1、M OTS60SX Semi Automatic Oil Test Set User Guide Guide de lutilisateur Gebrauchsanleitung Gua del usuario RadioFans.CN 收音机爱 好者资料库 ContentsGuide de lutilisateur - p14Gebrauchsanleitung - s24Gua del usuario - p34 Safety Warnings 3 General Description4 Preparing the Oil Test Set5 Operation5 Oil Samplin
2、g5 Preparing the Oil Test Vessel5 Control Panel6 Breakdown Testing6 Withstand Testing6 Warning messages7 Testing to National Standards7 Sampling and Testing Oil for Dielectric strength 8 Oil Testing Specifications10 Test Set Specification10 Accessories12 Repair and Warranty13 2. RadioFans.CN 收音机爱 好者
3、资料库 3. SAFETY WARNINGS Safety warnings and precautions must be read and understood before the instrument is used. They must be observed during use. The oil test set must be properly earthed. The test chamber must be kept scrupulously clean. If the test chamber cover is cracked or damaged in any way
4、the test set must not be used, but sent for repair to an authorised agent. The chamber door hinges are fitted with earth contact springs. These must not be damaged or corroded. The correct type of oil test vessel must be used, and correctly positioned on the support horns before carrying out any tes
5、ting. Replacement fuses must be of the correct type and rating. NOTE THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. Symbols used on the Test Set Caution: Refer to accompanying notes. Risk of electric shock. Equipment complies with relevant EU Directives RadioFans.CN 收音机爱 好
6、者资料库 Mains Input Panel General Description 4. The OTS60SX is a lightweight, semi-automatic, oil dielectric strength test set. The instrument is suitable for protected field use and can be powered from a range of mains supplies. The maximum 60 kV output allows tests to be performed on oil from a wide
7、 variety of electrical installations including transformers, circuit breakers and other equipment. The operation of the test set is extremely simple and the results are displayed on a bright LED display. A selection of vessels enables the instrument to be configured for a variety of tests. The semi
8、automatic operation allows all types of testing to be performed. An automatic one minute timer operates when the high voltage is paused for Withstand (proof) testing. Oil samples can also be tested according to breakdown specification by using a suitable sequence of tests. The instrument is construc
9、ted in a strong, sheet steel case. Handles are provided for ease of transportation. A pouch is located on the side of the instrument to contain accessories such as the power cord, additional electrodes and the spacing gauge. An optional carrying case incorporates a shoulder strap. A number of vessel
10、s are available suitable for testing to a wide range of national specifications. Three types of electrode are available; spherical (IEC type), mushroom (VDE/ASTM D1816) and cylindrical (ASTM D877). The oil sample can be stirred by a motor driven impeller available in selected vessels. The vessel is
11、located in the top of the instrument and covered by a transparent polycarbonate door with a mesh screen so that the oil breakdown can be viewed. A safety interlock ensures that the high voltage is disconnected when the chamber door is opened. Voltage Selector Mains power input Fuseholders Preparing
12、the Oil Test Set 5 Ensure that the instrument is properly earthed. The test set must be connected to a socket (receptacle) with a protective earth (ground) conductor. Before connecting the instrument to the mains power supply, ensure the voltage selector located on the mains input panel is set to th
13、e required voltage. The correct fuse value should be fitted for the supply voltage (See Specification). The fuse holders are located next to the supply inlet. Power cord If the power cord plug is not suitable for your type of socket outlets, do not use an adaptor. You should use a suitable alternati
14、ve power cord, or if necessary change the plug by cutting the cord and fitting a suitable plug. The colour code of the cord is: Earth (Ground)Yellow / Green NeutralBlue Phase (Line)Brown If using a fused plug, a 3 Amp fuse to BS 1362 should be fitted. Note: A plug severed from the power cord should
15、be destroyed, as a plug with bare conductors is hazardous in a live socket outlet. Operation Oil Sampling The relevant test specification will give advice on the best methods to use when sampling oil. Contamination of the oil may lead to misleading results. Always run off a small amount of oil befor
16、e sampling and ensure that the flow of oil is steady. Preparing the Oil Test Vessel Optional vessels are available for testing to national specifications (See table and Accessories). Some specifications require stirring of the oil sample as well as different test electrode shapes. Once the correct v
17、essel has been chosen for the required test the electrode gap should be set. This is achieved by adjusting the threaded nut on the side of the vessel. Small adjustment of the gap can be made by hand tightening the nuts and rotating the spindle. A slot is provided on the end of the spindle for this p
18、urpose. A set of individual 0,5mm feeler gauges is provided for accurate measurement of the gap. Gauges can be combined to measure from 1mm to 4mm. (see page 12) Clean the vessel in accordance with the instructions given in the relevant test specification. Allowing space for the lid to be fitted, fi
19、ll the vessel with the oil sample. The amount of oil required will depend upon the type of lid used. Load the vessel in the test chamber, connecting the motor power cable, if fitted, and carefully shut the door so that the safety interlock switch operates. Control Panel The control panel contains th
20、e supply on/off switch, the high voltage display, voltage ramp rate selection and the three control keys. The operation of the instrument is very simple. When the instrument is turned on, the on/off switch is illuminated, and the software edition code will be flashed up on the display. Breakdown Tes
21、ting The suitable ramp rate for the test to be performed should be selected using the L key. The START key will then turn on the high voltage. The red H.V. indicator will light to show the output voltage is on. The voltage will then increase at the selected ramp rate and the corresponding value will
22、 be shown on the display. If oil breakdown occurs the instrument will detect this and immediately shut off the high voltage. The breakdown voltage value will be left on the display until the next test is started. If no breakdown occurs, the high voltage will rise to 60 kV. At this point the test vol
23、tage is cut off and this maximum value left on the display. A test can be stopped at any time by pressing the STOP I button. This will remove the high voltage from the oil sample and leave the display at 0,0 kV. Withstand Testing A Withstand (proof) test can be carried out by pressing the PAUSE key.
24、 This will maintain the high voltage at the current value for one minute. After this time the high voltage will continue to rise until one of the following:- 1) Oil breakdown occurs 2) The STOP Ikey is pressed 6. Mains on/off switch HV on indicator Ramp rate selection Main controls LED Display L Con