1、_ *This bulletin supersedes TB 9-6625-2271-35, dated 5 March 1998. *TB 9-6625-2271-50 SUPERSEDED COPY DATED 5 MARCH 1998 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR AUDIO ANALYZER BOONTON, MODEL 1120-S/10 Headquarters, Department of the Army, Washington, DC 21 August 2000 App
2、roved for public release; distribution is unlimited REPORTING OF ERRORS You can help improve this publication by calling attention to errors and by recommending improvements and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be
3、mailed directly to Commander, U.S. Army Aviation and Missile Command, ATTN: AMSAM-TMD-LP, Redstone Arsenal, AL 35898-5000. You may also contact this office electronically. FAX to DSN 788-2313 (commercial 256-842-2313). A reply will be furnished directly to you. Paragraph Page SECTION I. IDENTIFICATI
4、ON AND DESCRIPTION Test instrument identification. 1 2 Forms, records, and reports. 2 2 Calibration description . 3 2 II. EQUIPMENT REQUIREMENTS Equipment required. 4 4 Accessories required . 5 4 III. CALIBRATION PROCESS Preliminary instructions. 6 5 Equipment setup and internal calibration. 7 5 Dc
5、measurement accuracy. 8 7 Ac measurement accuracy. 9 8 Frequency measurement . 10 11 Filters frequency accuracy. 11 12 Distortion measurement accuracy. 12 12 Output frequency accuracy. 13 15 Output level . 14 16 Output distortion. 15 19 SINAD measurement accuracy . 16 20 Final procedure. 17 22 Radio
6、Fans.CN 收音机爱 好者资料库 TB 9-6625-2271-50 2 SECTION I IDENTIFICATION AND DESCRIPTION 1. Test Instrument Identification. This bulletin provides instructions for the calibration of Audio Analyzer, Boonton, Model 1120-S/10. The manufacturers manual was used as the prime data source in compiling these instru
7、ctions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin. a. Model Variations. None. b. Time and Technique. The time required for this calibration is approximately 6 hours, using the dc and low frequency technique. 2. Forms, Records, and Reports
8、 a. Forms, records, and reports required for calibration personnel at all levels are prescribed by TB 750-25. b. Adjustments to be reported are designated (R) at the end of a sentence in which they appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those a
9、djustments made and designated with (R). 3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1. Table 1. Calibration Description Test instrument parameters Performance specifications Dc level measurement Range: 3, 30, 300 V F
10、S Accuracy: +1% or 3 counts whichever is greater Ac level measurement Range: 3, 30, 300 mV and 3, 30, 300 V Range: 50 Hz to 50 kHz (1 mV to 300 V) Accuracy: +1% Range: 20 to 50 Hz and 50 to 100 kHz (1 mV to 300 V) Accuracy: +2% Range: 10 to 20 Hz1 (1 mV to 300 V) Accuracy: +3% Range: 10 Hz to 100 kH
11、z1 (.3 mV to 300 V) Accuracy: +4 Flatness Range: 1 mV to 300V Range: 50 Hz to 50 kHz (1 mV to 300 V) Accuracy: +0.5% Range: 20 to 50 Hz and 50 kHz to 100 kHz (1 mV to 300 V) Accuracy: +1.0% Range: 10 to 20 Hz1 (1 mV to 300 V) Accuracy: +2.0% See footnotes at end of table. RadioFans.CN 收音机爱 好者资料库 TB
12、9-6625-2271-50 3 Table 1. Calibration Description - Continued Test instrument parameters Performance specifications Frequency counter Range: 20 Hz to 500 kHz Sensitivity: 3 mV Accuracy: +1 ppm/yr +1 count Audio filters Range: 30 kHz low-pass Accuracy: +2 kHz Range: 80 kHz low-pass Accuracy: +4 kHz R
13、ange: 220 kHz low-pass Accuracy: +20 kHz Range: 400 Hz high-pass Accuracy: +40 Hz Distortion measurement Range: 10 Hz to 20 kHz Accuracy: +1 dB Range: 20 to 100 kHz Accuracy: +2 dB Input voltage range: 10 mV to 300 V SINAD measurement Frequency range: 10 Hz to 100 kHz Input voltage range: 10 mV to 3
14、00 V Display range: -120.00 to 0.00 dB2 Accuracy: +1 dB, 20 Hz to 20 kHz +2 dB, 10 Hz to 100 kHz Output frequency Range: 20 Hz to 140 kHz Accuracy: +10 ppm +1 count Output level Range: 10 Hz to 50 kHz, 0.6 mV to 16 V Accuracy: +0.5% of setting +0.05% of range Range: 50 to 100 kHz, 0.6 mV to 16 V Acc
15、uracy: +1.0% of setting +0.05% of range Range: 100 to 140 kHz, 0.6 mV to 16 V Accuracy: +1.5% of setting +0.1% of range Flatness Ref 1 kHz (.30 mV to 8 V into 50) Accuracy: +0.5%; 10 Hz1 to 50 kHz Output distortion Range: 10 Hz to 20 kHz, 80 kHz bandwidth Accuracy: 0.01% (-80 dB) Range: 20 to 50 kHz
16、, 220 kHz bandwidth Accuracy: 0.02% (-74 dB) Range: 50 to 100 kHz, 500 kHz bandwidth Accuracy: 0.056% (-65 dB) Range: 100 to 140 kHz, 500 kHz bandwidth Accuracy: 0.1% (-60 dB) 1Not checked below 20 Hz. 2Not verified below -30 dB. RadioFans.CN 收音机爱 好者资料库 TB 9-6625-2271-50 4 SECTION II EQUIPMENT REQUI
17、REMENTS 4. Equipment Required. Table 2 identifies the specific equipment to be used in this calibration procedure. This equipment is issued with Secondary Reference Calibration Standards Set NSN 4931-00-621-7878. Alternate items may be used by the calibrating activity. The items selected must be ver
18、ified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the standard and TI. 5. Accessories Required. The acces
19、sories required for the calibration are common usage accessories, issued as indicated in paragraph 4 above and are not listed in this calibration procedure. Table 2. Minimum Specifications of Equipment Required Common name Minimum use specifications Manufacturer and model (part number) AC MEASUREMEN
20、T STANDARD Range: 20 Hz to 120 kHz 2.97 mV to 16.256 V Accuracy: .125% John Fluke, Model 5790A (5790A) CALIBRATOR Range: 3 to 300 V dc Accuracy: 0.2755% Range: 20 Hz to 100 kHz, 3 mV to 300 V Accuracy: .25% John Fluke, Model 5700A/CT (p/o MIS-35947); w/power amplifier, John Fluke, Model 5725A (5725A
21、); w/ac divider, John Fluke, Model 7405A-4207 (7405A-4207) MULTIMETER ACV range: 20 Hz to 100 kHz, 1 mV to 3 V rms Accuracy: 0.175% Hewlett-Packard, Model 3458A/E02 (3458A/E02) FREQUENCY COUNTER Range: 20 Hz to 140 kHz Accuracy: 2.5 ppm Hewlett-Packard, Model 5345A (MIS-28754/1 Type 1) SPECTRUM ANAL
22、YZER1 Range: 1 kHz -75 dBm Hewlett-Packard, Model 3585A (MIS-35951) SYNTHESIZER/FUNCTION GENERATOR Frequency range: 1200 Hz to 80 kHz Attenuator Accuracy: 0.25 dB 1200 Hz 0.5 dB 80 kHz Hewlett-Packard, Model 3325A (MIS-35932) TIME/FREQUENCY WORKSTATION Range: 10 MHz Accuracy: 0.25 ppm AUTEK Systems
23、Corp., Model 620 (MIS-38946) 1Used only for adjustments. TB 9-6625-2271-50 5 SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration process. Personnel should become familiar with the applicable sections before