1、 PN 601655 May 1996 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A. 5700A/5720A Series II Multi-Function Calibrator Remote Programming Reference Guide RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 1 Contents Syntax Rules.3 Command Summary by Function.5 Commands.10 Operating Sta
2、te Transitions.20 Status Register Summary.21 RadioFans.CN 收音机爱 好者资料库 2 3 Syntax Rules Spaces Use at least one space between a command and a parameter. Do not use spaces within a parameter, or within a number. Multiple Parameters When a command has more than one parameter, separate the parameters wit
3、h commas. For example: OUT 1 V, 100 HZ. Numeric Parameters Parameters may have up to 255 significant figures. Exponents may range from -32000 to +32000. The useful range for 5720A programming is 2.2 E-308 to 1.8 E308. Case Sensitivity All lower case letters are converted to upper case except in quot
4、ed strings. Extra Parameters Avoid using extra parameters. Too many parameters will cause a command error. Null Parameters Null parameters will cause a command error (e.g., adjacent commas in CLOCK 133700, ,071787). Expressions Do not use expressions as parameters. An example of an expression is (4+
5、2*13). Units Accepted units in command parameters are: A amperes DB decibels DBM decibels references to 1 milliwatt for ac voltage amplitude HZ hertz OHM ohms PCT percent PPM parts per million V volts 4 Syntax Rules (continued) Multipliers MA mega, or unit x 1,000,000) K kilo, or unit x 1,000) M mil
6、li, or unit 1,000; (except MOHM and MHZ) U micro, or unit 1,000,000) Examples: MOHM or MAOHM Megohms MHZ or MAHZ Megahertz MV millivolts MAV Megavolts MA millliamperes Multiple Commands To combine commands in one statement, use a semicolon (;). For example, instead of using these two statements: 20
7、PRINT 3,OUT 100 MV ! Select 100 mV dc 30 PRINT 3,OPER ! Activate the output You can accomplish the same with this single line: 20 PRINT 3,OUT 100 MV ; OPER Terminators Terminators sent by the 5720A: LF with EOI. Terminators received by the 5720A: LF or any character with EOI Responses The responses
8、described in the command table are correct for IEEE-488 remote control, and for serial remote control in COMPUTER mode. In serial remote control in TERMINAL mode, responses contain more descriptive text intended for an operator using a terminal interactively. 5 Command Summary by Function Error Mode
9、 Commands ADJOUT? Returns adjusted output magnitude and frequency. INCR Increments or decrements the output MULT Multiplies reference by a value and establishes the product as a new reference. NEWREF Establishes a new reference. OFFSET Sets and enables or disables an offset. OFFSET? Returns the valu
10、e of the present offset. OLDREF Sets the output to the present reference. OUT_ERR? Returns the UUT error computed by the 5720A. REFOUT? Returns the value of the present reference. SCALE Activates and deactivates scaling. SCALE? Returns scaling information. SCAL_ERR? Returns the scale error value if
11、scaling is active. Instrument Configuration Commands BTYPE Selects an amplifier type for voltage or current boost. BTYPE? Returns the amplifier types set for voltage and current boost. CLOCK Sets the clock/calendar. CLOCK? Returns the setting of the clock/calendar. CUR_POST Selects active binding po
12、st for current output. CUR_POST? Returns active binding post for current output. EXTGUARD Opens and closes an internal connection. between GUARD and LO. EXTSENSE Opens and closes an internal connection between SENSE and OUTPUT. FORMAT Restores calibration constants and other parameters to factory de
13、faults. LIMIT Sets positive and negative output limits. LIMIT? Returns the programmed positive and negative output limits. 6 Command Summary by Function (continued) *LRN? Returns a list of commands that can duplicate the present state. RANGE? Returns the present output range. RANGELCK Locks the pres
14、ent range, or selects autoranging. Output Commands BOOST Activates or deactivates an auxiliary amplifier. DBMOUT? Returns the output amplitude and frequency, but in dBm if ac V. OPER Activates the 5720A output if it is in standby. OUT Sets the output and establishes a new reference for error mode. O
15、UT? Returns the output amplitude and frequency. PHASE Sets the phase shift of the phase output signal. PHASE? Returns the phase shift of the phase output signal. PHASELCK Activates or deactivates phase locking to an external signal. PHASESFT Activates or deactivates variable phase output. RCOMP Acti
16、vates or deactivates two-wire compensation circuitry. STBY Puts the 5720A in standby. VOUT? Returns the output amplitude in volts if the active units are dBm. WBAND Activates or deactivates the Wideband AC Module (Option 5700A-03). General Functions *CLS Clear; clears status registers, any service r
17、equest, and flags. ECHO? Echoes a string to the remote port. GAL Go to Alternate Language; enters Fluke 5100B or 5200A emulation mode. *OPC Sets bit 0 in the ESR to 1 when pending operations are complete. 7 Command Summary by Function (continued) *OPC? Returns a 1 when all pending remote operations
18、are complete. *PUD Stores a string of user-selected characters in non-volatile memory. *PUD? Returns the contents of the PUD memory. RPT_STR Loads the user report string. RPT_STR? Returns the user report string. *RST Resets the 5720A to its default power-up state. UNCERT? Returns the absolute uncert
19、ainty of the output. *WAI Halts program execution until pending operations are complete. RS-232C Interface Parameter Setting Commands SP_EOF Sets the End-of-File (EOF) string. SP_EOF? Returns the End-of-File (EOF) string. SP_SET Sets RS-232C interface parameters. SP_SET? Returns RS-232C interface pa
20、rameters. SP_TIME Sets the time-out period. SP_TIME? Returns the time-out period. 5720A Calibration, Testing, and Diagnostics Commands CAL_ADJ Does the internal portion of calibration. CAL_CHK Starts a calibration check. CAL_CLST? Returns a group of calibration constant names and their values. CAL_C
21、ONF Sets the specification confidence level to 95% or 99%. CAL_CONF? Returns the active calibration confidence level. CAL_CONST? Returns the value of a particular constant. CAL_DATE? Returns the date of the most recent calibration. CAL_DAYS? Returns number of days since last calibration recall. CAL_
22、INTV Sets the calibration interval. CAL_INTV? Returns the calibration interval. 8 Command Summary by Function (continued) CAL_PR Prints a calibration report out the serial interface. CAL_REF Performs a calibration procedure using a 1, 10 k, or 10V standard. CAL_RNG Starts a self calibration procedur
23、e of a range. CAL_RPT? Returns a formatted calibration report. CAL_SHIFT? Returns the shift of a range due to calibration. CAL_SLST? Returns the shifts of a group of ranges. CAL_STORE Stores new calibration constants in nonvolatile memory. CAL_TEMP Sets the temperature for calibration. CAL_TEMP? Ret