FLUKE 5700A, 5720A Operator 电路图.pdf
《FLUKE 5700A, 5720A Operator 电路图.pdf》由会员分享,可在线阅读,更多相关《FLUKE 5700A, 5720A Operator 电路图.pdf(40页珍藏版)》请在收音机爱好者资料库上搜索。
1、PN 601648 May 1996 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A. 5700A/5720A Series II Multi-Function Calibrator Operator Guide RadioFans.CN 收音机爱 好者资料库 RadioFans.CN 收音机爱 好者资料库 1 Contents What is in this Guide? . 2 Safety Summary. 2 P Fuse and Line Voltage . 4 Basic Operation. 5
2、 A Short Calibration Exercise. 5 Warming up the Calibrator . 5 Exercise: Running dc Zero Calibration. 6 Connecting a Meter. 6 Applying a dc Voltage . 7 Checking the Calibrator Uncertainty . 7 Exercise: Activating Error Mode . 8 Checking the Meters Higher and Lower Ranges. 8 Exercise: Applying an ac
3、Voltage . 9 Front Panel Features. 10 Output Display (Left Side) . 10 Control Display (Right Side). 10 Display Screen Saver. 11 Front Panel Keys . 12 Front Panel Connectors . 21 Rear Panel Features . 23 Rear Panel Connectors. 23 Switches and Fuse Holder. 26 Cable Recommendations. 27 Connecting to a U
4、UT (Unit Under Test) . 28 Using Error Mode . 35 Operating in Remote . 37 RS-232-C Interface Parameters. 38 RadioFans.CN 收音机爱 好者资料库 2 What is in this Guide? The Operator Guide begins with safety information, a short meter calibration exercise for new users, then continues with a condensation of infor
5、mation from the Operator Manual. For complete information about the Calibrators features, functions, and operating procedures, refer to the Operator Manual. Safety Summary Warning Lethal voltage may be present on the terminals. Observe all safety precautions in this guide. To avoid electrical shock
6、hazard, the operator should not electrically contact the OUTPUT HI or SENSE HI binding posts. During operation, lethal voltages of up to 1100V ac or dc may be present on these terminals. Whenever possible, keep one hand away from the equipment to reduce the hazard of current flowing through vital or
7、gans of the body. Symbols Marked on Equipment Warning - Risk of electric shock. QGround Protective ground (earth) terminal. P Attention Refer to the manual for information about this feature. This symbol appears on the rear panel ground terminal and by the fuse holder. Use the Proper Fuse To avoid f
8、ire hazard, use only the fuse specified on the line voltage selection switch label. 3 Safety Summary (cont.) Grounding the Calibrator The Calibrator is a Safety Class I (grounded enclosure) instrument. The enclosure is grounded through the grounding conductor of the power cord. To avoid electrical s
9、hock, plug the power cord into a properly wired earth grounded receptacle before making any connections to the Calibrator terminals. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation. Do Not Remove the Cover To avoid personal injury, d
10、o not remove the Calibrator cover. Do not operate the Calibrator without the cover properly installed. There are no user-serviceable parts inside the Calibrator, so there is no need to ever remove the cover. 4 P Fuse and Line Voltage The correct fuse type for each voltage selection is indicated on t
11、he rear panel as shown in the following figure. VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 100V S2S3S4 110V 115V 120V VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 200V S2S3S4 220V 230V 240V CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING. VOLTAGE SELECTION FUSE-F1 T 125A
12、250V (SB) 100V S2S3S4 110V 115V 120V VOLTAGE SELECTION FUSE-F1 T 125A 250V (SB) 200V S2S3S4 220V 230V 240V CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING. Line Power Label and Switch Locations 5 Basic Operation To set the output, simply press the following sequence of
13、keys to select an output function and amplitude: numeric keys multiplier function E O For example, to set the output to 10 mV dc, press: 1 0 m V E O To set an ac output, press the following additional keys: numeric keys multiplier H E For example, to change the present 10 mV dc to 10 mV ac 1.8 kHz,
14、press: 1 . 8 K H E To change the output back to dc, press: 0 H E or + E A Short Calibration Exercise If you are not familiar with the Calibrator, you will find that the following exercise quickly teaches you some important front panel operations. Warming up the Calibrator When you first turn the Cal
15、ibrator on, you should let it warm up for 30 minutes. This ensures that the Calibrator meets or exceeds its specifications. 6 Exercise: Running dc Zero Calibration DC Zero is a brief internal process that removes offset errors. The specifications require that you run dc Zero at least every 30 days.
16、To run dc Zero, press the following sequence of softkeys: Setup Menus Cal Zero Press any key, then press P twice to return to normal operation. Connecting a Meter Typical connections for handheld and benchtop meters are shown on the next page. The diagrams show the use of Fluke Model 5440-7002 Low T
17、hermal Cables. Other types of cable may be used for this exercise. To connect either type of meter, proceed as follows: 1. Verify that the Calibrator is on, and is in standby mode (STANDBY indicator lit). If the Calibrator is not in standby, press r. 2. Set the meter to the lowest range that reads 1
18、0V dc. 3. Connect the shield lead (if your cables are shielded) to V GUARD on the Calibrator. 4. Verify that both X and x are off. Press them if they are lit. 5. Connect the meter input LO to the Calibrator OUTPUT LO. 6. Connect the meter input HI to the Calibrator OUTPUT HI. 7 Exercise: Connecting
19、a Meter (cont.) HI HI LO LO HI AUX CURRENT GUARD GROUND WIDEBAND Calibrator 10A 300mA V COM DMM NC EX SNS : OFF EX GRD : OFF OUTPUT V A SENSE V HI HI LO LO HI OUTPUT V A SENSE V AUX CURRENT GUARDGROUND WIDEBAND Calibrator DMM INPUT NC EX SNS : OFF EX GRD : OFF Applying a dc Voltage Apply 10V dc to t
20、he meter by pressing the following sequence of keys: 1 0 V E O Checking the Calibrator Uncertainty Press s. The total uncertainty for the 10V you are applying to the meter is displayed on the Control Display (the large dot-matrix display on the right side of the front panel). This number depends on
21、the setting of the calibration cycle and specification confidence level in a setup menu. Press P to clear the uncertainty display. 8 Exercise: Activating Error Mode Chances are the meter reads something other than exactly 10V. To quickly check the error of the meter, turn the rotary knob to obtain a
22、 meter reading of exactly 10V, and read the error off the Control Display. Turning the knob is all it takes to activate error mode. When you turn the knob, the least significant digit is highlighted. This digit changes as you turn the knob. To change a higher order digit, thereby speeding up the adj
23、ustment rate, press . The error on the Control Display is a combination of offset, scale, and linearity errors. You can view component parts of the error by pressing o and S. Checking the Meters Higher and Lower Ranges To check the meter error at 100V dc, first set the meter range (if applicable) to