1、 5700A/5720A Series II Multi-Function Calibrator Getting Started PN 1668111 March 2002 Rev. 2, 10/07 2002-2007 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies. RadioFans.CN 收音机爱 好者资料库 LIMITED WARRANTY the display shows linearity,
2、offset, and scale errors Keys that multiply and divide the output value by 10 to simplify work on meters with calibration points at decade multiples of a fraction of full-scale Programmable entry limits used for restricting the levels that can be keyed into the calibrator, preventing access to level
3、s that may be harmful to equipment or personnel A s key that provides the capability of displaying the instruments specification at the selected operating point, calibration interval, and specification confidence level 5700A/5720A Series II Getting Started 2 An auxiliary current binding post that al
4、lows you to calibrate meters with separate current inputs without moving cables Real-time clock and calendar for date stamping reports Offset and scaling modes that simplify linearity testing of multimeters Variable phase reference signal output and phase-lock input Interface for the Fluke 5725A Amp
5、lifier Standard IEEE-488 (GPIB) interface, complying with ANSI/IEEE Standards 488.1-1987 and 488.2-1987 Selectable normal remote mode or emulation of the Fluke 5100B and 5200A Series calibrators in functions and response to system controller software EIA Standard RS-232C serial data interface for pr
6、inting, displaying, or transferring internally-stored calibration constants, and for remote control of the calibrator Extensive internal self-testing and diagnostics of analog and digital functions A traceable calibration procedure for all modes and ranges that requires only 10 V, 1 , and 10 k exter
7、nal standards, with only occasional independent verification Fast, simple, automated calibration check providing added confidence between calibration recalls, and data that can be used to document and characterize the calibrators performance between calibration recalls Instruction Manuals The 5700A/
8、5720A Series II Calibrators ship with a complete manual set that contains information for the operator and service or maintenance technicians. The set includes: 5700A/5720A Series II Getting Started Manual (PN 1668111) 5700A/5720A Series II Operator Reference Guide (PN 601648) 5700A/5720A Series II
9、Remote Programming Reference Guide (PN 601655) 5700A/5720A Series II Operator Manual (provided on CD-ROM, PN 1668127, or a printed copy is available for purchase through the Fluke Service Department under PN 601622) 5700A/5720A Series II Service Manual (provided on CD-ROM, PN 1668127, or a printed c
10、opy is available for purchase through the Fluke Service Department under PN 105798) Order additional copies of these instruction manuals separately using the part numbers provided. For ordering instructions, refer to the Fluke Catalog or contact a Fluke sales representative. Getting Started Wideband
11、 AC Voltage Module (Option 5700A-03) 3 Wideband AC Voltage Module (Option 5700A-03) The Wideband AC Voltage Module (Option 5700A-03) can be installed in both the 5700A and 5720A Series II Calibrators. The module is a high-accuracy, low-noise, extremely flat ac voltage source for calibrating rf voltm
12、eters, with a frequency range of 10 Hz to 30 MHz. Output is in seven ranges from 300 V (-57 dBm) to 3.5 V (+24 dBm) through a Type-N coaxial connector into a 50 load. The output level is selected in volts or dBm through either the front panel controls or under remote control. The wideband module als
13、o functions with the calibrators output adjust controls that let display the error of a wideband meter in either percentage of output or in decibels. Included with the wideband module is a Type-N output cable and a 50 terminator. The wideband module is calibrated to the end of its standard-equipment
14、 output cable. Auxiliary Amplifier The Fluke Model 5725A Amplifier is available to extend the high voltage performance and current range of the calibrator: Interface connectors on the calibrators rear panel accept cables to directly operate a 5725A. Three amplifiers can be connected to the calibrato
15、r at the same time, but only one output can be active at a time. Once you have connected the amplifiers and configured the calibrator in a setup menu, amplifier operation is controlled by the calibrator. Operating instructions for the 5725A are provided in Chapter 4 of the Operators Manual. The Gene
16、ral Specifications include specifications for operating the calibrator with the 5725A. For other amplifier specifications, refer to their instruction manuals. Table 1 summarizes the extended capabilities offered by the 5725A. Brief descriptions of the extended capabilities follow. Table 1. Auxiliary
17、 Amplifier Data Model Mode Range 5725A Amplifier AC V 20 to 1100 V rms up to 70 mA, 40 Hz to 30 kHz (50 mA 5 kHz) 220 to 750 V rms up to 70 mA, 30 kHz to 100 kHz DC Amps 0 to 11 A DC Amps 1 to 11 A rms, 40 Hz to 10 kHz 5725A Amplifier The Fluke 5725A Amplifier is an external unit operating under cal
18、ibrator control to extend ac voltage drive capabilities and both ac and dc current output range. The amplifier adds the following capabilities to the calibrators 1100 V AC range with no compromise in accuracy: Frequency limits at higher voltage increase to 100 kHz at 750 V, 30 kHz at 1100 V. Load li
19、mit increases to 70 mA for frequencies above 5 kHz. Capacitive drive increases to 1000 pF, subject to the maximum output current. Extended-performance voltage is available at the calibrators front or rear binding posts, eliminating the need to change cables during a procedure. 5700A/5720A Series II
20、Getting Started 4 A separate set of binding posts on the front panel of the 5725A supplies extended-range ac and dc current outputs. Since most meters have a separate input terminal for the high current ranges, this eliminates the need to change cables during a procedure. The 5725A can also be confi
21、gured to source all current (both standard calibrator-generated current and its own current) through the 5725A binding posts. Support Equipment and Services Fluke supports your calibration needs with precision, high-quality equipment and a wide range of services. Depending on your needs, location, a
22、nd capabilities, you may decide to support your 5700A/5720A Series II Calibrator independently or use Fluke services for part, or all, of your support needs. The following paragraphs describe the support equipment and services offered by Fluke for the calibrator. For specifications and ordering inst
23、ructions for this support equipment and other Fluke instruments, refer to the Fluke catalog, or contact a representative at a Fluke Sales and Service Center. 732B Direct Voltage Reference Standard The Fluke 732B is a rugged, easily transported solid state direct voltage reference standard with a hig
24、hly predictable 10 V output. This predictability allows the Fluke Standards Laboratory, as well as many Fluke customers, to completely eliminate fragile, saturated standard cells. Laboratories still maintain standard cells using the 732A and 732B as a transportable voltage standard, eliminating the