous results. It is recommended that the instrument is not used in wet weather conditions. IThis instrument is not intrinsically safe and must not be used in hazardous atmospheres. 2 NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. RadioFans.
7、CN 收音机爱 好者资料库 Breakdown mode / burn mode15 Alarm limit mode15 Step voltage SV test16 Polarisation index PI test, and dielectric absorption ratio DAR test16 Dielectric discharge DD test16 Measurements above 100 G17 Circuit block diagram18 Specifications19 Accessories20 Repair and warranty21 Symbols u
8、sed on the instrument are: FCaution: risk of electric shock GCaution: refer to accompanying notes tEquipment protected throughout by Double Insulation (Class II) cEquipment complies with current EU directives. Terms used in this manual The word must is used to indicate that the instructions followin
9、g should be followed under all circumstances. Failure to follow these instructions could result in damage to the instrument and / or a hazard to the operator. The word should is used to indicate that the instructions indicate best practice. CONTENTS 3 Safety warnings2 Introduction4 General descripti
10、on4 Features4 Cleaning4 Power lead and battery charging5 Instrument controls and indicators6 Power On/Off button7 Test voltage L and M buttons7 Test start / stop button7 /I button7 Bbutton10 Fn button10 Mode button10 Timer L and M buttons11 Record button11 Recording to on-board memory11 Downloading
11、results11 Deleting results11 Recording to a PC12 High voltage warning LED12 Line input present LED12 Test terminals12 Guard terminal12 RS232 / USB connections13 Battery bar graph13 Voltage at terminals13 Timer indicator13 Digital display13 Analogue display14 Secondary display14 Pre-Test / during tes
12、t key action table14 Test mode summary15 Insulation resistance IR test15 RadioFans.CN 收音机爱 好者资料库 Features IContinuous resistance or current measurement on main display. IOptional resistance, current, or figure of merit measurements on secondary display. IStandard test voltages 250 V , 500 V , 1 kV ,
13、 2.5 kV , 5 kV , 10 kV INon-standard test voltages selectable between 50 V and 1 kV in 10 V steps, and selectable between 1 kV and 10 kV in 25 V steps. ITest modes include insulation resistance, polarisation index, step voltage and dielectric discharge. IThe dielectric absorption ratio is automatica
14、lly calculated if the corresponding timers are set. IEither burn or breakdown selectable in insulation resistance mode. IInsulation Alarm Limit available in insulation resistance mode. IProgrammable timers include a main test duration timer, plus T1 and T2 timers for time resistance method type test
15、s. ILoad capacitance and time constant measurements displayed at end of test. ILCD backlight. IData storage, data retrieval and real time data output. IUSB or RS232 communications. IBattery level / charge level indicator. Cleaning Disconnect the instrument and wipe it with a clean cloth slightly dam
16、ped with soapy water or Isopropyl alcohol (IPA). INTRODUCTION 4 General Description The S1-1052 is a microprocessor controlled 10 kV insulation tester offering measurement capability of up to 35 T. The instrument performs automatic tests and has data storage and data retrieval facilities. Polarizati
17、on index and dielectric discharge tests are performed automatically, and test duration and voltages can be adjusted according to user preference for these tests. A step voltage test can be performed automatically, with a default voltage of 1 kV , and test duration of 5 minutes. It is possible for th
18、e user to set different voltages and test durations for the step voltage test. The S1-1052 can be powered from the mains supply or by its own internal rechargeable battery, which provides for at least 4 hours of continuous testing. A battery level indicator on the LCD display indicates battery capac
19、ity. Connecting power to the mains supply connector will automatically charge the battery whether the instrument is switched on or off, except during testing. A high level of internal isolation allows the instrument to be used while the unit is supplied from the mains. An internal battery management
20、 system switches the instrument off after ten minutes of inactivity. If the battery approaches a very low state of charge the instrument turns itself off, and mains power must be applied before the instrument can be used again. Recorded test results and settings will not be lost when the instrument
21、is switched off. A comprehensive LCD display shows resistance, current, capacitance, time constant, voltage, timer measurements, and figure of merit measurements such as polarisation index. POWER LEAD AND BATTERY CHARGING 5 If the power lead supplied is not suitable for your mains connection, do not
22、 use an adaptor. Always use a power lead fitted with the correct plug. The instrument is fitted with a two-pin IEC60320 power inlet. Most power leads are made with three-core cable, so the ground connection will not be used. Power lead connection table ConnectionUK/InternationalUSA Earth/Ground Yell
23、ow/Green Green Neutral Blue White Live (Line) Brown Black If using a fused plug, ensure that it is fitted with a 3 Amp fuse. The instrument can be powered from 85 - 265 Vrms a.c. at 50/60 Hz. The battery will charge as long as the mains supply is connected, except when a test is in progress. The pow
24、er On/Off button has a green LED, which illuminates when mains power is present. For optimum battery life, charge the battery after each use. A completely exhausted battery will take 14 hours to recharge. INSTRUMENT CONTROLS AND INDICATORS 6 Measurement terminals -veguard+ve RS232USBLine input Time