1、 PN 3474006 May 1996, Rev. 1, 5/09 2009 Fluke Corporation. All rights reserved. Printed in USA. All product names are trademarks of their respective companies. 5700A/5720A Series II Multi-Function Calibrator Operator Guide RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 i Table of Contents Title Pag
2、e What is in this Guide?. 1 Safety Summary. 1 WFuse and Line Voltage . 2 Basic Operation. 2 A Short Calibration Exercise. 2 Warming up the Calibrator. 3 Exercise: Running dc Zero Calibration. 3 Connecting a Meter . 3 Applying a dc Voltage. 3 Checking the Calibrator Uncertainty. 4 Exercise: Activatin
3、g Error Mode. 4 Checking the Meters Higher and Lower Ranges. 4 Exercise: Applying an ac Voltage. 4 Front Panel Features . 5 Output Display (Left Side). 5 Control Display (Right Side). 5 Display Screen Saver. 5 Front Panel Keys . 5 Front Panel Connectors . 10 Rear Panel Features . 11 Rear Panel Conne
4、ctors . 11 Switches and Fuse Holder. 12 Cable Recommendations . 13 Connecting to a UUT (Unit Under Test) . 14 Using Error Mode. 19 Operating in Remote. 19 RS-232-C Interface Parameters . 20 RadioFans.CN 收音机爱 好者资料库 5700A/5720A Operator Guide ii Series II Multi-Function Calibrator What is in this Guid
5、e? 1 What is in this Guide? The Operator Guide begins with safety information, a short meter calibration exercise for new users, then continues with a condensation of information from the Operator Manual. For complete information about the Calibrators features, functions, and operating procedures, r
6、efer to the Operator Manual. Safety Summary WXWarning Lethal voltage may be present on the terminals. Observe all safety precautions in this guide. To avoid electrical shock hazard, the operator should not electrically contact the OUTPUT HI or SENSE HI binding posts. During operation, lethal voltage
7、s of up to 1100V ac or dc may be present on these terminals. Whenever possible, keep one hand away from the equipment to reduce the hazard of current flowing through vital organs of the body. Symbols Marked on Equipment X Warning - Risk of electric shock. J J Ground Protective ground (earth) termina
8、l. W Attention Refer to the manual for information about this feature. This symbol appears on the rear panel ground terminal and by the fuse holder. Use the Proper Fuse To avoid fire hazard, use only the fuse specified on the line voltage selection switch label. WX Grounding the Calibrator The Calib
9、rator is a Safety Class I (grounded enclosure) instrument. The enclosure is grounded through the grounding conductor of the power cord. To avoid electrical shock, plug the power cord into a properly wired earth grounded receptacle before making any connections to the Calibrator terminals. A protecti
10、ve ground connection by way of the grounding conductor in the power cord is essential for safe operation. WXDo Not Remove the Cover To avoid personal injury, do not remove the Calibrator cover. Do not operate the Calibrator without the cover properly installed. There are no user-serviceable parts in
11、side the Calibrator, so there is no need to ever remove the cover. 5700A/5720A Operator Guide 2 WWFuse and Line Voltage The correct fuse type for each voltage selection is indicated on the rear panel as shown in the following figure. VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 100V S2S3S4 110V 115V 1
12、20V VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 200V S2S3S4 220V 230V 240V CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING. VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 100V S2S3S4 110V 115V 120V VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 200V S2S3S4 220V 230V 240V CAUTION
13、 FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING. 5720og01.eps Line Power Label and Switch Locations Basic Operation To set the output, simply press the following sequence of keys to select an output function and amplitude: numeric keys multiplier function E O For example, to s
14、et the output to 10 mV dc, press: 1 0 m V E O To set an ac output, press the following additional keys: numeric keys multiplier H E For example, to change the present 10 mV dc to 10 mV ac 1.8 kHz, press: 1 . 8 K H E To change the output back to dc, press: 0 H E or + E A Short Calibration Exercise If
15、 you are not familiar with the Calibrator, you will find that the following exercise quickly teaches you some important front panel operations. Series II Multi-Function Calibrator Exercise: Running dc Zero Calibration 3 Warming up the Calibrator When you first turn the Calibrator on, you should let
16、it warm up for 30 minutes. This ensures that the Calibrator meets or exceeds its specifications. Exercise: Running dc Zero Calibration DC Zero is a brief internal process that removes offset errors. The specifications require that you run dc Zero at least every 30 days. To run dc Zero, press the fol
17、lowing sequence of softkeys: Setup Menus ? Cal ? Zero Press any key, then press P twice to return to normal operation. Connecting a Meter Typical connections for handheld and benchtop meters are shown on the next page. The diagrams show the use of Fluke Model 5440-7002 Low Thermal Cables. Other type
18、s of cable may be used for this exercise. To connect either type of meter, proceed as follows: 1. Verify that the Calibrator is on, and is in standby mode (STANDBY indicator lit). If the Calibrator is not in standby, press r. 2. Set the meter to the lowest range that reads 10V dc. 3. Connect the shi
19、eld lead (if your cables are shielded) to V GUARD on the Calibrator. 4. Verify that both X and x are off. Press them if they are lit. 5. Connect the meter input LO to the Calibrator OUTPUT LO. 6. Connect the meter input HI to the Calibrator OUTPUT HI. HI HI LO LO HI AUX CURRENT GUARD GROUND WIDEBAND
20、 Calibrator 10A 300mA V COM DMM NC EX SNS : OFF EX GRD : OFF HI HI LO LO HI SENSE V AUX CURRENT GUARDGROUND WIDEBAND Calibrator DMM INPUT NC : OFF EX GRD : OFF EX SNS OUTPUT VA SENSE V OUTPUT VA 5720og10.eps Applying a dc Voltage Apply 10V dc to the meter by pressing the following sequence of keys:
21、1 0 V E O 5700A/5720A Operator Guide 4 Checking the Calibrator Uncertainty Press s. The total uncertainty for the 10V you are applying to the meter is displayed on the Control Display (the large dot-matrix display on the right side of the front panel). This number depends on the setting of the calib