1、THIS COPY IS A REPRINT, WHICH INCLUDES CURRENT PAGES FROM CHANGES 1 THROUGH 5 _ *This bulletin supersedes TB 9-6625-994-50, 28 July 1969 and TB ORD 1026/981, 12 October 1964 including all changes. *TB 9-6625-1866-50 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR DIFFERENTIAL VOL
2、TMETERS AN/USM-98, AN/USM-98A, AND AN/USM-98B (JOHN FLUKE MODELS 801, 801R, 801B, 801BR, 801B/AG) Headquarters, Department of the Army, Washington, DC 19 September 1970 REPORTING OF ERRORS You can help improve this publication by calling attention to errors and by recommending improvements and stati
3、ng your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be mailed directly to Commander, U.S. Army Aviation and Missile Command, ATTN: AMSAM-TMD-EP, Redstone Arsenal, AL 35898-5000. FAX to DSN 788- 2313 (commercial 256-842-2313). A reply will
4、 be furnished directly to you. ParagraphPage SECTIONI.GENERAL Purpose and scope.12 Reporting of equipment publication improvements.22 Descriptive data .32 General Instructions.43 Differences among models.54 II.CALIBRATION OF DIFFERENTIAL VOLTMETERS AND/USM-98 (J-F 801 and 801R) Equipment and accesso
5、ry required.64 Preliminary procedure.74 Dc VTVM and null range.85 500-volt range and stability.97 50-volt range.109 5-volt range.119 0.5-volt range.1220 RadioFans.CN 收音机爱 好者资料库 TB 9-6625-1866-50 2 ParagraphPage Voltage dividers .1310 Final procedure .1412 III.CALIBRATION OF DIFFERENTIAL VOLTMETERS A
6、N/USM-98A AND AN/USM-98B (J-F 801B, J-F 801 BR, AND J-F 801B(AG) Equipment and accessories required.1512 Preliminary procedure.1612 Dc VTVM and null range.1713 500-volt range and stability.1816 50-volt range.1916 5-volt range.2017 0.5-volt range.2118 Voltage dividers .2219 Final procedure .2320 SPEC
7、IFIC ITEM IDENTIFICATION Identification.2420 Secondary transfer calibration standards set.2520 SECTION I GENERAL 1. Purpose and Scope. This bulletin provides information for the periodic calibration of Differential Voltmeters AN/USM-98, AN/USM-98A, and AN/USM-98B (John Fluke Models 801, 801R, 801B,
8、801BR, and 801B/AG). It is to be used by personnel trained and qualified in the use of calibration equipment. 2. Reporting of Equipment Publication Improvements. The reporting of errors, omissions, and recommendations for improving this publication by the individual user is encouraged. Reports shoul
9、d be submitted on DA Form 2028 (Recommended Changes to Publications) and forwarded direct to: Commander, US Army Missile Materiel Readiness Command, ATTN: DRSMI-MFPA, Redstone Arsenal, AL 35809. 3. Descriptive Data a. Identification NomenclatureDIFFERENTIAL VOLTMETER Military designationAN/USM-98, A
10、N/USM-98A, and AN/USM-98B ManufacturerJohn Fluke Model number801, 801R, 801B, 801BR, and 801B/AG ReferenceManufacturers instruction manual, TM 11-6625-599-12, and TM 11-6625-599-45 RadioFans.CN 收音机爱 好者资料库 TB 9-6625-1866-50 3 b. Specifications AN/USM-98 (J-F 801 AND J-F 801R): Input requirement1100 t
11、o 130 vac, 60 Hz, 60 w Stability of reference supply10.01% for change in line voltage (105 to 130 v); 0.01% per hour after warm-up (adjustable via front panel CAL control). As a differential voltmeter: Input voltage ranges0 to 500 vdc in 4 ranges Accuracy0.05% from 0.1 to 500 vdc (0.1% of input volt
12、age or 0.05 mv whichever is greater below 0.1 vdc. Null voltage ranges0 to 10, 0 to 1, 0 to 0.1 and 0 to 0.01 vdc Dc input resistance1Infinite at null (all ranges). As a vacuum tube voltmeter: Basic input voltage ranges0 to 500 vdc in 4 ranges Accuracy4% of full scale (all ranges). Additional voltag
13、e ranges20 to 0.01, 0 to 0.1, 0 to 1, and 0 to 10 vdc. AN/USM-98A and AN/USM-98B (J-F 801B, 801BR, and 801B/AG): Power input requirements115 or 230 vac, 10%, 50 to 440 Hz, approx. 70w. (801B and B/AG, S/N 123 to 1632, operate only on power-line frequencies of 50 to 60 Hz). Stability of meter zero10.
14、5% of full scale for 10% line charge. As a differential voltmeter: Input voltage ranges0 to 500 vdc in 4 ranges Accuracy0.05% from 0.1 to 500 vdc (0.05% +50v) below 0.1 vdc. Null voltage ranges0 to 10, 0 to 1, 0 to 0.1 and, 0 to 0.01 vdc. Dc input resistanceInfinite at null (all ranges). As a vacuum
15、 tube voltmeter: Basic input voltage ranges0 to 500 vdc in 4 ranges. Accuracy3% of full scale (all ranges). Additional voltage ranges20 to 0.01, 0 to 0.1, 0 to 1, 0 to 10 vdc. c. Calibration Time required3 hours (approx.) TechniqueDc and low frequency - 1 This specification is for information only a
16、nd is not necessarily verified in this bulletin. 2 These range are obtained by setting NULL switch to desired range and readout dials to zero. 4. General Instructions a. DA Form 2416 (Calibration Data Card). During the use of this bulletin, annotate DA Form 2416 in accordance with TM 38-750. b. Unit
17、 Under Test. Differential Voltmeters AN/USM-98 (John Fluke Model 801 and 801R) will be referred to as UUT (unit under test) in section II. Differential voltmeters RadioFans.CN 收音机爱 好者资料库 TB 9-6625-1866-50 4 AN/USM-98A and AN/ USM-98B (John Fluke Models 801B, 801BR, and 80 1B/AG) will be referred to
18、as UUT in section III. c. Equipment and Accessory Identification. The equipment and accessories referred to throughout this bulletin are identified in tables 1, 2, 6 and 7 and section IV. d. Equipment Setup. Disconnect instructions are not contained in this bulletin. 5. Differences Among Models. Dif
19、ferential Voltmeter 801R is a rack-mounted version of the 801. AN/USM-98 is the military designation for the 801. Differential Voltmeter 801B contains a standard cell; the 801B/AG contains a Zener diode. The 801BR is the rack- mounted version of the 801B. AN/USM-98A is the military designation for t
20、he 801B and 801B/AG. The AN/USM-98B is similar to the 801B. Front panel designations may vary among models. SECTION II CALIBRATION OF DIFFERENTIAL VOLTMETERS AN/USM-98 (J-F 801 AND 801R) 6. Equipment and Accessory Required. Table 1 lists minimum use specifications of equipment required for calibrati
21、on performance checks and adjustments. Table 2 lists the required accessory. Tables 1 and 2 are provided to assist in the selection of required equipment and accessory. For specific item identification refer to section IV. Table 1. Minimum Specifications of Equipment Required Item NumberCommon NameM
22、inimum Use Specifications A1AUTOTRANSFORMERRANGE: 105 to 125 vac ACCURACY:1% A2DC VOLTAGE STANDARDRANGE: 0 to 520 v ACCURACY:0.016% Table 2. Accessory Required Item NumberCommon NameDescription B1CABLE30-in., RG-58/U; double banana plug terminations 7. Preliminary Procedure NOTE Personnel should fam
23、iliarize themselves with the entire bulletin prior to performing calibration. a. With no power applied to UUT, mechanically zero meter, using meter mechanical zero adjustment screw. TB 9-6625-1866-50 5 b. Connect UUT to autotransformer (A1, table 1) and adjust for 115 volts ac. c. Turn VOLTS RANGE,