FLUKE 5820a 电路图.pdf
《FLUKE 5820a 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 5820a 电路图.pdf(164页珍藏版)》请在收音机爱好者资料库上搜索。
1、 5820A Oscilloscope Calibrator Service Manual PN 673142 June 1999 Rev.2, 11/06 1999-2006 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 LIMITED WARRANTY ANSI/ISA-S82.01-1994; CAN/CSA- C22.2 No. 1010.1-
2、92 Output Impedance The 5820A is designed to drive both 50 e and 1 Me loads Analog Low Isolation 20 V EMC Complies with EN 61326-1 Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V Line Frequency: 47 to 63 Hz Line Power Line Voltage Variation: 10 % about line voltage setting Power Consumption 25
3、0 VA Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6 inch) for feet on bottom of unit. Width: 43.2 cm (17 inches), standard rack width Dimensions Depth: 47.3 cm (18.6 inches) overall Weight 20 kg (44 pounds) Absolute Uncertainty Definition The 5820A specifications include stabi
4、lity, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total specification of the 5820A for the temperature range indicated. Specification Confidence Interval 99 % 1 Aft
5、er long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be required. 5820A Service Manual 1-8 1-8. Voltage Output Specifications Volt Function DC Signal Square Wave Signal 1 Load into 50 e e into 1 Me e into 50 e e into 1 Me e Amplitude Character
6、istics 0 V to 6.6 V 0 V to 130 V 1 mV to 6.6 V p-p 1 mV to 130 V p-p Range Resolution 1 mV to 24.999 mV 1 V 25 mV to 109.99 mV 10 V 110 mV to 2.1999 V 100 V 2.2 V to 10.999 V 1 mV 11 V to 130 V 10 mV Adjustment Range Continuous 1-Year Absolute Uncertainty, tcal 5 C 0.25 % of output +40 V) 0.025 % of
7、 output +25 V) 0.25 % of output +40 V) 0.05 % of output +5 V) 2 Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV) Square Wave Frequency Characteristics Range 10 Hz to 10 kHz 1-Year Absolute Uncertainty, tcal 5 C (0.33 ppm of setting) Typical Aberration (from 50 % of leading/trailing edge) 25 mV to 130 V: w
8、ithin 4 s 10 mV to 25 mV: within 8 s 1 mV to 10 mV: within 14 s 1 Me input impedance (measure voltage across 1 Me input resistor) Introduction and Specifications Specifications 1 1-9 1-10. Edge Specifications Note The GHz Option offers a Fast Edge function. The specifications for the Fast Edge funct
9、ion can be found in Chapter 8 of the 5820A Operators Manual. Edge Characteristics into 50 e e load 1 Year Absolute Uncertainty, tcal 5 C Rise Time 300 ps 0/- 100 ps Amplitude Range (p-p) 4.0 mV to 2.5 V (2 % of output +200 V) Resolution 4 digits Adjustment Range 10 % around each sequence value (indi
10、cated below) Sequence Values 5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80 mV, 100 mV, 200 mV, 250 mV, 300 mV, 500 mV, 600 mV, 1 V, 2.5 V Frequency Range 1 kHz to 10 MHz (0.33 ppm of setting) Typical Jitter, edge to trigger 3 ps (p-p) within 2 ns from 50 % of rising edge (3 % of output + 2 mV) 2 to 5 ns (2 %
11、 of output + 2 mV) Leading Edge Aberrations 2 5 to 30 ns (1 % of output + 2 mV) after 30 ns 600 MHz) table in Chapter 8 of the 5820A Operators Manual. Frequency Range Leveled Sine Wave Characteristics into 50 e e 50 KHz (reference) 50 KHz to 100 MHz 100 MHz to 300 MHz 300 MHz to 500 MHz 500 MHz to 6
12、00 MHz Amplitude Characteristics Range (p-p) 5 mV to 5.0 V 1 V p-p 1 1 V p-p 1 1 V p-p 1 1 V p-p 1 1 V p-p Typical Jitter (p-p) 10 ppm 1 ppm 1 ppm 1 ppm 1 ppm Sequence 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms) Adjustment Range At least 10 % around each sequence value indicated above Ampl
13、itude Resolution 4 digits 1-Year Absolute Uncertainty, tcal 5 C 1 (2.5 ppm + 5 Hz) 20.33 ppm 0.33 ppm 0.33 ppm 0.33 ppm 1 Time marker uncertainty is 50 ppm when measured off of cardinal points: 5 s, 2 s, 1 s, 500 ms, 200 ms, 100 ms, 50 ms, 20 ms, 10 ms, 5 ms, 2 ms, 1 ms, 500 s, 200 s, 100 s, 50 s, 2
14、0 s, 10 s, 5 s, 2 s, 1 s, 500 ns, 200 ns, 100 ns, 50 ns, 20 ns, 10 ns, 5 ns and 2 ns. 2 As an example, a 5s 0.2 Hz marker would have an uncertainty of: 0.2 Hz * 2.5 ppm + 5 Hz = 5.5 Hz. The valid values are 0.2000055 to 0.1999945 Hz or 4.9998625 s to 5.0001375 s. Introduction and Specifications Spec
15、ifications 1 1-11 1-13. Wave Generator Specifications Wave Generator Characteristics Square Wave and Sine Wave into 50 e e or 1 Me e Triangle Wave into 50 e e or 1 Me e Amplitude into 1 Me: 1.8 mV to 55 V p-p into 1 Me: 1.8 mV to 55 V p-p Range into 50 e: 1.8 mV to 2.5 V p-p into 50 e: 1.8 mV to 2.5
16、 V p-p 1-Year Absolute Uncertainty, tcal 5 C, 10 Hz to 10 kHz (3 % of p-p output + 100 V) (3 % of p-p output + 100 V) Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV) 1-2-5 (e.g., 10 mV, 20 mV, 50 mV) Typical DC Offset Range 0 to (40 % of p-p amplitude) 1 0 to (40 % of p-p amplitude) 1 Ramp Linearity 2 Be
17、tter than 0.1 % 10 Hz to 10 kHz 3 Frequency Range 0.01 Hz to 100 kHz 0.01 Hz to 100 kHz Resolution 4 or 5 digits depending upon Frequency 4 or 5 digits depending upon Frequency 1-Year Absolute Uncertainty, tcal 5 C 5 (2.5 ppm + 5 Hz) 4 (2.5 ppm + 5 Hz) 4 1 The DC offset plus the wave signal must not
18、 exceed 30 V rms. 2 Applies to the 10 % to 90 % of the triangle waveform 500 mV p-p to 10 V p-p. 3 No specification below 10 Hz or above 10 kHz. 4 With 10 MHz external reference selected, the uncertainty becomes that of the external clock plus 5 Hz. 5 Uncertainties below 10 Hz are typical. 6 Square
19、wave rise/fall time typically less than 500 ns. 1-14. 1 ns Pulse Generator Specifications Pulse Generator Characteristics Positive pulse into 50 e e Typical Rise/fall Time 500 ps Typical Available Amplitudes 1.5 V, 600 mV, 150 mV, 60 mV, 15 mV Pulse Width Range 1 1 ns to 500 ns Uncertainty 5 % 200 p
20、s Pulse Period Pulse width 1 ns 20 ms to 200 ns 1 ns Pulse width 9.9 ns 20 ms to 200 ns 10 ns Pulse width 79.9 ns 20 ms to 2 s 80 ns Pulse width 500 ns 20 ms to 10 s Resolution 4 or 5 digits depending upon frequency and width 1-Year Absolute Uncertainty, tcal 5 C 0.33 ppm Pulse Skew with Trigger 3 R
21、ange 2 +30 ns to -10 ns with 250 ps resolution Uncertainty 4 500 ps 1 May generate pulses below 1 ns but pulse width accuracy is not specified 2 Assumes that trigger is used in divide by 1 mode. Other divide modes are not specified. 3 Pulse skew measured from 30 % of trigger signal amplitude to 30 %
22、 of pulse range amplitude. 4 Uncertainty specification applies only for pulse periods that are 3 s or greater in duration. Otherwise, skew uncertainty is typical. 1-15. Trigger Signal Specifications (Pulse Function) Pulse Period Division Ratio Amplitude into 50 e e (p-p) Typical Rise Time 20 ms to 2