1、MEGGERBMM2000 Series Insulation Multimeters User Guide SAFETY WARNINGS Safety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use. The circuit under test must be de-energised and isolated before connections are made except for voltage
2、measurement. Circuit connections must not be touched during a test. After insulation tests, capacitive circuits must be allowed to discharge before disconnecting the test leads. The Live Circuit Warning and Automatic Discharge are additional safety features and should not be regarded as a substitute
3、 for normal safe working practice. Replacement fuses must be of the correct type and rating. Failure to fit the correctly rated fuse will result in damage to the instrument in the event of an overload. Test leads, including crocodile clips, must be in good order, clean and have no broken or cracked
4、insulation. Ensure that hands remain behind guards of probes/clips when testing. U.K. Safety Authorities recommend the use of fused test leads when measuring voltage on high energy systems. NOTE THE INSTRUMENTS MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS Users of this equipment and/o
5、r their employers are reminded that Health and Safety Legislation require them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as from inadvertent short circuits. Where the assessments show that th
6、e risk is significant then the use of fused test leads constructed in accordance with the HSE guidance note GS38 Electrical Test Equipment for use by Electricians should be used. Symbols used on the instruments are: Caution, risk of electric shock. Caution, refer to User Guide. Equipment protected t
7、hroughout by Double Insulation (Class II). Equipment complies with current EU Directives. Equipment must not be connected to installations 500V. 500V RadioFans.CN 收音机爱 好者资料库 GENERAL DESCRIPTION The MEGGER BMM2000 Series instruments are battery powered Insulation and Continuity testers, with a measur
8、ement capability from 0,01 Continuity to 200G Insulation. Offering multi-voltage facilities, the instruments take full advantage of microprocessor technology and feature a large liquid crystal display combining digital and analogue readings. The analogue display has the benefit of indicating trends
9、and fluctuations in readings, while the digital readout gives direct accurate results. The display is also backlit giving clear visibility even in low light conditions. The BMM2000 Series instruments have the unique capability of being able to measure voltages down to a resolution of 0,1mV. This giv
10、es the user the option to fit a wide variety of transducers to further enhance the capabilities of the BMM Series instruments, eg temperature or humidity measurement. A customised connector on the top of the instrument enables the optional MEGGER SP1 Switched Probe to be used for two handed probe op
11、eration. The 250V, 500V and 1000V ranges can be used to test electrical installations in compliance with BS7671 (16th Edition IEE Wiring Regulations) IEC364 and HD384, since each range has a 1mA minimum test current at the minimum pass values of insulation specified in these documents. The 100V rang
12、e is ideal for testing telecommunications equipment which would be damaged by higher voltages. The 50V range is useful for testing sensitive equipment, such as electronic components, and computer peripherals. The BMM2000 series instruments have a current facility which enables up to 500mA to be meas
13、ured, (not ESD models), this together with , V and mV ranges means that the instrument can realistically be used in situations where previously a multimeter would be needed.For higher currents an optional current clamp is available. Designed to IEC1010-1 the BMM2000 Series are protected against conn
14、ection to a 500V Category III supply. The instruments have a basic accuracy of 2% at 20C. The instruments are waterproof and dustproof to IP54. This helps maintain accuracy and ensures maximum reliability in harsh environments. OPERATION Refer to Safety Warnings before using the instrument Testing i
15、s automatically inhibited if: An external voltage 25V is present when switched to any insulation range position. An external voltage 10V is present on all other ranges (excluding OFF/V). The external voltage is indicated on the display, on insulation ranges an audible bleeper will sound if a test is
16、 attempted. RadioFans.CN 收音机爱 好者资料库 Live Circuit Warning When more than 25V (10V on 10V insulation range) is applied to the terminals in the insulation ranges, the instrument defaults to a voltmeter and gives an audible warning if a test is attempted.On all other switch positions except OFF/V when a
17、pprox 10V is applied the default voltmeter will be activated. Testing will be inhibited. Voltage Testing on High Energy Systems Use extreme care when using or measuring voltages above 30V, particularly in high energy systems. Fused test leads are available as optional accessories for local situation
18、s where increased protection is required. Auto-shut Off To conserve battery life, Auto-shut Off (preceded by a series of bleeps) operates after approx. 10 minutes of instrument inactivity on insulation, 5 minutes on all other ranges. If the instrument is switched on whilst holding thekey, the Auto s
19、hut-off time is extended to 60 minutes. To restore operation after Auto-shut Off, either select OFF followed by the required switch position. Note: It is recommended that the instrument is switched to the OFF position when not in use. Backlight The backlight is activated by pressing thekey. The back
20、light will remain illuminated for approx. one minute before automatically switching off to conserve battery life, alternatively thekey can be re-pressed. Insulation Tests (M) (See fig. 1) The insulation tests apply a known voltage to the circuit under test and measure the resulting leakage current.
21、The circuit under test must be completely de-energised and isolated before test connections are made. Insulation tests are only initiated when the TEST button is pressed. 1. Set the range switch to the test voltage required. 2. Connect the test leads, first to the instrument, and then to the isolate
22、d item under test. 3. Press the TEST button to activate the test voltage. Take the reading. 4.Release the TEST button at the end of the test. 5. Any capacitive circuits charged during a test will automatically discharge. If significant voltage remains the voltage warning will occur and the voltage p
23、resent displayed. 6. Remove the test leads only when no voltage is indicated. Locking Test Button (ltb) When it is desired to do a long insulation test, the test can be locked on by pressing thekey while the test button is held down. The warningwill appear on the display and both buttons may be rele
24、ased whilst the test continues. The next press of the test button will terminate the test. RadioFans.CN 收音机爱 好者资料库 Note: There is a short delay on the first operation of 1000V range, each time the range is selected. This is to prevent accidental application of 1kV. The M range features a leakage cur