1、 Application Note Temperature measurement and calibration: What every instrument technician should know F ro m t h e F l u k e D i g i t a l L i b r a r y w w w. f l u k e . c o m / l i b r a r y Introduction Temperature may be the most commonly measured physical parameter. Yet there have never been
2、 so many ways to measure it as there are today. With so many options its natural to have a few questions. How do I measure temperature? How accurate is my measurement? What temperature range is required? What type of device best measures tempera- ture? Does my instrument require certification? These
3、 are very common ques- tions when confronted with the need to measure temperature. A variety of measurement devices may be used for temperature; liquid-in-glass thermometers (LIG), thermocouples (TCs), thermistors, resistance tempera- ture detectors (RTDs), platinum resistance thermometers (PRTs) an
4、d standard platinum resistance thermometers (SPRTs). This appli- cation note focuses on electronic temperature measurements, and helps clarify the answers to some of these nagging questions. More information on these subjects is available at . How do I measure temperature? After inserting a temperat
5、ure sensor into the area to be measured, it takes time before the temperature reading is stabilized. For the thermom- eter to stabilize at the correct temperature, the probe must be sufficiently immersed. Some thermometers require more immersion depth than others. Most precision thermometers require
6、 four to six inches when inserted into a liquid or snug fitting well, depending on the diameter of the probe. Best results in terms of accuracy and stabilization time occur when the probe can be inserted into a stirred liquid. Air pockets between probes and solid sur- faces lead to longer stabilizat
7、ion times and require more immer- sion than would be required in a liquid. Specialized thermom- eters are needed for measuring temperatures on surfaces and for situations where the probe cable will be exposed to extreme temperatures. Often devices that measure and display temperature need to be veri
8、fied or calibrated against a reference thermom- eter. Accuracy is improved when the distance between the two thermometers is reduced. A best practice is to align the centers of the sensing elements of the reference thermometer and the device under test. Be aware that the location of the center of th
9、e sensor depends on the sensor type and model (i.e. PRT, ther- mocouple, bimetallic). A common method of calibrat- ing temperature sensors is to remove them from where they are installed and place them in a dry-well calibrator or a Micro- Bath. These calibrators provide a stable temperature environm
10、ent over a range of temperatures to compare the thermometer under test to the calibrator display or to a reference thermometer for more accuracy. Alternatively, temperature sensors may be calibrated or verified without RadioFans.CN 收音机爱 好者资料库 2 Fluke Calibration Temperature measurement and calibrati
11、on: What every instrument technician should know removing them from their installed location. Usually this is done by inserting a reference thermometer into a thermowell, immersion well, or thermometer pocket installed next to the ther- mometer to be tested. In other cases the sensing element of the
12、 reference thermometer must be placed inside of the freezer, oven, or environmental cham- ber being verified, calibrated or adjusted. In these cases it is often necessary to record data over a period of time such as a few hours to verify performance. Statistics such as average value, maximum and min
13、imum, or standard deviation are some- times recorded. Testing the energy perfor- mance of steam systems, cooling towers, heat exchangers, and refrigeration systems, turbines, and internal and external combustion engines requires measuring differences between inlet and outlet temperatures. Sometimes
14、these measurements have to be made from outside the pipe using thermocouples, thin-film sensors, or infrared temperature measurements. However, the best accuracy will be achieved when a thermowell has been properly installed in both the inlet and outlet pipes so that a probe can be inserted and suff
15、iciently immersed. Because pipe diameters are sometimes a limiting factor for immersion, the best location for a thermowell is at an elbow in the piping so that the probe can be inserted parallel to fluid flow with as much immer- sion depth as needed. How much accuracy is needed? Decisions about acc
16、uracy should be made carefully. Inaccuracy leads to mistakes and mistakes cost money. Mistakes may lead to down time, excessive energy costs, high product defect rates, safety hazards, and public health threats. Thermometers are specified by design engineers for tempera- ture monitoring or control.
17、These specifications should include the accuracy of the thermom- eters. A design engineer, quality engineer or metrologist should also specify the calibration requirements. However, it is not uncommon for instrument tech- nicians to receive a calibration job and little or no information about calibr
18、ation requirements. A common calibration strategy is to reduce mistakes by keeping the uncertainty of the calibration standards to a low percentage of the accuracy of the thermom- eter under test. This percentage is usually described as a Test Uncertainty Ratio (TUR). For example, the 4:1 TUR used b
19、y the military and other industries keeps the collective uncertainty of the calibration standards to 25 % of the thermometer under test accuracy. For comparison, a TUR of 2:1 means that the uncertainty is 50 % of the thermometer accuracy, and if the reference thermometer has the same accuracy as the
20、 thermom- eter under test, then the TUR is 1:1. The latter TUR is never recommended for calibration and would produce unreliable results. With a more accurate calibra- tion standard you can identify more actual out-of-tolerance field devices. Table 1 illus- trates the expected frequency of making mi
21、stakes at various TURs. Table 1 is based on a scenario where 950 of 1,000 instruments are truly in toler- ance. For example, if all 1,000 are calibrated with a 2:1 TUR then we expect that 926 will be found in tolerance (accepted), 12 of which are truly out of toler- ance (false accept). Of the 74 ex
22、pected to be rejected, 41 are expected to be truly in tolerance (false reject). The cost incurred for each of those falsely rejected instruments could range from $50/each for a calibration house to $10,000/each in down time in the chemical process industry. Temperature rangeAccuracyCost Noble-metal
23、thermocouples (Special tolerances) R, S: 50 C to 1760 C 0.6 CMed Base-metal thermocouples (Special tolerances) B: 0 C to 1820 C E: 270C to 1000 C J: 210 C to 1200 C K: 270 C to 1370 C N: 270 C to 1300 C T: 270 C to 400 C 0.25 % 1 C 1.1 C 1.1 C 1.1 C 0.5 C Low Low Low Low Low Low PRTs and SPRTsIndust
24、rial: 80 C to 480 C Reference: 200 C to 660 C High temp: 0C to 1000 C 0.05 - 0.1 C 0.001 0.02 C 0.01 0.02 C Low - Med Med - High Med High Precision thermistors0 C to 100 C0.002 CMed Table 2. Temperature sensor tradeoffs among temperature range, accuracy and cost. The most accurate sensors are the mo