1、 1659 RLC Digibridge Instruction Manual FORM 1659-0120-02/Bl QuadTech, Inc., 1992 5 Clock Tower Place, 210 East Maynard, Massachusetts, U.S.A. 01754 March, 2000 Telephone 978-461-2100 Sales 800-253-1230 Facsimile 978-461-4295 Website The material in this manual is for informational purposes only and
2、 is subject to change, without notice. QuadTech assumes no responsibility for any enor or for consequential damages that may result from the misinterpretation of any procedures in this publication. Contents Specifications Warranty Introduction -Section 1 Installation -Section 2 Operation -Section 3
3、Theory -Section 4 Service -Section 5 Parts Lists and Diagrams -Section 6 RadioFans.CN 收音机爱 好者资料库 These supplementary pages contain information of improvements or modifications not documented in the current manual. All references to GenRad in the manual now apply to QuadTech, Inc. Page iv -Table of C
4、ontents (Parts Lists and Diagrams -Section 6) Power supply assembly, board layout and schematic (pages 6-13 0 to 50 degrees C, 1 us, and (if START is not debounced) a-c the settling time or programmed delay. For single measurements, the DUT can be disconnected after e. The selected BIN line goes low
5、 at f; the other BIN lines stay high. AVERAGE measurement routines, ACQ OVER goes low (e) at the end of the last measurement. 2.7.2 Timing Figure 2-3. Refer to the accompanying figure for timing guidelines. Notice that START must have a duration of 1 us (minimum) in each state (high and low). If STA
6、RT is provided by a mechanical switch without debounce circuitry, the Digibridge will make many false starts; if START does not settle down (low) within the default settling time after the first transition to high, the measurement time may increase substantially. Measurement starts at time d, which
7、is essentially the same as time b or c; measurement is completed at g. (The START signals are expanded for clarity.) Interval a-e, during which the DUT must remain connected for data acquisition, is considerably shorter than the total measurement time a-g. The DUT can be changed after e (indexing on
8、 ACQ, to save time) or after g (indexing on EOT, for a simpler test setup), as explained below. After the calculation interval e-f, measurement results are available for sorting, i.e., one of the BIN lines goes low. A few micro-seconds later, EOT goes low (can be used to set a latch holding the bin
9、assignment). ACQ OVER, the selected BIN line, and EOT then stays low until the next start command. The time required for measurement depends on test conditions and operating selections. Interval a-e can be less than 15 ms; the cycle a-g can be less than 40 ms; refer to paragraph 3.5 for details. Set
10、 up the handler either of two ways: indexing on EOT or indexing on ACQ, as follows. The handler must supply a signal (here called start next measurement) when it has completed connection of the DUT to the test fixture. 2-6 INSTALLATION Indexing on EDT. Set up the handler to respond to the EOT signal
11、 from the Digibridge, which occurs at the end of test, when the bin assignment is available for sorting. Set up the Digibridge to receive its START signal from the handlers start next measurement signal. This setup is simpler than the one below. NOTE The Digibridge requires that a non-zero value be
12、entered for nominal value to enable generation of the EOT signal and indication by the GO/NO-GO lights; see paragraphs 3.8.3,3.8.4. Indexing on ACQ. Set up the handler to respond to the ACQ OVER signal from the Digibridge, which occurs when the data acquisition is complete. The handler can then remo
13、ve the DUT from the test fixture and replace it with another DUT, while the Digibridge is calculating the result. In addition, set up an interface that provides a START signal to the Digibridge by logical combination of the EOT signal from the Digibridge AND the start next measurement signal from th
14、e handler. Indexing on ACQ results in higher measurement rate than indexing on EOT. INSTALLATION 2-7 2.8 IEEE-488 INTERFACE (OPTION) 2.8.1 Purpose FigFigure 2-4. If you have the 1658-9620 interface option, you can connect this instrument to a printer or into a system (containing a number of devices
15、such as instruments, apparatus, peripheral devices, and generally a controller or computer) in which each component meets IEEE Standard 488-1978, Standard Digital Interface for Programmable Instrumentation. A complete understanding of this Standard (about 80 pages) is necessary to understand in deta
16、il the purposes of the signals at the IEEE- 488 INTERFACE connector on the rear panel of this instrument. NOTE For copies of the Standard, order IEEE Std 488- 1978, IEEE Standard Digital Interface for Programmable Instrumentation, from IEEE Service Center, Department PB-8, 445 Hoes Lane, Piscataway,
17、 N, J.08854, To make connection to a single device like a printer, use a IEEE-488 cable, which fits the rear-panel connector labeled IEEE-488 INTERFACE. For larger systems, each device is connected to a system bus, in parallel, usually by the use of several stackable cables. Refer to the figure for
18、a diagram of a hypothetical system. A full set of connections is 24 (16 signals plus shield and ground returns), as tabulated below and also in the Standard. Suitable cables, stackable at each end, are available from Component Manufacturing Service, Inc., West Bridgewater, MA 02379, U.S.A. (Their pa
19、rt number 2024/1 is for a 1-meter-long cable.) This instrument will function as either a TALK/LISTEN or a TALK ONLY device in the system, depending on the position of the TALK switch. TALK/LISTEN denotes full programmability and is suited for use in a system that has a controller or computer to mana
20、ge the data flow. The handshake routine assures that the active talker proceeds slowly enough for the slowest listener that is active, but is not limited by any inactive (unaddressed) listener. TALK ONLY is suited to a simpler system -e.g. Digibridge and printer -with no controller and no other talk
21、er. Either mode provides measurement results to the active listeners in the system. 2-8 INSTALLATION Figure 2-5. The handshake process, illustrated by timing diagrams of the pertinent signals, for a system with one talker and several listeners. For details, refer to the standard. 2.8.2 Interface Fun
22、ctions Figure 2-5. The following functions are implemented. Refer to the Standard for an explanation of the function subsets, represented by the identifications below. For example, T5 represents the most complete set of talker capabilities, whereas PPO means the absence of a capability. SH1, source
23、handshake (talker) AR1, acceptor handshake (listener) T5, talker (full capability, serial poll) L4, listener (but not listen-only) SR1, request by device for service from controller RL2, remote control (no local lockout, no return-to-Iocal switch) PP0, no parallel poll DC, device clear DT1, device t