Biddle MHP10_UG 电路图.pdf
《Biddle MHP10_UG 电路图.pdf》由会员分享,可在线阅读,更多相关《Biddle MHP10_UG 电路图.pdf(18页珍藏版)》请在收音机爱好者资料库上搜索。
1、Withstand Tester MHP10 User Guide MEGGER RadioFans.CN 收音机爱 好者资料库 Safety Warnings2 General Description4 Features, Controlsand Connections5 Operation Testing Precautions11 Before each testing sequence12 Carrying out a test12 Switched Low Voltage probe LP213 120 k resistance Box14 Specification15 Acces
2、sories16 Repair and Warranty17 Symbols used on the instrument Caution:Refer to User Guide Risk of Electric shock Danger of death Protective Ground (Earth) Terminal Equipment complies with current EU Directives. Contents 2 RadioFans.CN 收音机爱 好者资料库 3 SAFETY WARNINGS High Voltage testing is hazardous. S
3、afety Warnings and Precautions must be read and understood before the instrument is used. They must be observed during use. Testing must only be carried out in a dedicated, demarcated, High Voltage testing area which is clearly marked. National and local safety codes and standards must be observed.
4、For safety, the instruments must be properly grounded (earthed). A separate protective ground (earth) lead must be correctly connected from the Protective ground (Earth) terminal to a reliable ground (earth) before testing. This is in addition to the ground (earth) wire in the power (mains) lead whi
5、ch must also be connected to provide a double level of safety. Suitable external interlocks must be used. The remote control unit is not a safety interlock and must not be used as such. The instrument, HV test probe, Low voltage test lead (or LV probe) and any remotely connected equipment must not b
6、e used if any part is damaged. The HV test probe and Low Return lead (or LV probe) must be confirmed as being in good order, clean, with no broken or cracked insulation before each test sequence. Exposed metal parts, including the HV probe tip, LV probe tip, Low return lead clip and the item under t
7、est must not be touched during testing, or until completely discharged. The High Voltage probe must only be held with the fingers and thumb behind the hand guard. Replacement fuses, parts and accessories must be of the correct type and rating. NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINE
8、D AND COMPETENT PERSONS RadioFans.CN 收音机爱 好者资料库 The MHP10 is designed for general Withstand (Flash) testing of electrical equipment and insulating materials. The instrument is operated from a nominal 230 V 50/60 Hz mains supply and has a variable output voltage providing up to 3 kV a.c. The instrume
9、nt is self contained in a sheet steel cabinet, with simple to use controls on the front face of the unit. Customer terminals, accessed behind a rear panel, enable connection for external relay control, external indicator beacon and external safety interlocks. An optional remote switch facility is av
10、ailable. The instruments have been designed for indoor use only and as such should not be used in harsh conditions or outside. The instruments are designed for Installation Category II use. Installation Category II relates to transient overvoltages likely to be met on grounded (earthed) receptacles
11、(socket outlets). The instrument is supplied with a 3 wire power cord. The instrument must be used with a grounded (earthed) supply, and all 3 wires must be connected. The colour code of the cord is :- U.S.A.U.K./International Ground (Earth)GreenYellow/Green Neutral WhiteBlue Line (Phase)BlackBrown
12、Safety features include:- Interlock terminals enable automatic disconnection of the test voltage when accessing the device under test. High Voltage warning terminals provide a connection for a high visibility beacon. Two methods of control allow the user to select the best method for the type of tes
13、ting undertaken. Two independent methods of grounding (earthing), supply cord ground and an additional protective Ground (Earth) terminal located at the rear of the instrument, both of which must be used. General Description 4 Features, Controls and Connections 5 FAIL VOLTS POWER TRIP mA 0.5 3.5 10
14、20 30 40 50 HI-POT MHP10MEGGER BREAKDOWN BREAKDOWN TRIP/ TRIP STOPSTARTSTOP 0 0 2 1 34 11 9 675 8 1.Illuminated mains On / Off switch 2.Voltage control knob 3. Trip Level Switch Sets the current trip limit when testing with the trip or breakdown/trip mode. This level can be set to: 0.5 mA, 3.5 mA, 1
15、0 mA, 20 mA, 30 mA, 40 mA, 50 mA peak a.c. Warning:- The Trip level Switch does not limit the instantaneous current that can flow from the instrument prior to switchoff. Up to 12cycle delay will occur before switch off. 4.Test mode select switch There are two methods of automatically disconnecting t
16、he supply:- i) Breakdown The test voltage is applied until insulation breakdown (a sharp increase in current independent of continuous leakage current), or user intervention. ii) Breakdown / Trip As for Breakdown above, but the test voltage will also be removed automatically if the leakage current e
17、xceeds the limit set on the Trip Level Switch. iii) Trip The test voltage is automatically removed if the leakage current exceeds the limit set on the Trip Level Switch. Note:-The test voltage will only be removed automatically if the leakage current exceeds the limit set on the Trip Level Switch. 5
18、. Visual Test Fail indicator Illuminates to indicate when test conditions have been exceeded. An internal buzzer will also sound when the visual indication is active. 6. Output volt meter Used in conjunction with the Voltage control to set, and give an accurate indication of output test voltage. 10
19、7. High Voltage indicator When the instrument is generating high voltage the indicator is illuminated. The voltage on the terminals will be indicated on the voltmeter at all times. 8. High voltage a.c. output receptacle The HV receptacles used in the instrument are designed to accept a locking plug.
20、 To insert the plug, align the arrows on the plug with the open padlock symbol, push in, and then turn the plug through 90 to align the arrows with the closed padlock symbol. This will now lock the plug into the instrument. To remove, turn the plug to align the arrows with the open padlock symbol an
21、d withdraw. High Voltage probe The High Voltage probe is designed to lock into the High voltage receptacle and is fitted with a hand guard, and a spring loaded electrode cover which retracts as the probe is pressed to the item being tested. See Accessories for replacement probe. Warning:- The probe
22、must not be used in a damaged or dirty state. Do not touch any part of the HV probe in front of the guard or the device under test, until completely discharged. Warning:- Care must be taken not to touch the probe to any adjacent ungrounded (unearthed) metalwork such as pipes, window frames etc. as t
23、his may become hazardous live. 9.Low Return terminal The Low Return lead must be connected to this terminal. The terminal is internally connected to the instrument ground (earth). The Low Return lead is fitted with a hook connector so that it will not pull off the terminal. To check this lead, see B
24、efore each testing sequence. A replacement low voltage lead is available, see Accessories. 10. Start button Controls the high voltage output. Either of two modes of operation can be selected and set using a moveable link situated behind the rear panel cover:- i)The latched mode (link to the right) a