《8035XA Ops Manual_REVF 电路图.pdf》由会员分享,可在线阅读,更多相关《8035XA Ops Manual_REVF 电路图.pdf(110页珍藏版)》请在收音机爱好者资料库上搜索。
1、Giga-tronics Incorporated v 4650 Norris Canyon Road v San Ramon, California 94583 925.328.4650 or 800.726.4442 v 925.328.4700 (Fax) v 800.444.2878 (Customer Service) v 925.328.4702 (CS Fax) .Certified ProductISO 9001. Certified Process Registrar: BSI, Certification No. FM 34226 v Registered 04 June
2、1996 v Amended 01 March 2000 Manual Part Number: Revision: Print Date: Series 8035XA Peak Power Sensors Operation 80351A = 100 x larger, 80352A = 1000 x larger Zero Set: 1.0 mW, Peak; 0.05 mW, CW Zero Drift: 1.0 mW, Peak 0.05 mW, CW in 1 hour at constant temperature, 24 hour warmup Noise Uncertainty
3、: 1.0 mW, Peak; 0.05 mW, CW at constant temperature, measured over a 1 minute interval, 24 hour warmup Sample Delay Timing Delay Range:-20 ns to 104 ms Delay Resolution:0.5 ns Delay Jitter:2.0 ns Trigger Level Set Range: Internal:-30 to +20 dBm Resolution:0.01 dB External:0.0V to 4.0V Resolution to
4、0.01V Trigger Jitter: 2.0 ns Settling Time:(50% to within 3%) 500 s) as shown in Figure 2-12. Figure 2-12: Pulse Profile and Sample Delay Test Setup Use the Sample Delay output as an oscilloscope trigger source. This waveform rises at the trigger point and falls at the sample point; thus, it provide
5、s both a stable scope trigger source and a precise indicator of the trigger point and sample point. The time length of the sample delay pulse is the sum of the sample delay which is displayed on the Series 8540 power meter front panel, and the sample delay offset which is available through the menu.
6、 Series 8035XA Peak Power Sensors 2-12Manual 21568, Rev. F, March 2008 2.3.7Sample Delay Offset In addition to compensating for delay line triggering variations or external triggering cables, sample delay offset can be used to set a 0.0 ns time reference point after the trigger point (see Figure 2-1
7、3). Figure 2-13: Sample Delay total = sample delay + sample delay offset total = 20,000s + 0.120s = 200 120s total = 200,000s + (-0.010s) = 199.990s t t t or total t The use of a digital oscilloscope can permit better viewing of data. There are two small markers injected onto this waveform. The firs
8、t is a small triggering marker which is added slightly after the trigger point. The second marker on the waveform is the sample marker. The sample marker is located slightly behind the actual sample point. There may be small markers at the end of the sample transfer and when the trigger signal occur
9、s. Because the visibility of these markers varies greatly with signal level and horizontal sweep rate, the use of the SAMPLE DELAY output is recommended. In Figure 2-14, triggering occurs at the frame start of a pulsed TDMA communications signal. The trigger level is set such that triggering can onl
10、y occur on the highest amplitude pulse; this provides stable triggering. A sample delay offset is used to set a 0.0 ns reference point at the start of the third data burst pulse. Figure 2-14: Using SD to Offset a 0 ns Time Reference Detector Out Connection Sample Delay Offset Sample Delay Sample Del
11、ay Internal Trigger Level Sample Delay Connection Operation Manual 21568, Rev. F, March 2008 2.3.8Measuring Pulse Droop Pulse characteristics such as droop, ripple, and overshoot can be measured quickly using referenced measurements. This involves the use of the front panel REL key for the Series 85
12、40 power meter users, or the CALC#:REF:COLL function for 58542 power meters. 1.Connect the 8035XA Peak Power Sensor to the power meter and the CALIBRATOR output. 2.Press CAL/ZERO to calibrate the sensor to the meter. 3.Upon successful completion of power sweep calibration, connect the sensor to a pu
13、lsed signal source. The power level must be above the trigger level. 4.Connect the 8035XA Detector Out and Sample Delay leads to a digital oscilloscope. 5.Set the sample delay (DLYA) to the t1 position just after the rising edge as shown in Figure 2-15. 6.Press REL. The display should now read appro
14、ximately 0.00 dBm or 100%. 7.Set the sample delay to the t2 position just before the falling edge of the pulse. The display is now reading the pulse-top amplitude variation. Figure 2-15: SD Setting for Measuring Pulse Droop Detector Out Trigger Level t1 t2 Series 8035XA Peak Power Sensors 2-14Manual
15、 21568, Rev. F, March 2008 2.3.9Measuring 3 dB Pulse Width Pulse width and other pulse timing parameters can be measured using the REL key and the sample delay offset control (see Figure 2-16). 1.Preset the delay offset to 0 nS. 2.Set the sample delay equal to about half the pulse width. 3.Press REL
16、 to set the 0.0 dB reference level (100% for Watts display). 4.Set the sample delay to a position on the rising edge of the pulse. Increment or decrement the sample delay value until the power level display reads approximately -3.00 dBr. 5.Remember or write down the sample delay value. 6.Press MENU,
17、 and use the up/down arrow keys to display SETUP MENU. Press ENTER. Select PEAK SNSR SETUP, then select the current triggering method (INT or EXT). When DLY Offset appears, set the offset value to the same time value from Step 4. This sets the 3 dB down time point to a 0.0 ns reference position. 7.P
18、ress ENTER to return to the measurement display. 8.Set the sample delay to a position on the falling edge of the pulse. Increment or decrement the sample delay value until the display again reads approximately -3.00 dBr. The sample delay currently displayed is the signals 3 dB pulse width. Figure 2-
19、16: Using SD to Measure a 3 dB Pulse Width 100% Reference Level 3 dB down (50%) 3 dB down (50%) Sample Delay Offset Total Delay Trigger Level Sample Delay = Pulse Widtht t= Operation Manual 21568, Rev. F, March 2008 2.3.10Measuring Rise-Time Rise time measurements can be performed using a technique
20、similar to the pulse width measurement. This example uses a linear Watts display readout rather than the more common logarithmic dBm readout for convenient identification of the 10% and 90% levels. 1.Preset the delay offset to 0 ns. 2.Press dBm/mW to obtain a linear, Watt, display readout. 3.Set the
21、 sample delay equal to about half the pulse width. 4.Press REL to set the 100% reference level. 5.Set the sample delay to a position on the rising edge of the pulse. Increment or decrement the sample delay value until the power level display reads approximately 10%. 6.Remember or write down the samp
22、le delay value. 7.Press MENU. Select SETUP MENU then PEAK SNSR SETUP. Proceed to the SAMPLE DELAY OFFSET selection. Set the offset value to the same value from Step 6. 8.Return to the measurement display and increment the sample delay until the display reads approximately 90%. The sample delay curre
23、ntly displayed is the 10% to 90% rise time. 2.4GPIB Commands These commands supplement the commands given in the Series 8540C Universal Power Meter Operation and Maintenance Manual. 2.4.1Setting Trigger Modes These commands set the trigger method for the 8035XA sensor. The sensor can be set to trigg
24、er on the rising RF envelope of the power signal. This is the internal trigger mode. An external TTL trigger can be used, or the sensor can “free run” and allow a CW measurement mode with no trigger required. The terms digital filter (as used in some instrument instructions) and averaging buffer (as