FLUKE pm6685r 电路图.pdf
《FLUKE pm6685r 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE pm6685r 电路图.pdf(94页珍藏版)》请在收音机爱好者资料库上搜索。
1、Programmable Frequency Counter PM6685 rather, it is concerned primarily with those parts of the instrument which are essential for determining the function of the instrument. It is not necessary to remove the cover of the instrument to perform this procedure. If the test is started less than 20 minu
2、tes after turning on the instru- ment, results may be out of specification, due to insufficient warm-up time. Recommended Test Equipment *) Two of the cables must have 10 ns difference in delay, for ex- ample: 5 ns and 15 ns. Preparations Power up your instruments at least 20 minutes be- fore beginn
3、ing the tests to let them reach normal operating temperature. Failure to do so may result in certain test steps not meeting equipment specifi- cations. Front Panel Controls Power-On Test At power-on the counter performs an automatic self-test of the fol- lowing: Microprocessor RAM ROM Measuring circ
4、uits Display If a GPIB interface is installed, the GPIB address is displayed. If there are any test failures, an error message is shown. Turn on the counter and check that all segments light up on the display and that no error message appears. Internal Self-Tests The different built-in test routines
5、 invoked by the power-on test can also be activated from the front panel as follows: Enter the Auxiliary Menu by pressing AUX MENU. Select the test submenu by pressing DATA ENTRY up or down. Enter the test menu by pressing the ENTER key. Selections for internalself-tests are: 1TEST ALL (Test 2 to 5
6、in sequence) 2TEST RO (ROM) 3TEST RA (RAM) 4TEST LOGIC (Measuring Logic) 5TEST DISP (Display Test) Use DATA ENTRY up/down to select TEST ALL, then press ENTER. If any fault is detected, an error message appears on the dis- play and the program halts. If no faults are detected, the program returns to
7、 measuring mode. 2-2Performance Check, General Information Type of instru- ment Required Specifications Suggested Equipment LF SynthesizerSquare; Sine up to 10 MHz Power Splitter50 ?PM9584/02 T-piece Termination50 ?PM9585 Reference oscilla- tor 10 MHz 0.1 Hz for standard oscillator Fluke counter wit
8、h calibrated option PM9691 10 MHz 0.01 Hz for PM9691 press menu again. Short Form Specification Test Sensitivity and Frequency Range Press the PRESETkey to set the counter in the default setting. Then confirm by pressing ENTER. Turn off AUTO. Select IMP A = 50 ? and maximum sensitivity. Connect a si
9、gnal from a HF generator to a BNC power splitter. Connect the power splitter to your counter and an oscilloscope. Set input impedance to 50 ? on the oscilloscope. Adjust the amplitude according to the following table. Read the level on the oscilloscope. The counter should display the correct frequen
10、cy. Reference Oscillators X-tal oscillators are affected by a number of external conditions, suchasambienttemperatureandsupplyvoltage,buttheyarealsoaf- fected by aging. Therefore, it is hard to give limits for the allowed frequency deviation. You must decide the limits depending on your application,
11、 and recalibrate the oscillator accordingly. See the Pre- ventive Maintenance in the Repair chapter, Chapter 5. To check the accuracy of the oscillator you must have a calibrated reference signal that is at least five times as stable as the oscillator that you are testing, see the following table. P
12、ress the PRESETkey, then press the ENTER key to set your counter in the Default setting. Performance Check, Short Form Specification Test2-3 Key(s)DisplayNotePass /Fail STAND-BYDisplay OffRed LED beside the key On ONBacklight on PRESET ENTER DEFAULT? NO SIGNAL Default setting EXT REFEXT REF Input A
13、FILTERFILTER 50? 50 ? (2 times) SENS ? (2 times) Bar graph: ? SENS ? (2 times) Bar graph: ? AUTOAUTO TRIG Other PRESET ENTER DEFAULT? NO SIGNAL Default setting MEAS TIME2003s DATA ENTRY ?5003s DATA ENTRY ?2003s ENTERNO SIGNAL DISPLAY HOLDHOLD DISPLAY HOLD SINGLESINGLE FUNCTION ?DUTY F A FUNCTION ?TO
14、T A MAN FUNCTION ?DUTY F A FUNCTION ?FREQ A AUX MENURECALL MEAS RESTARTNO SIGNAL PRESET ENTER DEFAULT? NO SIGNAL Default setting CHECK10.000000006Hz*Start counting NULLNULL NULL10.000000006Hz* BLANK DIGITS (3 times) 10.00000? ? ?6Hz* MENUDisplays all avail- able functions, pro- cesses and input cont
15、rols. Selected items are blinking. Table 2-2Keyboard Test. FrequencyLevelPass/Fail MHzmVPPmVRMSdBmInput A 1301027 25301027 50301027 150602021 200903017 25015050-13 30015050-13 Table 2-3Sensitivity of input A at various frequencies. OscillatorMax. tempera- ture dependence Max. aging per month Max. ag
16、ing per year Standard100 Hz5 Hz50 Hz PM96910.05 Hz0.1 Hz0.75 Hz PM96920.025 Hz0.03 Hz0.2 Rubidium0.003 Hz0.0005 Hz0.002 Hz Table 2-4Deviation (for PM9691 and PM9692 after a warm-up period of 48 hours). Connect the reference to input A. Check the readout against the accuracy requirements of your appl
17、ication. ?Acceptance Test As an acceptance test, the following table gives a worst case figure after a 30 minute warm up time. All deviations that can occur in a year are added together. ?Acceptance Test, PM6685R To fully test the accuracy of the PM6685R, a reference signal of ex- tremely high stabi
18、lity is needed. Examples of such references are Cesium Atomic references, or transmitted signals from a nationally or internationally traceable source, like the GPS satellites. Recommended Test Equipment TypeStabilityModel 10 MHz refer- ence ? 1x10-10910R with satellite contact dur- ing the last 72
19、hours. Test Procedure Connect the counter to the line power. Check that the UNLOCK indicator turns on, and then turns off again within 6 minutes after connecting line power. Connect the 10 MHz reference signal to input A of the counter. Select FREQUENCY A measurement. Select 2 s measuring time. Chec
20、k that the displayed frequency is 10.00000000 MHz 0.05 Hz ? 10 minutes after connection to line power. Rear Input/Output INT REF Output Connect an oscilloscope to the 10 MHz output on the rear of the counter. Use coaxial cable and 50 ? termination. The output voltage is sinusoidal and should be abov
21、e 2.8 VPP. EXT REF Input Press the PRESET key, then press the ENTER key to set your counter in the Default setting. Apply 10 MHz sine to input A equipped with a T-piece and to Ext Ref input at the rear, terminated with 50 ?. Amplitude on 10 MHz signal; 200 mVRMS, (560 mVPP) Press the Ext Ref key. Th
22、e display should show 10.000000006Hz 5 LSD. EXT ARM INPUT Press the PRESETkey, then press the ENTER key to set your counter in the Default setting. Select 50 ?input impedance. Apply 10 MHz 500 mVRMS, (1.4 VPP) sine to input A The counter measures and displays 10 MHz. Press the AUX MENU key. Press th
23、e DATA ENTRY UP/DOWN keys until the display shows Ar. Start, confirm by pressing the ENTER key. Press DATA ENTRY UP/DOWN keys until the display shows POS, confirm by pressing the ENTER key. Press the ENTER key once more. The counter does not measure. Connect a pulse generator to Ext Arm input. Setti
24、ngs for pulse generator: single shot pulse, amplitude TTL = 0 - 2 VPP, and duration = 10 ns. Apply one single pulse to Ext Arm input. The counter measures once and shows 10 MHz on the display. Measuring Functions Preparation for Check of Measuring Function is as follows: Connect a 10 MHz sine wave s