GoldStar os7020 电路图.pdf
《GoldStar os7020 电路图.pdf》由会员分享,可在线阅读,更多相关《GoldStar os7020 电路图.pdf(100页珍藏版)》请在收音机爱好者资料库上搜索。
1、GoldStar MODEL OS- 7020 OSCILLOSCOPE SERVICE MANUAL GoklStar Praclslon Co . Ltd. RadioFans.CN 收音机爱 好者资料库 GoldStar MODEL OS- 7020 OSCILLOSCOPE SERVICE MANUAL GoldStar Pn!dslon Co . Ltd. RadioFans.CN 收音机爱 好者资料库 CONTENTS 1. General - Page 1 2. Specifications - 2 3. Accessories- 5 4. External View - - 7
2、 5 P rev en t i v e Ma i n ten an c e - - - - - - - - - - - - - - - - - - - - - 9 5-1 Cleaning - 9 5-2 Visual inspection - 10 5-3 Periodical calibration - 10 6. Measuring instruction- 11 6-1 Probe phase adjustment - 11 6-2 Voltage measurement - 12 6-3 Frequency measurement - 14 6-4 Rise (fall) time
3、measurement - 15 JI 6-5 Pulse width measurement- 16 6-6 Phase measurement- 16 7. Circuit description- 18 8. Calibration - 21 8-1 Period of check and calibration - 21 8-2 Test equipment required- 21 8-3 Preliminary procedure - 21 8-4 Preliminary control settings 23 8-5 Initial starting procedure- 23
4、9. Semiconductor lead configurations- 34 10. Electrical parts list -35 ( 1) ATTENUATOR - - - - - - - - - - - - - 35 ( 2) CBl INPUT AMP - 36 RadioFans.CN 收音机爱 好者资料库 (3) CH2 INPUT AMP.- 39 (4) VERTICAL CONTROL- 42 (5) VERTICAL MAIN AMP.- 44 (6) TRIGGER AMP. - 46 (7) SWEEP GENERATOR- 47 (8) TIMING CIRC
5、UIT- 49 (9) HORIZONTAL MAIN AMP. - 50 (10) CHOP PULSE GENERATOR- 52 (11) HIGH VOLT :-, . ro I l I l WAN.g=- - . (J CJ u - IGoldStor1 osouoscoPE os.1010 - l - - :.,-.;- 0 0 Ol + =i - Q, = I 1 - r 1 I - -.:=:=:=j - ./-. .,._ _,- UL. 1 I =:=:=:=: _J 5. PREVENTIVE MAINTENACE These are the precautions on
6、 preventive maintenance which will protect the instrument from trouble and keep it clean for a long time. 5-1 Cleaning Since the extent of instrument smearing depends on the ambient condition in which the oscilloscope is used, the frequjncy of the cleaning operations cannot be defined initially. Cle
7、aning should be accomplished as required. Dust deposited inside the instrument affects the normal flow of cooling air and invites the local overheating of components parts of results in trouble. Smeared switch contacts of connector pins can be the cause of defective contact, and smearing of the circ
8、uit can cause arcing between circuits, particularly in moist season. Cleaning agents recommended and those prohibited are listed as below. Recornrneded agent Alcohol, gasoline of keroscene Prohibited agent I I I Acetone, tri-ethyl-ketone lacquer thinner or agent containing ketone series so L vent Cl
9、eaning the cver Normal sruedriGg of the cover can be cleaned by washing the cover with a neutral chemical cleaner. For greasy smearing, use a recommended agent with a soft cloth. Front and other panels. Clean the smeared panel with a soft cloth moistened with a recommended agent. Dust in-instrument
10、The best way to remove dust inside the instrument is the blowing of compressed air. Remove persistent dust with a soft painting brush and blow the air again. CRT The CRT screen will be smeared if it is used for too long without the filter. Ordinary dust of finger print left on the CRT can be cleaned
11、 by a soft cloth. Use a soft cloth moisted with alcohol to remove persistent smearing. Filter The filter may be clogged when used for a long time. Dust or fingerprint can be cleaned with a dry soft cloth. Use a soft cloth moisted with alcohol to remove persistent smearing. 5-2 Visual inspection The
12、instrument should be inspected occasionally for such defects as broken connections, improperly seated transistors, damaged circuit boards, and heat-damaged parts. The correc- tive procedure for most visible defects is apparent ; howe- ver, articular care must be taken if heat-damaged components are
13、found. Overheating usually indicates other trouble in the instrument; therefore, correcting the cause of the overheating is impor- tant to prevent the recurrance of the damage. 5-3 Periodical Calibration In order to use the oscilloscope accurately, periodical calibration of circuit is essential. Whe
14、n this oscilloscope is used frequently, a calibration will be required at every l,000hours of operation. 10 The frequency of calibration may be extended to 6 months, if the oscilloscope is used infrequently. 6. MEASURING INSTRUCTION 6-1 Probe phase Adjustment Proper adjustment of the probe phasing i
15、s required before measurements as the incorrect phasing results in eroneous measurements when the XLO position is selected. 1. Set the controls as follows VOLTS/DIV VARIABLE TIME/DIV (sweep)VARIABLE O.lV CAL o.sms CAL (Horizontal)POSITION CENTER 2. Set the probe switch to x10 and connect the probe b
16、etween the input connector and the CAL OUT terminal. 3. Display the calibration voltage waveform with 5 divisions amplitude 4. Adjust the variable capacitor of the probe using the small driver, and set the top of the waveform to the correct phasing shown in Fig. 6-1 I I I I I I - - - - _ ._ - - - -
17、I I I I : I . . . . . I . I I I I - - - - - - _ ,. _ - - - - - I , r y r r I I r r r! r r I . - - - . - - -t- - - - - I . . . I I I correct OVER -Shoot UNDER - Shoot Fig 6-1. Phasing of probe o Volt1ge is pplitd t I I I I I l I I I I I- o - volt rlfererce llnR I AC-CiND-DC switch to CiND) :i . 1 . r
18、 . +-+- T I I I I I l - I I I Fig 6-2. DC Voltage measurement 12 6-2-2 AC voltage measurement The measurement of the voltage waveform is performed as follows ; Set the VOLTS/DIV switch in order to obtain the amplitude for easy reading, read the ampl- itude as shown in Fig. 6-3 and calculate by equat
19、ion(!) or (2). When the waveform superimposed on DC current is measu- red, set the AC-GND-DC switch to DC in order to measure the value including DC component, or set this switch to AC in order to measure AC component only. The measured value by means of this procedure is peak value(Vp-p). Effective
20、 value(Vrms) of a sine wave signal can be Effective voltage(Vrms) = given by Equation(). Peak voltage(Vp-p) 2 ff - (3) Vp-p rz l . r r 1 T I I I I I Fig 6-3. AC Votae measurement 13 6-3 Frequency measurement The first method; Calculate the one cycle time(interval) of the input signal and obtain the
21、frequency by equation(4) Frequency(Hz) = l (C) - (4) PERIOD(S) The second method; Count the repetition number N per 10 divisions in the viewing area, and calculate the frequency by Equation (5) N(C) ( 5 Frequency(Hz) = - TIME/DIV setting value(s/div)xl0(div . When N is large(30-50), the second metho
22、d can give a high accuracy level than that obtained with the first method. This accuracy is approximately equal to the rated accuracy of sweep rate. However, when N is small, the count below decimal point becomes very ambiguous, which results in considerable error. For the measurement of comparative