1、MEGGER CFL510E - TDR1000 Time Domain Reflectometer MEGGER User Guide Manuel Utilisateur Bedienungsanleitung Gua del usuario Guida per lutente Gebruikersgids Brugervejledning Kyttohjeet Brukerhndbok Anvndarmanual RadioFans.CN 收音机爱 好者资料库 Introduction6 User Controls and Display8 Operation10 Balance con
2、trol12 Velocity factor12 Pulse widths13 Techniques to improve accuracy14 Test and cable from both ends14 Care and maintenance14 Specification15 Repair and Warranty 18 MeterCenter (800) 230-6008 2 Contents RadioFans.CN 收音机爱 好者资料库 Symbols used on the instrument Caution: Refer to accompanying notes Equ
3、ipment protected throughout by Double or reinforced Insulation Instrument flash tested to 3.7 kV r.m.s. for 1 min. Equipment complies with current EU Directives 3 3.7 RadioFans.CN 收音机爱 好者资料库 SAFETY WARNINGS This instrument meets the safety requirements of IEC 61010 part 1 to 150V cat III. If it is t
4、o be used in situations where hazardous live voltages may be encountered then an additional blocking filter must be used. CAUTION (Risk of electric shock) Although this tester does not generate any hazardous voltages, circuits to which it can be connected could be dangerous due to electric shock haz
5、ard or due to arcing (initiated by short circuit). While every effort has been made by the manufacturer to reduce the hazard, the user must assume responsibility for ensuring his, or her, own safety. 4 The instrument should not be used if any part of it is damaged. Test leads, probes and crocodile c
6、lips must be in good order, clean and with no broken or cracked insulation. Check that all lead connections are correct before making a test. Disconnect the test leads before accessing the battery compartment. Refer to operating instructions for further explanation and precautions. Safety Warnings a
7、nd Precautions must be read and understood before the instrument is used. They must be observed during use. NOTE THE INSTRUMENTS MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. 5 Thank you for purchasing this quality AVO product. Before using your new instrument please take the time to
8、read this user guide, ultimately this will save you time, advise you of any precautions you need to take and could prevent damage to yourself and the instrument. The MEGGER TDR1000 is an advanced instrument capable of identifying a wide range of cable faults. The instrument uses a technique called T
9、ime Domain Reflectometry (TDR) which in many ways is similar to radar. Narrow pulses of electrical energy are transmitted along a pair of conductors in a cable. The pulse travels through the cable at a velocity determined by the insulation between the conductors and this resistance to the flow of th
10、e pulse is characterised as impedance for the cable. Changes in cable impedance will cause a proportion of the pulse to be reflected. The pulse velocity is normally described as a fraction of the speed of light and is called the Velocity Factor. By measuring the time between the transmitted pulse an
11、d the reception of the reflected pulse, and multiplying this by the speed of light and the velocity factor, the actual distance to the reflection point can be given. Faulty cables, poor joints or discontinuities will all cause a change in impedance. Impedances higher than the cables cause a normal r
12、eflection; Impedances lower than the cables cause an inverse reflection. Matched terminations absorb all the pulse hence no reflection will occur, the cable appearing endless. Open or Short circuits will reflect all the pulse energy and the TDR will not see the cable beyond that fault. 6 Introductio
13、n As a pulse is transmitted down a cable, the size and shape of that pulse is gradually attenuated by losses in the cable: the pulse gets smaller in height and more spread out. The level of attenuation is determined by the cable type, the condition of the cable and any connections along its length.
14、The limit of how far you can see is determined by the point beyond which you will not discern a reflection. To maximise the instruments range, the TDR1000 has an adjustable gain setting on its input to allow you to discern a reflection from farther away. By combining this variable gain with increasi
15、ng pulse widths, the TDR1000 can discern faults up to 3 Km away. The MEGGER TDR1000 can be used on any cable consisting of at least two insulated metallic elements, one of which may be the armouring or screen of the cable. The TDR1000 has internal matching networks to allow testing of 25 , 50 , 75 a
16、nd 100 cables. (These correspond to power, coaxial data and data/telecoms cable). The instrument can be closely balanced to the cable using the balance control; this allows long lengths of cable to be easily tested. The velocity factor can be adjusted to match the cable, thus allowing an accurate di
17、stance measurement to be directly read from the instrument. To enable a wider range of faults to be detected, the gain of the instrument is adjustable; this allows more minor faults to be identified along the entire length of the cable. Other setting options include changing the distance units betwe
18、en metres and feet, changing the propagation velocity units between a ratio and a distance per microsecond. Display contrast is fully adjustable to compensate for all viewing conditions. A backlight aids viewing in low ambient light conditions. The batteries to power the instrument are housed in the
19、 compartment on the case back, the cover is held in place with two screws. The batteries are held in a carrier, which hold the batteries securely, and allow a quick change of rechargeable battery packs. The instrument can be powered by manganese-alkali, nickel-cadmium or nickel-metal-hydride batteri
20、es. All cells must be of the same type. 7 User Controls and Display: The controls of the TDR have been arranged such that the instrument is easy to use and easy to learn how to use. The instrument controls consist of the following: 1) Instrument Display: The display shows the user the current settin
21、gs of the instrument and the reflected energy trace from the cable connected. 2) Balance: This is an analogue control which allows the user to match the instrument impedance to that of the cable under test more closely, thus enabling faults to be more easily detected. 3) Cursor Left: This control mo
22、ves the cursor left or selects a lower value depending on which mode the instrument is in. 4) Menu: This control is a bi-directional switch and can be used to navigate around the various control options. The control option to be adjusted is shown in reverse video and its name is displayed in the top
23、 left-hand corner of the display. Menu options are cursor, range, VF, V unit, Zo the second is a short circuit at 2000m away. 11 CURSOR0.67R 3000m2000m100 CURSOR0.67R 3000m2000m100 Balance Control Without Balance Control (point 2 in the User Controls and Display section) the transmitted pulse would
24、be visible at the beginning of the trace, swamping any reflections within the pulse length (the dead zone). The balancing circuit attempts to match the characteristic impedance of the cable under test to produce an equivalent pulse. Subtracting this equivalent pulse from the transmitted pulse effect