1、Manual Part Number:20790 Revision Level: C Configuration Code:27 Print Date:November 1998 Operation Model 8502A has dual detector inputs for making ratio measurements. All technical and operation descriptions in this manual apply equally to both models unless otherwise stated. Performance specificat
2、ions for the Series 8500A are in Sectio n 1.2. Detector specifications are contained in Appendix D. Options, if any will be detailed in Appendix E. Power requirements are 100/120/220/240 Vac 10%, 48-400 Hz. See Section 2.2 for Safety Precautions and details to set the voltage and install the correct
3、 fuse for the area in which the instrument will be used. Series 8500A Peak Power Meters 1-2Manual 20790, Rev. C, November 1998 1.1.1Environmental Requirements Series 8500A power meters are type tested to MIL-T-28800E, Type III, Class 5, Style E for Navy shipboard, submarine, and shore applications e
4、xcept as follows: Operating temperature range is 0 C to 50 C (calibrator operating temperature range is 5 C to 35 C). Non-operating (storage) temperature range is -4 0C to +7 0C. Relative humidity is limited to 95% non-condensing. Altitude and EMI requirements are not specified. 1.1.2Items Furnished
5、 In addition to options and/or accessories specifically ordered, items furnished with the instrument are: 1 ea. - Power Cord 1 ea. - Detachable Detector Cable (for Model 8501A), or 2 ea. - Detachable Detector Cables (for Model 8502A) 1 ea. - Operation and Maintenance Manual 1.1.3Items Required The 8
6、500A requires an external power detector; see Appendix D for power detector specifications. 1.1.4Tools and Test Equipment No special tools are required to operate the Series 8500A. Test equipment required for calibration or performance verification is described in Chapter 4. 1.1.5Cooling No cooling
7、is required if the instrument is operated within its specified operating temperature range (0 to 50 C). 1.1.6Cleaning The front panel can be cleaned using a cloth dampened with a mild detergent; wipe off the detergent residue with a damp cloth and dry with a dry cloth. Solvents and abrasive cleaners
8、 should not be used. 1.1.7Receiving Inspection Use care in removing the instrument from the carton and check immediately for physical damage, such as bent or broken connectors on the front and rear panels, dents or scratches on the panels, broken extractor handles, etc. Check the shipping carton for
9、 evidence of physical damage and immediately report any damage to the shipping carrier. Each Giga-tronics instrument must pass rigorous inspections and tests prior to shipment. Upon receipt, its performance should be verified to ensure that operation has not been impaired during shipment. The perfor
10、mance verification procedure is described in Chapter 5 of this manual. Introduction Manual 20790, Rev. C, November 1998 1-3 1.1.8Returning an Instrument If you are returning an instrument to Giga-tronics for any reason, including service, first contact Giga-tronics Customer Service at (800) 444-2878
11、 or Fax at (925) 328-4702 so that a return authorization number can be assigned. You can also contact Customer Service over their e-mail address . To protect the instrument during reshipment, use the best packaging materials available. If possible use the original shipping container. If this is not
12、possible, a strong carton or a wooden box should be used. Wrap the instrument in heavy paper or plastic before placing it in the shipping container. Completely fill the areas on all sides of the instrument with packaging material. Take extra precautions to protect the front and rear panels. Seal the
13、 package with strong tape or metal bands. Mark the outside of the package “FRAGILE DELICATE INSTRUMENT”. If corresponding with the factory or local Giga-tronics sales office regarding reshipment, please reference the full model number and serial number. If the instrument is being reshipped for repai
14、r, enclose all available pertinent data regarding the problem that has been found. Series 8500A Peak Power Meters 1-4Manual 20790, Rev. C, November 1998 1.2System Specifications Frequency Range30 MHz to 40 GHz depending on detectors (See the Detector Specifications in Appendix D) Power Range(Direct
15、input to detector): PulseMode:-20 to +20 dBm CW Mode:-40 to +20 dBm Absolute Maximum Limit:+23 dBm (200 mW) AccuracyThe uncertainty of microwave power measurements depends on several factors, the most important is the effective mismatch of both the power detector and the RF source. Excluding mismatc
16、h effects, the measurement uncertainties of the instrument and its power detectors are as follows: Calibrator Power Uncertainty (at 0 dBm):1.5% (0.065 dBm) Linearity after Automatic Calibration:3% (at stable temperature) Temperature coefficient of linearity at ambien t5 C, CW and Peak, typical:-10 d
17、Bm negligible, 0 to 50 C -10 dBm 0.5%/deg.C, 15 to 50 C 1.0%/deg.C, 0 to 15 C (The instrument indicates if the 5 C calibration range is exceeded) Uncertainty (due to zeroing and noise):CW: 10 nW, 15 to 50 C (Avg. = 500) 20 nW, 0 to 15 C Peak: 3.5 mW, 15 to 50 C (Avg. = 100) 5.0 mW, 0 to 15 C Single
18、Pulse:15 mW, 15 to 50 C (typical) 2 dB and 20 dB below the peak level of the signal. Duty Cycle:50% Trigger Rate:10 Hz to 10 MHz External:TTL levels, Max PRF = 1 MHz Graph Display ModePlots the outline of the detected pulse on the EL display. Also provides readout of amplitude and timing information
19、. Fast Measurement ModeAvailable under GPIB control to provide fast data acquisition and output. For an averaging number = 1, typically will be between 70 and 120 measurements/second. Also can be used to provide fast data acquisition and throughput via a rear panel analog output for using an 8500A w
20、ith a network analyzer to make swept frequency response tests. Calibrator Specifications Frequency:1 GHz 5% Power Uncertainty at 1mW:1.5% Return Loss at 1 mW:Greater than 25 dB Self Calibration Time:Less than 1 minute Connector:Type N Series 8500A Peak Power Meters 1-6Manual 20790, Rev. C, November
21、1998 Auxiliary Inputs/Outputs Monitor:Provides a voltage proportional to the detected RF envelope. Rise time is typically 20 ns, output impedance is nominally 50 ohms. Analog Output Coefficient:100 mV/dB 0.5 mV (0 V = 0 dBm) Offset:10 mV Trigger Input:TTL RF Blanking:TTL open collector low during ze
22、roing. Used to control power source. VPROPF:(V/GHz) Allows direct entry of frequency from RF power sources equipped with a VpropF output. GPIB Interface:Per IEEE STD 488-1978 GPIB Indicators:REM, TLK, LSN, SRQ, LLO Remote OperationComplete setup and measurement capabilities accessible via GPIB (IEEE
23、 488). Reporting of errors, malfunctions, operational status, and self-test diagnostics available through serial poll capability. Direct Plot Output:Outputs hardcopy pulse profile plots, with time, date and part identification, to a GPIB plotter. GPIB Address:Selectable from front panel IEEE Interfa
24、ce Functions:SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, TE0, LE0 General Specifications Stored Setups:Saves settings at power down, and has 10 additional setups in non-volatile memory. Power-On Self-Test (POST):POST is optionally performed at any time. A diagnostic code indicates the cause and locat