1、PIN NO: 009048-004 TB 9-4931-383-50 Change 4 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARAD, JOHN FLUKE, MODELS 332A, 332B, AND 332B/AF Headquarters, Department of the Army, Washington, DC 9 March 1988 TB 9-4931-383-50, 26 July 1976, is changed as follows: 1. R
2、emove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page. Remove pagesInsert pages 3 through 83 through 8 11 and 1211 and 12 2. File this change sheet in front of the publication for reference purposes. By Order of the
3、Secretary of the Army: CARL E. VUONO General, United States Army Chief of Staff Official: R. L. DILWORTH Brigadier General, United States Army The Adjutant General Distribution: To be distributed in accordance with Std. IDS No. RLC-1500, 17 Jun 85. RadioFans.CN 收音机爱 好者资料库 PIN NO: 009048-003 TB 9-493
4、1-383-50 Change 3 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARAD, JOHN FLUKE, MODELS 332A, 332B, AND 332B/AF Headquarters, Department of the Army, Washington, DC 18 February 1983 TB 9-4931-383-50, 26 July 1976, is changed as follows: 1. Remove old pages and ins
5、ert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page. Remove pagesInsert pages 19 and 2019 and 20 2. File this change sheet in front of the publication for reference purposes. By Order of the Secretary of the Army: E. C. MEYER General, Un
6、ited States Army Chief of Staff Official: ROBERT M. JOYCE Major General, United States Army The Adjutant General Distribution: To be distributed in accordance with Std. IDS. No. RLC-1500, 9 September 1982. RadioFans.CN 收音机爱 好者资料库 This is a reprint, including Changes 1 and 2. _ *This bulletin superse
7、des TB 9-4931-383-50, 31 January 75, including all changes. *TB 9-4931-383-50 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROCEDURE FOR VOLTAGE STANDARD, JOHN FLUKE MODELS 332A, 332B, AND 332B/AF Headquarters, Department of the Army, Washington, DC 26 July 1976 REPORTING OF ERRORS You can
8、help improve this publication by calling attention to errors and by recommending improvements and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be mailed directly to Commander, U.S. Army Aviation and Missile Command, ATTN: AMSA
9、M-TMD-EP, Redstone Arsenal, A1 35898-5400. You may also contact this office electronically. E-mail address is tmde- eredstone.army.mil. FAX to DSN 788-2313 (commercial 256-842-2313). A reply will be furnished directly to you. ParagraphPage SECTIONI.IDENTIFICATION AND DESCRIPTION Test instrument iden
10、tification.12 Calibration data card, DA Form 2416.22 Calibration description.32 II.EQUIPMENT REQUIREMENTS Equipment required .43 Accessories required.53 III.CALIBRATION PROCESS FOR MODEL 332A Preliminary instructions .64 Regulation.77 Zero voltage .89 Linearity.910 Output accuracy and tracking.1012
11、Panel meter, voltage trip, and current limit.1116 Power supply .1217 Final procedure.1318 RadioFans.CN 收音机爱 好者资料库 TB 9-4931-383-50 2 IV.CALIBRATION PROCESS FOR MODELS 332B AND 332B/AF Preliminary instructions . 1418 Regulation. 1521 Zero voltage . 1622 Linearity (model 332B). 1723 Linearity (model 3
12、32B/AB). 1826 Output accuracy and tracking. 1927 Panel meter (voltage). 2030 Voltage trip. 2131 Panel meter (current) and current limit. 2232 Power supply (model 332B). 2333 Power supply (model 332B/AF) . 2434 Final procedure. 2534 SECTION I IDENTIFICATION AND DESCRIPTION 1. Test Instrument Identifi
13、cation. This bulletin provides instructions for the calibration of Voltage Standard, John Fluke Models 332A, 332B, and 332B/AF. The manufacturers instruction manuals were used as the prime data source in compiling these instructions. The equipment being calibrated will be referred to as the TI (test
14、 instrument) throughout this bulletin. a. Model Variations. On model 332A with serial numbers 123 through 332, the X position of voltage dials 2 through 7 is labeled 10. b. Time and Technique. The time required for this calibration is approximately 4 hours, using the dc and low frequency technique.
15、2. Calibration Data Card, DA Form 2416 a. Forms, records, and reports required for calibration personnel at all levels are prescribed by TM 38-750. DA Form 2416 must be annotated in accordance with TM 38-750 for each calibration performed. b. Adjustments to be reported on DA Form 2416 are designated
16、 (R) at the end of the sentence in which they appear. When adjustments are in tables, the (R) will follow the designated adjustment. Report only those adjustments made and designated with (R). 3. Calibration Description. TI parameters and performance specifications which pertain to this calibration
17、are listed in table 1. TB 9-4931-383-50 3 Table 1. Calibration Description Test Instrument ParametersPerformance Specifications1 Voltage range10-v range; 0 to 11.111110 v (1-v steps) 100-v range; 0 to 111.11110 v (10-v steps) 1000-v range; 0 to 1111.1110 v (100-v steps) Resolution Accuracy0.1 ppm of
18、 range (1 v max) 10-v range; (0.003% of setting +10 v) 100-v range; (0.003% of setting +20 v) 1000-v range; (0.003% of setting +40 v) Over-voltage protectionAutomatically disables output voltage if level exceeds selected value (10 to 110% of selected range). Output current0 to 50 ma (any voltage set
19、ting) Over-current protectionAutomatically limits output current to present level; between 1 and 60 ma. Line and load regulation0.0005% of setting or 25 v for 10% change in line voltage from nominal. 1The specifications for models 332B and 332B/AF have been derated to the same as model 332A which ar
20、e, reflected in this calibration description. SECTION II EQUIPMENT REQUIREMENTS 4. Equipment Required. Table 2 identifies the specific equipment used in this calibration procedure. This equipment is issued with secondary reference standards calibration set NSN 4931-00-621-7878, and is to be used in
21、performing this procedure. Alternate items may be used by the calibrating activity when the equipment listed in table 2 is not available. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the mi
22、nimum use specifications listed in table 2. When accuracies listed in table 2 provide a four-to-one accuracy ratio between the standard and TI. Where the four-to-one ratio cannot be met, the accuracy of the equipment selected is shown in parenthesis. 5. Accessories Required. The accessories listed i
23、n table 3 are issued as indicated in paragraph 4 above and are to be used in this calibration procedure. When necessary, these items may be substituted by equivalent items unless specifically prohibited. Table 2. Minimum Specifications of Equipment Required ItemCommon NameMinimum Use Specifications