1、 43B Power Quality Analyzer Service Manual 4822 872 05385 December 2001 2001 Fluke Corporation, All rights reserved. Printed in the Netherlands All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 SERVICE CENTERS To locate an authorized service center, visit us on
2、the World Wide Web: or call Fluke using any of the phone numbers listed below: +1-888-993-5853 in U.S.A. and Canada +31-402-678-200 in Europe +1-425-446-5500 from other countries RadioFans.CN 收音机爱 好者资料库 i Table of Contents ChapterTitlePage 1Safety Instructions. 1-1 1.1 Introduction. 1-3 1.2 Safety P
3、recautions. 1-3 1.3 Caution and Warning Statements. 1-3 1.4 Symbols. 1-3 1.5 Impaired Safety. 1-4 1.6 General Safety Information. 1-4 2Characteristics . 2-1 2.1 Introduction. 2-3 2.2 Safety Specifications. 2-3 2.3 Function Specifications. 2-4 2.3.1 Electrical functions. 2-4 2.3.2 Scope. 2-5 2.3.3 Me
4、ter. 2-7 2.3.4 Record. 2-7 2.4 Miscellaneous . 2-8 2.5 Current Probe. 2-8 2.6 Environmental Conditions . 2-9 2.7 Electromagnetic Immunity. 2-10 3Circuit Descriptions . 3-1 3.1 Introduction. 3-3 3.2 Block Diagram. 3-3 3.2.1 Input 1 - Input 2 Measurement Circuits. 3-3 3.2.2 Trigger Circuit. 3-5 3.2.3
5、Digital Circuit. 3-5 3.2.4 Power Circuit. 3-6 3.2.5 Start-up Sequence, Operating Modes. 3-7 3.3 Detailed Circuit Descriptions. 3-9 3.3.1 Power Circuit. 3-9 3.3.2 Input 1 - Input 2 Measurement Circuits. 3-14 3.3.3 Trigger Circuit. 3-20 RadioFans.CN 收音机爱 好者资料库 43B Service Manual ii 3.3.4 Digital Circu
6、it. 3-24 4Performance Verification. 4-1 4.1 Introduction. 4-3 4.2 Equipment Required For Verification . 4-3 4.3 How To Verify. 4-3 4.4 Display and Backlight Test. 4-4 4.5 Input 1 and Input 2 Tests in the SCOPE MODE. . 4-5 4.5.1 Input 1 Trigger Sensitivity Test. 4-6 4.5.2 Input 1 Frequency Response U
7、pper Transition Point Test. 4-7 4.5.3 Input 1 Frequency Measurement Accuracy Test. 4-7 4.5.4 Input 2 Frequency Measurement Accuracy Test. 4-9 4.5.5 Input 2 Trigger Level and Trigger Slope Test. 4-10 4.5.6 Input 2 Trigger sensitivity Test. 4-12 4.5.7 Input 1 Trigger Level and Trigger Slope Test. 4-13
8、 4.5.8 Input 1 and 2 DC Voltage Accuracy Test. 4-15 4.5.9 Input 1 and 2 AC Voltage Accuracy Test. 4-17 4.5.10 Input 1 and 2 AC Input Coupling Test . 4-19 4.5.11 Input 1 and 2 Volts Peak Measurements Test . 4-20 4.5.12 Input 1 and 2 Phase Measurements Test. 4-21 4.5.13 Input 1 and 2 High Voltage AC 5
9、0 V; 500 V; 1250 V minimum step.1 mV; 10 mV; 100 mV; 1 V (1 % + 10 counts) Current ranges (auto).50.00 A to 500.0 kA, 1250 kA minimum step.10 mA; 0.1 A; 1 A; 100 A, 1 kA (1 % + 10 counts) Frequency range .10.0 Hz to 15.0 kHz 40.0 to 70.0 Hz.(0.5 % + 2 counts) resolution. 0.1 Hz; 1 Hz; 10 Hz; 100 Hz
10、CF Crest Factor range.1.0 to 10.0 (5 % + 1 count) Power (1 and 3 phase / 3 conductor / balanced loads) Readings .Watt, VA, VAR, PF, DPF, Hz Watt, VA, VAR ranges (auto).250 W to 250 MW, 625 MW, 1.56 GW when selected: total ( %r):.(2 % + 6 counts) when selected: fundamental ( %f):.(4 % + 4 counts) DPF
11、.0.00 to 1.00 0.00 to 0.25.not specified 0.25 to 0.90. 0.04 0.90 to 1.00. 0.03 Characteristics 2.3 Function Specifications2 2-5 PF.0.00 to 1.00 0.04 Frequency range .10.0 Hz to 15.0 kHz 40.0 to 70.0 Hz.(0.5 % + 2 counts) Harmonics Number of harmonics.DC.21, DC.33, DC.51 Readings / Cursor readings Vr
12、ms.fund. (3 % + 2 counts) .31st (5 % + 3 counts).51st (15 % + 5 counts) Irms.fund. (3 % + 2 counts) .31st (5 % + 3 counts).51st (15 % + 5 counts) Watt.fund. (5 % + 10 counts).31st (10 % + 10 counts) 51st (30 % + 5 counts) Frequency of fundamental. 0.25 Hz Phase.fund. 3 . 51st 15 K-factor (in Amp and
13、 Watt). 10 % Sags so only one of these functions needs to be verified. 4.2 Equipment Required For Verification The primary source instrument used in the verification procedures is the Fluke 5500A. If a 5500A is not available, you can substitute another calibrator as long as it meets the minimum test
14、 requirements. Power Quality Option 5520A-PQ is not strictly necessary for the tests in this chapter, but it offers useful test signals to test modes such as Harmonics and Sags the test pattern as shown in Figure 4-1 may not be visible or hardly visible. Observe the display closely, and verify that
15、no light pixels are shown. Figure 4-1. Display Pixel Test Pattern Performance Verification 4.5 Input 1 and Input 2 Tests in the SCOPE MODE.4 4-5 11. Press . The test pattern is removed; the test tool shows Contrast (CL 0110):MANUAL 12. Press (CAL) . The test tool shows the display test pattern shown
16、 in Figure 4-1, at default contrast. Observe the test pattern closely, and verify that the no pixels with abnormal contrast are present in the display pattern squares. Also verify that the contrast of the upper left and upper right square of the test pattern are equal. 13. Press . The test pattern i