FLUKE 45 电路图.pdf
《FLUKE 45 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 45 电路图.pdf(184页珍藏版)》请在收音机爱好者资料库上搜索。
1、 45 Dual Display Multimeter Service Manual For IEC 1010 Meters Only PN 609203 March 1999 Rev.1, 1/04 1999-2004 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 LIMITED WARRANTY effective troubleshooting
2、may require not only reference to the troubleshooting procedures in Chapter 5, but also some understanding of the detailed Theory of Operation in Chapter 2 and some tracing of circuit operation in the Schematic Diagrams presented in Chapter 9. Often, scanning the table of contents will yield an appr
3、opriate place to start using the manual. A comprehensive table of contents is presented at the front of the manual; local tables of contents are also presented at the beginning of each chapter for ease of reference. If you know the topic name, the index at the end of the manual is probably a good pl
4、ace to start. The following chapter descriptions serve to introduce the manual: Chapter 1. Introduction and Specifications Introduces the Fluke 45 Dual Display Multimeter, describing its features, options, and accessories. This chapter also discusses use of the Service Manual and the various convent
5、ions used in describing the meters circuitry. Finally, a complete set of specifications is presented. Introduction and Specifications Conventions1 1-5 Chapter 2. Theory of Operation This chapter first categorizes meter circuitry into functional blocks, with a description of each blocks role in overa
6、ll operation. A detailed circuit description is then given for each block. These descriptions explore operation to the component level and fully support troubleshooting procedures defined in Chapter 5. Chapter 3. General Maintenance Provides maintenance information covering handling, cleaning, and f
7、use replacement. Access and reassembly procedures are also explained in this chapter. Chapter 4. Performance Testing and Calibration This chapter provides performance verification procedures that are tied to the specifications presented in Chapter 1. To maintain these specifications, a full calibrat
8、ion procedure is also presented. Chapter 5. Diagnostic Testing and Troubleshooting The troubleshooting procedures presented in this chapter rely closely on both the Theory of Operation presented in Chapter 2 and the Schematic Diagrams shown in Chapter 9. Chapter 3 provides access information. Chapte
9、r 6. List of Replaceable Parts Includes parts lists for all standard assemblies. Information on how and where to order parts is also provided. Chapter 7. Option -01 Battery Pack Each option is allocated a separate chapter: 7 for the Battery Pack Option -01 and 8 for the IEEE-488 Interface Option -05
10、. Option -15 incorporates both Options -01 and -05. Chapter 7 includes the full range of Service Manual topics (specifications, theory of operation, maintenance, list of replaceable parts, etc.) for the Battery Pack option. Schematic diagrams for the options are found in Chapter 9. Chapter 8. Option
11、 -05 IEEE-488 Interface Includes the full range of Service Manual topics (specifications, theory of operation, maintenance, list of replaceable parts, etc.) for the IEEE-488 Interface option. Schematic diagrams for the options are found in Chapter 9. Chapter 9. Schematic Diagrams Includes schematic
12、diagrams for all standard and optional assemblies. A list of mnemonic definitions is also included to aid in identifying signal name abbreviations. 1-5. Conventions Throughout the manual set, certain notational conventions are used. A summary of these conventions follows: Instrument Reference The Fl
13、uke 45 Dual Display Multimeter is usually called the meter. Printed Circuit Assembly 45 Service Manual 1-6 The term pca is used to represent a printed circuit assembly and its attached parts. Signal Logic Polarity On schematic diagrams, a signal name followed by a * is active (or asserted) low. Sign
14、als not so marked are active high. Circuit Nodes Individual pins or connections on a component are specified with a dash (-) following the component reference designator. For example, pin 19 of U30 would be U30-19. User Notation For front panel operation, XXXAn uppercase word or symbol without paren
15、theses indicates a button to be pressed by the user. Buttons can be pressed in four ways: 1. Press a single button to select a function or operation. 2. Press a combination of buttons, one after the other. 3. Press and hold down a button, then press another button. 4. Press multiple buttons simultan
16、eously. For computer interface operation, XXXAn uppercase word without parentheses identifies a command by name. Angle brackets around all uppercase letters mean press the key. (xxx)When associated with a keyword, a lowercase word in parentheses indicates an input required by the user. 1-6. Specific
17、ations The following contains the specifications for the Fluke 45 Dual Display Multimeter. These specifications assume: A 1-year calibration cycle An operating temperature of 18 to 28C Relative humidity not exceeding 90% (non-condensing) Accuracy is expressed as (percentage of reading + counts). Rea
18、ding Rates and Display Counts RateReadings per SecondFull Range Display Counts Slow Medium Fast 2.5 5 20 99,999* 30,000 3,000 * Ohms full range will typically be 98,000 counts Introduction and Specifications Specifications1 1-7 Response Times Refer to Chapter 4 of the Users Manual for detailed infor
19、mation. DC Voltage ResolutionAccuracy Range SlowMediumFast(6 months)(1 year) 300 mV 3 V 30 V 300 V 1000 V - - - - - 10 V 100 V 1 mV 10 mV 100 mV 100 V 1 mV 10 mV 100 mV 1 V 0.02% + 2 0.02% + 2 0.02% + 2 0.02% + 2 0.02% + 2 0.025%+2 0.025% +2 0.025%+2 0.025% +2 0.025%+2 100 mV 1000 mV 10 V 100 V 1000
20、 V 1 V 10 V 100 V 1 mV 10 mV - - - - - - - - - - 0.02% + 6 0.02% + 6 0.02% + 6 0.02% + 6 0.02% + 6 0.025%+6 0.025% +6 0.025%+6 0.025% +6 0.025%+6 Input Impedance 10 M in parallel with 80 dB at 50 or 60 Hz, slow and medium rates 54 dB for frequencies between 50-440 Hz, slow and medium rates 60 dB at
21、50 Hz, fast rate (Note: Fast rate has no filtering) Maximum Allowable AC Voltage While Measuring DC Voltage Peak Normal Mode Signal RangeMax Allowable Peak AC Voltage NMRR* 80dB*NMRR 60 dB* 300 mV100 mV20 V15 V15 V 3 V1000 mV20 V15 V15 V 30 V10 V1000 V50 V300 V 300 V100 V1000 V50 V300 V 1000 V1000 V
22、1000 V200 V1000 V * NMRR is the Normal Mode Rejection Ratio * Normal Mode Rejection Ratio at 50 or 60 Hz 0.1% Common Mode Rejection Ratio 90 dB at dc, 50 or 60 Hz, (1 k unbalanced, medium and slow rates) 45 Service Manual 1-8 Maximum Functional Input 1000 V dc or peak ac on any range True Rms AC Vol
23、tage, AC-Coupled Range Resolution SlowMediumFast 300 mV-10 V100 V 3 V-100 V1 mV 30 V-1 mV10 mV 300 V-10 mV100 mV 750 V-100 mV1 V 100 mV1 V- 1000 mV10 V- 10 V100 V- 100 V1 mV- 750 V10 mV- Accuracy FrequencyLinear AccuracydB AccuracyPower* Max Input at Upper Freq SlowMediumFastSlow/MedFast 20-50 Hz1%
24、+ 1001% + 107% + 20.150.722% + 10750 V 50 Hz -10 kHz0.2% + 1000.2% + 100.5% + 2 0.080.170.4% + 10 750 V 10 - 20 kHz0.5% + 1000.5% + 100.5% + 2 0.110.171% + 10750 V 20 - 50 kHz2% + 2002% + 202% + 30.290.344% + 20400 V 50 - 100 kHz5% + 5005% + 505% + 60.700.7810% + 50200 V * Error in power mode will n