FLUKE t5ServiceInfo 电路图.pdf
《FLUKE t5ServiceInfo 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE t5ServiceInfo 电路图.pdf(9页珍藏版)》请在收音机爱好者资料库上搜索。
1、PN 686953 March 1998 Rev.1, 9/01 1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies. T5-600/T5-1000 Electrical Tester Service Information Introduction This service information sheet provides the following information for the
2、 T5-600 and T5-1000 Electrical Testers (hereafter referred to as the tester). Safety information Parts and service information Specifications Cleaning procedure Required equipment Performance tests Parts and accessories list Battery replacement procedure For operating instructions, refer to the T5-6
3、00/T5-1000 Electrical Tester Instruction Sheet. Safety Information Warning To avoid possible electric shock or personal injury, follow these guidelines: Do not use the tester if it is damaged or operating abnormally. Protection may be impaired. Before each use: Make sure the battery door is closed a
4、nd latched. Inspect the tester and test leads. Look for cracks, missing plastic, exposed metal, or damaged insulation. Replace damaged test leads before using the tester. Verify the testers operation by measuring a known voltage. Replace the batteries as soon as the low battery indicator (?) appears
5、. Do not use the tester around explosive gas, vapor or dust. Do not apply more than the rated voltage, as marked on the tester, between terminals or between any terminal and earth ground. Refer servicing to qualified personnel. Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc. Wh
6、en using the probes, keep your fingers behind the finger guards on the probes. Connect the common test lead before you connect the live test lead. Disconnect the live test lead first. The display will not display hazardous voltages in display hold. The voltage indicator LED () continues to operate.
7、99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentD RadioFans.CN 收音机爱 好者资料库 T5-600/T5-1000 Electrical Tester 2 Parts and Service The tester is warranted to be free from defects in material and workmanship for two years, while under normal
8、 use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet. To order parts, or for warranty service, contact Fluke as follows: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Eur
9、ope: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Flukes Web site at . Specifications Accuracy is specified for one year after calibration, at 18 C to 28 C (64 F to 82 F) with relative humidity to 90 %. AC conversions are ac-coupled,
10、 average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows: ( % of reading + number of least significant digits) Temperature coefficient of 0.1 x (specified accuracy)/ C for 28 C (82.4 F) Serial Numbers 79560000 and Above General Specifica
11、tions CalibrationOne-year calibration cycle. Maximum Voltage Between any Terminal and Earth Ground T5-600: 600 V rms, Overvoltage Category III, Pollution Degree 2 T5-1000: 1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV, Pollution Degree 2 Maximum Voltage Between Current For
12、k and Earth Ground T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.) TemperatureOperating: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F) AltitudeOper
13、ating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft) Relative Humidity0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F); 0 % to 45 %, 40 C to 50 C (104 F to 122 F) Battery Type and LifeAA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chlorid
14、e Shock, Vibration1 m drop at -10 C to + 50 C (14 F to 122 F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum) Surge ProtectionT5-600: 6 kV per IEC 61010, T5-1000: 8 kV per IEC 61010 Enclosure RatingIP 52 per IEC
15、60529, no vacuum applied RadioFans.CN 收音机爱 好者资料库 Test Equipment Depot? 99 Washington Street? Melrose, MA 02176-6024? ? ? 800-517-8431? 781-665-0780 FAX Service Information Serial Numbers 79559999 and Below 3 General Specifications (continued) RF Field Specification0.5 % full scale + (specified accur
16、acy) at 3 V/m SafetyComplies with ANSI/ISA-S82.01-94, UL Classified to IEC 61010, CSA/CAN C22.2 No.1010.1- 92, and EN61010-1 1995. EMCEN 61326 Certifications In accordance with IEC 61010-1 54CJ Resolution and Accuracy FunctionT5-600 RangeT5-1000 RangeResolutionAccuracy ?600 V rms1000 V rms1 V(1.5 %
17、+ 2 digits) ?600 V1000 V1 V(1 % + 1 digit) ?100.0 A100.0 A0.1 A(3 % + 3 digits) ?1000 1000 1 (1 % + 2 digits) Input Characteristics Input Protection FunctionT5-600T5-1000 Input impedance (nomial ? 600 V rms1000 V rms1 M, 100 pF ac-coupled ? 600 V rms1000 V rms1 M, 100 pF ? 600 V rms1000 V rms Open C
18、ircuit Test VoltageShort Circuit Current ?1.65 V dc (nominal)600 A CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in large buildings. CAT IV protects against transients from a primary supply such as an electricity meter or an
19、overhead or underground utility service. Serial Numbers 79559999 and Below General Specifications CalibrationOne-year calibration cycle. Maximum Voltage Between any Terminal and Earth Ground T5-600: 600 V rms, Overvoltage Category III T5-1000: 1000 V rms, Overvoltage Category III TemperatureOperatin
20、g: -10 C to +50 C (14 F to 122 F); Storage: -30 C to +60 C (-22 F to +140 F) AltitudeOperating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft) Relative Humidity0 % to 95 %, 5 C to 30 C (41 F to 86F); 0 % to 75 %, 30 C to 40 C (86 F to 104 F); 0 % to 45 %, 40 C to 50 C (104 F to 122 F) Battery Type a
21、nd LifeAA (2); 400 hours continuous with alkaline; 200 hours continuous with zinc chloride RadioFans.CN 收音机爱 好者资料库 T5-600/T5-1000 Electrical Tester 4 General Specifications (continued) Shock, Vibration1 m drop at 15 C to 35 C (59 F to 95 F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1993. Sinusoidal vi
22、bration per MIL-PRF-28800F for a Class 2 instrument (5 Hz to 55 Hz, 3 g maximum) Surge ProtectionT5-600: 6 kV per IEC 1010-1, 1990-09: T5-1000: 8 kV per IEC 1010-1, 1990-09 Enclosure RatingIP 52 per IEC 60529, no vacuum applied RF Field Specification0.5 % full scale + (specified accuracy) at 3 V/m S