1、 hph 8/23/06 On the Calibration of Digibridges and the Verification of their High-Frequency Specifications (footnotes in superscript, references in parenthesis) Part 1, Introduction Traditionally, resistance calibrations have been made at dc, but the GR/IETLabs Digibridges are calibrated at 1 kHz. T
2、hus it is reasonable to question if this ac calibration is valid. Moreover, some the Digibridges have specified accuracies to 100 kHz and 200 kHz. It is also reasonable to ask how these accuracies are determined and verified. This memo hopes to answer these questions. The brief answer is that the Di
3、gibridge specifications have been verified by NIST calibrations. However, there is more that that to say than that. NIST and other national labs cannot possibly calibrate resistors, capacitors and inductors of all values at all frequencies. Values are scaled by known ratio techniques over very wide
4、ranges. Changes in capacitor and resistor values2 between official calibrations at different frequencies are determined by extrapolation between these calibrations using the known behavior of the component. This is especially convincing if there is no change between these values. This must be consid
5、ered valid for if not, NIST would have to make an infinite number of calibrations on each device. Extrapolation of values to higher frequencies can also be valid, in fact it is encouraged by NIST. In a NBS Technical Note, Jones (1) shows how the values of air3 capacitors can be extrapolated to highe
6、r frequencies by determining their inductance and applying a simple formula. He states that “it is possible for a laboratory to perform its own high frequency calibrations” and that “mutual benefits can accrue” from such independent calibrations. He goes on to suggest that if, the inductance is accu
7、rately determined, the value of the capacitor can be determined up to a frequency where the change is 10% by using the formula C = Co/(1-w2LCo). The tolerance of the value so determined is based on a plus or minus deviation of 10% in the measured value of the inductance. He determined the inductance
8、 of GR 1401 and 1402 capacitors by measuring their resonant frequency with a grid-dip meter. It is argued (3) that a similar extrapolation method can be used for most types of precision resistors, especially small film resistors, particularly if a wide tolerance on the calculated error is used. More
9、over, it is recommended that new high-frequency values should not be assigned to the standards (as Jones suggested), but rather the calculated changes be used only to determine the increased tolerances of the high-frequency values. Even though the equivalent circuit of such resistors is somewhat mor
10、e complicated than that of air capacitors, the frequency errors can be reduced to simple formulas in most cases, see Appendices A, B and C. The high-frequency specifications of the Digibridges were determined mainly by this method (see part 3). 1 RadioFans.CN 收音机爱 好者资料库 Part 2, The AC Calibration of
11、 Digibridges The more accurate Digibridges have four ranges and are calibrated by four external precision resistors (Vishay type S102). The four internal standard resistors, used to measure the current through the DUT, are of the same values and type. The standards are calibrated by comparison to se
12、condary standards whose values are derived from a dc NIST calibration of a precision 10 k resistor. However, the Digibridge is made at 1 kHz. The ac-dc difference is assumed to be zero because published specifications for the maximum inductance and capacitance of these resistors would cause errors o
13、f less that 1 ppm when used in the equations derived from the equivalent circuit. At ac, any resistor that is not pure (Q = 0) has two equivalent values, series (Rs) and parallel (Rp). The relations ship between then is Rp = Rs(1 + Q2). The series value of low-valued resistors changes less with freq
14、uency while the parallel value of high-valued resistors changes less. If we calculate the 1 kHz error of Rs for low-valued resistors and of Rp for high-values ones, the changes are all less than one part per billion. It is hard to doubt that the dc value may be used in this ac calibration. Note that
15、 the calibration kit includes both an open and a short and both open and short calibrations are made. This removes, almost completely, any capacitance test fixture and resistor mounting and any mutual inductance between the leads as long as they arent moved. Part 3. High-Frequency Accuracy Specifica
16、tions The high-frequency accuracy specifications were determined before we had high- frequency NIST calibrations on resistors of the same values as those used in the Calibration Kit. I believe GR did have NIST calibrations on many values of GR 1442 Coaxial Resistors but have no idea where there is a
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- OntheCalibrationofDigibridgesandtheVerificationHPH 电路图
收音机爱好者资料库所有资源均是用户自行上传分享,仅供网友学习交流,未经上传用户书面授权,请勿作他用。
链接地址:http://www.radiofans.cn/p-201331.html



Marshall-9100-62-02-Schematic电路原理图.pdf