1、 Digital Storage Oscilloscopes Operators Manual Gould 400 Series RadioFans.CN 收音机爱 好者资料库 Gould 400 Series Digital Storage Oscilloscopes Operators Manual Copyright1990 Gould lnc., Instruments Division (Gould Electronics Ltd.) RadioFans.CN 收音机爱 好者资料库 400 Operators Manual Contents Introduction Getting
2、Started 1 Advanced Features 2 The Menus 3 Performance Checking 4 Waveform Processing 5 Battery Unit Operation 6 Alphabetical Summary 7 Appendix 1 Error Messages A1 Appendix 2 Specification A2 Appendix 3 Remote Operation A3 Appendix 4 Front the 400 the 420 and the 450. All versions can be fitted with
3、 a battery unit which provides the instrument with a fully automatic built in Nickel Cadmium battery and charger which allows uninterrupted operation of the DSO in the event of an AC supply failure and complete operation independent of an AC supply. Obtaining a trace is especially simple - just conn
4、ect the signal and press the Auto Setup button - the 400 does the rest. Having obtained a trace, readily accessible datum lines and a cursor make it easy to take automatic timing and voltage measurements directly from the display. On the 420 and 450 models, the waveform processing function increases
5、 the power of the cursor measurements in terms of both capture and post storage analysis and measurement functions. The innovative use of ergonomically designed pressure sensitive push button controls provides a combination of precision and flexibility for ease of operation. More advanced features o
6、f the 400 series include a com- prehensive range of menu-controlled functions. For example, the Display and Trigger menu operates features such as the trigger delay and pre-trigger display functions. The pre-trigger display function allows the signal prior to the trigger point to be captured and dis
7、played. Three complete traces can be stored for future use and recalled to the display via the Save Trace and Recall Trace menus respectively. With the built in battery back up facility, these will be retained even when the instrument is switched off. In addition to the above features, the 420 and 4
8、50 instru- ments have a built in 4 colour plotter and a battery backed Real Time Clock. This provides a simple and convenient method of obtaining permanent hard copy plots of the screen display. The plots will contain the date and time of acquisition together with the date and time of plotting. The
9、450 has all the features of the 420 and has a signal bandwidth of 50MHz and an extra timebase range of 50ns/div. Numbers circled in the text refer to the controls shown on the front panel picture in Appendix 3. Getting Started 1.1-1.1.6 400 Series Operators Manual 1.1 Safety and Power Requirements t
10、hey should be removed only by suitably qualified personnel for maintenance and repair purposes. 1.1.1 International Safety Warning (as required for I.E.C. 348 Cat I) This instrument has been designed and tested in accordance with IEC publication 348, and has been supplied in a safe condition. This m
11、anual contains information and warnings which must be observed to keep the instrument in a safe condition. The instrument should not be switched on if it is damaged and it should not be used under wet conditions. For the correct and safe use of this instrument it is essential that both operating and
12、 service personnel follow generally accepted safety procedures in addition to the safety precautions specified in this manual. Whenever it is likely that safety-protection has been impaired, the instrument must be made inoperative and be secured against any unintended operation. Qualified maintenanc
13、e or repair personnel should be informed. Safety protection is likely to be impaired if, for example the instrument shows visible damage or fails to perform the intended measurements correctly. WARNING: Removing the covers may expose voltages in excess of 8000V at the side of the display tube; these
14、 may be present for up to one minute after the instrument has been disconnected from the power source. 1.1.4 Ventilation and Dust The instrument relies on forced air cooling via a fan and ventilation slots. Adequate ventilation can usually be achieved by leaving a 75mm (3 gap) around the instrument.
15、 The instrument should not be operated in dusty environments. If the CRT filter requires cleaning it can be easily removed by pressing in its right hand edge as shown by the moulded arrow. 1.1.5 Operating Temperatures The instrument is designed to be operated in an environment having an ambient temp
16、erature of between 0 and 50 degrees C, (0 to 45 degrees if the battery unit is fitted) and to operate with full accuracy between 15 and 35 degrees C. Note: Direct sunlight, radiators and other heat sources should be taken into account when assessing the ambient temperature. The instrument may occasi
17、onally be subjected to temperatures between 0 and -10C without degradation of its safety. 1.1.2 Grounding THE INSTRUMENT MUST BE GROUNDED. AC The instrument must be operated with a protective ground connected via the yellow/green conductor of the supply cable. This is connected to the instrument bef
18、ore the line and neutral connections when the supply plug is inserted into the socket on the back of the instrument. If the final connection to the supply is made elsewhere, ensure that the ground connection is made before line and neutral. DC If the unit is disconnected from the AC supply, and powe
19、red from an isolated DC source or the internal battery unit, the unit will not be grounded. Independent provision must be made to maintain the case at a safe potential, by grounding the safety ground terminal on the DSO rear panel. The safety ground terminal is connected to the instrument case, to t
20、he ground of the input signals (outer of the BNC connectors), and also internally to the negative side of the DC supply. 1.1.6 Power and Frequency Requirements The instrument uses less than 85V A (200V A if battery unit fitted) and operates from line voltages of 90V to 130V, and 190V to 265V, at 45
21、to 400Hz. Under the extreme conditions of 90V and 45Hz, the instrument will still operate correctly even if there is a half cycle dropout in the line supply. The instrument may be powered from a direct current supply in the range 12V to 33V. The unit cannot be damaged by applying power to both input
22、s simultaneously. Before connecting the instrument to the supply, ensure that the rear panel AC supply voltage selectors are set to the appropriate voltage. Access to the voltage selector can only be made if the AC supply connector is removed, and is by hinging open the panel connector from its top
23、edge. The selector barrel can then be moved to the required setting. This same operation provides access to the AC supply fuse, which must be changed to suit the supply voltage as shown in table 1.1. 7. The AC power connection is via a standard IEC, CEE 22 connector and the DC power input is via the
24、 supplied 0.25 spade connector (Gould part No. 457839.) See appendix 4 for polarity details. WARNING: Any interruption of the protective ground conductor inside or outside the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited. Signal connections to the ins