1、 PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT TEL:(516)334-5959(800)899-8438FAX:(516)334-5988 534MainStreet,Westbury,NY11590 IET LABS, INC. HARS-LX SeRieS Laboratory Standard Decade Resistance Substituter User and Service Manual Copyright 2009 IET Labs, Inc. Visit for manual revision updates HARS-
2、LX im/April 2009 RadioFans.CN 收音机爱 好者资料库 PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT TEL:(516)334-5959(800)899-8438FAX:(516)334-5988 534MainStreet,Westbury,NY11590 IET LABS, INC. RadioFans.CN 收音机爱 好者资料库 WARRAntY We warrant that this product is free from defects in material and workmanship and, wh
3、en properly used, will performinaccordancewithapplicableIETspecifi cations.Ifwithinoneyearafteroriginalshipment,itisfound not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when returned to IET. Changes in this product not approved by IET or application of
4、 voltages or currents greater than those allowed bythespecifi cationsshallvoidthiswarranty.IETshallnotbeliableforanyindirect,special,orconsequential damages, even if notice has been given to the possibility of such damages. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCL
5、UDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTIBILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. RadioFans.CN 收音机爱 好者资料库 WARninG OBSERVE ALL SAFETY RULES WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES. Dangerous voltages may be present inside this instrument. Do not open the case Refer ser
6、vicing to qulifi ed personnel HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT WHENEVER HAZARDOUS VOLTAGES ( 45 V) ARE USED, TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS. USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUME
7、NT. Use extreme caution when working with bare conductors or bus bars. WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND KEEP UNREQUIRED PERSONNEL SAFELY AWAY. CAUtion DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THIS INSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON THE FRO
8、NT PANEL OR THE OPERATING GUIDE LABEL. Contents Chapter 1 Introduction .1 Chapter 2 Specifications .2 Model Information .3 Chapter 3 Installation .5 3.1 Initial Inspection . 5 3.2 Installation .5 3.3 Repackaging for Shipment . 5 3.4 Storage . 5 Chapter 4 Operation .6 4.1 Initial Inspection and Setup
9、 .6 4.2 Connection .6 4.2.1 General Considerations .6 4.3 Electrical Considerations .6 4.3.1 Thermal emf Considerations .6 4.4 Dial Setting .7 4.5 Power Considerations . 7 4.6 Environmental Conditions . 7 4.7 Switch Conditioning .7 4.8 Meter Shunt Applications . 8 4.9 Kelvin Bridge Applications . 8
10、Chapter 5 Maintenance .9 5.1 Maintainability and Reliability .9 5.2 Preventive Maintenance .9 5.3 Calibration .9 5.3.1 Calibration Interval .9 5.3.2 General Considerations .10 5.3.3 RequiredEquipment . 10 5.3.4 Calibration Procedure . 10 5.4 Adjustments. 11 5.4.1 Trimming Considerations .11 5.4.2 Tr
11、imming Procedure . 11 5.5 Replaceable Parts List .13 Figures and Tables Figure 1-1: HARS-LX Standard Decade Resistance Substituter .1 Figure 2-1: Typical Operating Guide Affixed to Unit .4 Figure 4-1: Kelvin Bridge Connections .8 Table 5-1: Trimming Potentiometers .12 Figure 5-1: Typical Trimmer Boa
12、rd .12 Table 5-2: Replaceable Parts List .13 Figure 5-2: HARS-LX Replaceable Parts .13 1 HARS-LX Series 1introduction Chapter 1 intRoDUCtion The Laboratory Standard (HARS-LX) Series of resistance decade substituters (Figure 1.1) is a fam- ily of instruments providing a very broad choice of high-perf
13、ormance resistance sources. Any number of decades from one to ten is available. The HARS-LX-8-0.01-RH is a direct replacement for the model RS925D. The HARS-LX substituter is a precision resistance source with excellent characteristics of accuracy, stability,temperaturecoefficient,andpowercoeffi- ci
14、ent. All these features serve to make it a laboratory resistance standard, exceeded in performance only by stand-alone standard resistors. Hermetically sealed wirewound resistors are used for 1 steps and over. These resistors are carefully conditioned under power and temperature in order to develop
15、the best stability characteristics. Actual experience has shown them to exhibit a storage sta- bility of better than 5 ppm/year, improving as they age. The low-resistance resistors are constructed with resistance wire with a minimum of copper resistance inseriestolimittemperaturecoefficienteffects.
16、Theunithasafixedminimumresistanceof10m. This is implemented by mechanically limiting the 10 m decade from going below the “1” position. In this manner, no zero resistance subtractions have to be made, and the accuracy given is for the absolute reading. The HARS-LX series employs completely enclosed
17、dust-tight very low contact resistance switches. They featuresolidsilveralloycontactsandquadruple- leaf silver alloy wipers which keep switch contact resistance to under 1 m per decade, and more im- portantly, keep switch contact resistance reproducible, insuring repeatable instrument performance. H
18、ighqualitygoldplatedtelluriumcopperbinding posts minimize the thermal emf effects which would artificiallyreflectachangeindcresistancemeasure- ments. All other conductors within the instrument, as well as the solder employed, contain no metals or junctions that contribute to thermal emf problems. Th
19、e HARS-LX is designed to allow very convenient maintenance of calibration over time. Most decades are calibratable without changing components or sol- deringresistors.Thedecadesforthe100through 100 k steps are calibrated with convenient trimmers. Trimming of the lower decades is also possible. With