1、PN 644226 July, 1997 Rev.1, 3/99 1997,1999 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. 1 16 Multimeter with Temperature Calibration Information WWarning To avoid electrical shock, remove test leads and any input signals fr
2、om the Meter before opening the case. Caution This Meter contains parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive devices. Introduction The 16 Calibration Information instruction sheet provides the information necessary to calibrate and veri
3、fy the Fluke Model 16 Multimeter with Temperature (hereafter known as the Meter). This instruction sheet provides the following information: Product specifications Disassembly and reassembly Performance test procedures Calibration procedures and adjustments Replacement parts See the instruction shee
4、t for operating instructions. Service The Meter is warranted to be free from defects in material and workmanship for one year. The complete warranty statement is in the Meters instruction sheet. To contact Fluke, call one of the following telephone numbers: USA and Canada: 1-888-99-FLUKE (1-888-993-
5、5853) Europe: +31-402-678-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-356-5500 Or, visit Flukes Web stie at . Specifications Specifications are in Table 1. Accuracy is specified for a period of one year after calibration, at 18C to 28C (64F to 82F) with relative
6、humidity to 90%. AC conversions are ac- coupled, average responding, and calibrated to the rms value of a sine wave input. Table 1. Specifications Maximum Voltage Between any Terminal and Earth Ground (excludes 10% tolerance) 600 V ac or dc Display3-3/4-digits, 4000 counts, updates 4/sec Operating T
7、emperature-10C to 50C Storage Temperature-30C to 60C indefinitely (to -40C for 100 hrs) Temperature Coefficient0.1 x (specified accuracy)/C (28C) Relative Humidity0% to 90% (-10C to 35C) 0% to 70% (35C to 50C) Altitude Operation/Non-operation2000 m Battery Type9V, NEDA 1604 or IEC 6F22 Battery Life6
8、50 continuous hours with alkaline 450 continuous hours with carbon-zinc Shock, Vibration3 meter drops: Sinusoidal vibration up to 3 G 55 Hz Size (HxWxL)1.35 in x 2.75 in x 5.55 in (3.46 cm x 7.05 cm x 14.23 cm) Weight10 oz (286g) EMI RegulationsComplies with FCC Part 15B, Class B, EN50081-1, EN50082
9、-1. SafetyDesigned to Protection Class II requirement of UL1244, ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231, and VDE 0411, and IEC 1010-1 overvoltage category III (CAT III), 600 V CertificationP $ RadioFans.CN 收音机爱 好者资料库 16 Multimeter with Temperature 2 Accuracy specifications are given as follows: (%
10、 of reading + number of least significant digits). Table 1. Specifications (cont) FunctionRangeResolutionAccuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.0 V 400.0 V 600.0 V 1 mV 0.001 V 0.01 V 0.1 V 0.1 V 1 V (1.9%+3) (1.9%+3) (1.9%+3) NA (1.9%+3) (1.9%+3) E 4000 mV 1 4.000 V 40.00 V 300.0 V
11、 400.0 V 600 V 1 mV 0.001 V 0.01 V 0.1 V 0.1 V 1 V (0.9%+2) (0.9%+2) (0.9%+1) NA (0.9%+1) (0.9%+1) J 400.0 4.000 k 40.00 k 400.0 k 4.000 M 40.00 M 0.1 0.001 k 0.01 k 0.1 k 0.001 M 0.01 M (0.9%+2) (0.9%+1) (0.9%+1) (0.9%+1) (0.9%+1) (1.5%+3) K 1.000 F 10.00 F 100.0 F 1000 F 10000 F 0.001 F 0.01 F 0.1
12、 F 1 F 10 F (1.9%+2) (1.9%+2) (1.9%+2) (1.9%+2) (10%+90) typical 0 to 200 A0.1 A(1% + 2 counts) (50 to 400 Hz) 0 to 200 A0.1 A2% + 3 counts) Temperature 3 (Type K Thermocouple) -10C to 400C 14F to 75F -40C to -10C -40F to 14F 0.1C 0.2F 0.1C 0.2F (1% + 0.8C) (1% + 1.5F) (5% + 1.5C) Typical (5% + 3.3F
13、) Typical C 2 2.000 V0.001 V(0.9%+2) 1. The 4000 mV range can be entered only in the manual range mode. Use the 4000 mV range with accessories. 2. The beeper is guaranteed to come on at 250. The Meter detects opens or shorts of 250 s or longer. 3. Error does not include Type K Thermocouple errors. R
14、adioFans.CN 收音机爱 好者资料库 Specifications 3 Table 1. Specifications (cont) FunctionOverload Protection 1 Input Impedance (Nominal) Common Mode Rejection Ratio (1 k Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc 5 M 2 k 60 dB at dc 50 Hz or 60 Hz E 600 V dc 10 M 2 k 100 dB at dc, 50 Hz or 60 Hz 50 dB a
15、t 50 Hz or 60 Hz Open Circuit Test Voltage Full Scale Voltage To 4.0 M 40 M Short Circuit Current J 600 V dc1.5 V dc450 mV dc1.5 V dc500 A L 2 600 V dc2.4-3.0 V dc2.400 V dc0.95 mA (typical) Burden VoltageCommon Mode Rejection Ratio Normal Mode Rejection 600 V rms or dc100 dB dc 50 Hz or 60 Hz 50 dB
16、 50 Hz or 60 Hz 600 V rms or dc60 dB at dc, 50 Hz or 60 Hz . 1. 3 x 10 6 V Hz maximum. 2. 2 k input impedance up to 50 V. Impedance increases with input voltage to 300 k at 600 V. MIN MAX Recording Accuracy and Response Time Specified accuracy of the measurement function 12 digits in dc for changes
17、200 ms in duration (40 digits in ac). Typical 100 ms response to 80%. Example 1: this would mean 1.2 when recording temperature. Example 2: This would mean 12 A when recording A or 12 A if used with a dc amp probe. MIN MAX Recording with Elapsed Time Elapsed Time: 0 to 100 hours (99.59) Resolution:
18、1 minute Accuracy: 0.3% typical Continuity Capture Detects opens or shorts of 250 s or longer. RadioFans.CN 收音机爱 好者资料库 16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. A Phillips-head screwdriver and small flat-blade screwdriver are
19、 required. 1. Remove the test leads and set the rotary knob to OFF. 2. Remove the Phillips-head screws from the case bottom. 3. Separate the case top from the case bottom. 4. To replace the battery: lift the battery from the case bottom and insert a new 9V battery. Be sure the positive and negative
20、battery posts are oriented correctly. 5. To remove the pca: insert a small, flat-blade screwdriver between the edge of the case top and the pca where shown in Figure 2. Gently unsnap a side of the case top from the pca. Repeat on the other side of the pca. Unsnap the case from the top of the pca las
21、t. Lift the pca from the case top by its edges. If the elastomeric contact strips for the switch assembly and LCD are stuck to the pca, remove them without touching the conductive edges. Note Before reinserting the pca, make sure that the rotary knob is in the OFF position. 7. To reinsert the pca: P
22、lace the pca over the four screw posts in the case top, then press gently on the center of the pca while using the small flat-edge screwdriver to shoehorn the pca under the snap on a side of the case top. Repeat on the other side and the top. 8. To remove the switch support, use a small, flat-blade
23、screwdriver to gently unsnap the sides and top of the switch support from the snaps shown in Figure 2. 9. The LCD, switch assembly, and elastomeric contact strips for the LCD and switch assembly are accessible and can be replaced as needed. Do not allow the LCD to get wet. Before installing a new LC
24、D, make sure that all connector contact points are clean. Caution Do not touch the conductive edges of the elastomeric strips or the contacts on the switch assembly. If they are contaminated, clean them with isopropyl alcohol. 9. Reassembling the Meter is the reverse of disassembling it. After the M