1、PN 735985 February 1985 Rev.6, 11/02 1985-2002 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companies. Introduction The Fluke 80TK is a thermocouple converter for use with any K-type thermocouple and a digital multimeter. The 80TK accept
2、s the output of any K-type thermocouple and converts it to 1 millivolt per degree (Celsius or Fahrenheit). A three-position switch acts as a power switch as well as selecting Fahrenheit or Celsius scaling for the output. In addition, the OFF position of the power switch allows you to read the batter
3、y condition via the external multimeter. The thermocouple input accepts wire leads or standard mini- thermocouple connectors. Temperature is measured by exposing or attaching the thermocouple to the surface to be measured. The multimeter displays the temperature directly in degrees Fahrenheit or Cel
4、sius. Specifications Electrical Measurement Range: -50 to 1000 degrees Celsius -58 to 1832 degrees Fahrenheit Resolution: 2000 count display:0.1 C/F in 200 mV range 1.0 C/F in 2V range 3200 count display:0.1 C/F in 300 mV range 1.0 C/F in 3V range 4000 count display:0.1 C/F in 400mV range 1.0 C/F in
5、 4V range Note: Ignore decimal point on 2V, 3V, or 4V range. Maximum Voltage at Thermocouple Input: 60V DC, 24V AC RF Field: Exposure to an RF field interferes significantly with this modules capacity to accurately measure temperature with thermocouples. The interference is temporary and the module
6、suffers no harm when exposed to an RF field of 3 V/m. Accurate temperature measurement capability is completely restored when the module is removed from the RF field. Common sources of RF fields are handheld 2-way radios and cellular telephones. If such a source is suspected of interfering with this
7、 module, either turn off the transmitter or increase the separation between the transmitter and the module. Environmental Ambient Operating Range: 0 to 50C Storage Temperature: -40 to 60C -40 to 140F Humidity:0% to 90% (0 to 35C) 0% to 70% (35 to 50C) Basic Accuracy: ( 23 5C Calibration)* (Applies f
8、or one year after purchase or recalibration) RANGEACCURACY (% of reading + degrees), 1 year -20 to 350C0.5% 2C -4 to 662F0.5% 3.6F 351 to 500C1.75% 2C 664 to 932F*1.75% 3.6F 501 to 1000C2.0% 2C 933 to 1832F2.0% 3.6F -50 to -21C2.5% 2C typical -58 to -5.8F2.5% 3.6F typical Temperature Coefficient: *0
9、C to 18C and 28C to 50C ambient, multiply the basic accuracy specification by 0.1 for each degree above 28 or below 18. For accuracy enhancement between 351C to 500C, subtract 3 degrees from the reading. The accuracy is now 0.75% 2C. * For accuracy enhancement between 663F and 932F, subtract 5.4 deg
10、rees from the reading. The accuracy is now 0.75% 3.6F. Note The basic accuracy specification does not include the error of the probe or the DMM. Please refer to the probe accuracy specification for additional details. General Weight: 4.5 ounces, 126 grams Overall length: 4 5/8 inches, 11.75 mm Batte
11、ry: Standard 9V Battery (NEDA #1604, 6F22, 006P) Battery Life: 1600+ hours, 6.5V minimum Output: 3/4 inch spaced banana plugs, 10 megohm minimum load resistance Accessories 80PK-1 (included): Bead Probe 4 foot K-type thermocouple bead probe, with Teflon tape insulation. Maximum insulation temperatur
12、e: 260C. Accuracy: 2.2C or 0.75% (whichever is greater) from 0 to 800C. 80PK-2A: Immersion Probe (8 or 20.32 cm from tip to probe handle) 80PK-3A: Surface Probe (for flat or curved surfaces) 80PK-4A: Air Probe (protected by stainless steel baffle) 80PK-5A: Piercing Probe (suitable for food use, liqu
13、ids and gels) 80PK-6A: Exposed Junction Probe (exposed bead attached to 6 sheath) 80TK Thermocouple Module Instruction Sheet RadioFans.CN 收音机爱 好者资料库 . 80PK-7: Industrial Surface Probe 80PK-8 Pipe Clamp Probe (For pipe diameter 1/4” to 1 7/8”) 80CK-M: Male Mini-Connector (for attachment of K-type the
14、rmocouple wire) Use and Operation Thermocouples rely upon the difference in temperature between the measuring junction and a reference junction. Traditionally the reference junction is another thermocouple placed in an ice bath, however, the 80TK uses an electronic reference junction. The converter
15、(80TK) must be operated in a thermal environment consistent with its specifications. Measuring Technique Making accurate temperature readings using a thermocouple is a matter of identifying and minimizing the causes of error. The largest contributor to errors is the thermal interface between the mea
16、surement surface and the thermocouple. The next largest contributor is the temperature difference between the thermocouple probe and its environment. Here are some suggestions for improving the accuracy of your temperature measurements. Ensure that there is a good connection between the probe and th
17、e surface you are measuring. You can do this in many ways: 1. Increase the mounting pressure. 2. Use an interface agent (such as silicone grease) between the thermocouple probe and the surface you are measuring. 3. Use a thermal epoxy to attach the thermocouple. When measuring temperatures higher th
18、an ambient, adjust the connection of the probe to the surface until you get the highest temperature reading. Use any of the suggestions given above to do so. When measuring temperatures lower than ambient, adjust the connection of the probe to the surface until you get the lowest temperature reading
19、. Use any of the suggestions given above to do so. When measuring temperatures near ambient, make the reading when the multimeter display is most stable. Warning To avoid electrical shock, do not use this instrument when voltages exceeding 24V ac or 60V dc are present. The probe tip is electrically
20、connected to the output terminals. Temperature Measurement Follow this procedure to correctly use the 80TK. 1. Plug the 80TK into the multimeter. Ensure that the banana plug with the red dot above it plugs into the voltage input and the other plug is in the COM input. 2. Set the multimeter to the 20
21、0, 300, or 400 millivolt range. This setting will read temperatures equal to the number of millivolts the range has (i.e., 320C or F if the millivolt range reads 320 mV. 3. Use the dc volt setting and ignore the decimal for higher temperatures than can be accomodated by the millivolt range. Less res
22、olution is available when using the volt range. 4. Plug the thermocouple into the 80TK thermocouple input connector. If the thermocouple does not have a plug attached, connect the thermocouple wires to the thermocouple input screws inside the 80TK. The yellow coded wire (chromel) connects to the scr
23、ew marked ”+”. 5. Turn on the 80TK and the multimeter. 6. Adjust the thermal connection between the measurement surface and the thermocouple as described previously. 7. Read the temperature on the multimeter. Open Thermocouple Reading If the thermocouple probe has open circuited (i.e. is broken), th
24、e mulimeter will read a steady, constant voltage. The voltage readings are: RangeReading C 1.0V F 2.2V Battery Condition Test 1. Set the power switch to the OFF position. 2. Connect the 80TK to the multimeter. 3. Set the multimeter to the 200, 300, or 400 mV dc range. 4. Read the battery test voltag