KEI 414A INST RevB(Model414A) 电路图.pdf
《KEI 414A INST RevB(Model414A) 电路图.pdf》由会员分享,可在线阅读,更多相关《KEI 414A INST RevB(Model414A) 电路图.pdf(28页珍藏版)》请在收音机爱好者资料库上搜索。
1、INSTRUCTION MANUAL MODEL 4 1 4 A PICOAMMETER 0 COPYRIGHT 1976, K E I T H L E Y INSTRUMENTS, INC. PRINTED MAR. 1978, CLEVELAND, OHIO, U.S.A. RadioFans.CN 收音机爱 好者资料库 CONTENTS MODEL 414A CONTENTS 1. 2. 3. 4 . 5. 6. 7. 1 4 6 7 9 14 17 ii 0575 RadioFans.CN 收音机爱 好者资料库 MODEL 414A ILLUSTRATIONS I LLU ST RAT
2、 I 0 N S Fig. No. T i t l e Page 3 3 5 6 10 12 13 21 21 0974 iii RadioFans.CN 收音机爱 好者资料库 S PEC IF ICAT ION S MODEL 414A S P E C I F ICAT 1 0 N S RANGE: 10-lo ampere full scale to l o 2 ampere in seventeen lx and 3x ranges, positive or negative currents. ACCURACY: *4% of full scale on 3 x 10-9 to 10-
3、10 ampere ranges. ZERO DRIFT: Less than 0.5% of full scale per week plus 0.05% per C on any range after a 10-minute warm-up and with source voltages greater than 1 volt. INPUT VOLTAGE DROP: Less than 1 millivolt for full-scale meter deflection on any range when properly zeroed. EFFECTIVE INPUT RESIS
4、TANCE: Less than 0.1 ohm on ampere range, increasing to less than lo7 ohms on 10-lo ampere range. RISE TIME, MAXIMUM: Seconds, from 10% to 90%. Range, ampere Across Input 10.10 0.75 10.9 0.15 10.8 0.13 10-7 0.09 10 - 6 0.05 10.5 and above With up to 500 pF less than 0.001 MAXIMUM INPUT OVERLOAD: Tra
5、nsient: 1000 volts for up to 3 seconds using a Keithley (or other 10 mA limited) High Voltage Supply. Continuous: 600 volts on 10-10 to 10-6 ampere ranges, de- creasing to 12 volts on the ampere range. CONNECTORS: Input: Teflon-insulated UHF type. Output: Amphenol 8OPC2F. RECORDER OUTPUT: Output: il
6、 volt or 1 milliampere for full-scale meter deflection. Output polarity is opposite input polarity. Noise: 414s: Less than 0.3% of full scale peak-to-peak. 414A: Less than 1 % of full scale peak-to-peak. POWER: 105-125 or 210-250 volts (switch selected), 50-60 Hz, 5 watts. DIMENSIONS, WEIGHT: 5%w hi
7、gh x 8% wide x 10 deep: net weight, 8 pounds. ACCESSORIES SUPPLIED: Mating input and output con- nectors. iv 0974 MODEL 414A GENERAL DESCRIPTION SECTION 1. GENERAL DESCRIPTION 1-1. GENERAL. a. The Keithley Model 414A is a completely solid- state picoammeter which measures currents over 17 ranges fro
8、m LO- to ampere full scale. Accuracy is *2% of full scale on the to of full scale on the 3 x lo- to lo- ampere ranges. am ere ranges and *4% b. The Picoammeter employs matched MOS FET transis- tors in the input followed by a differential amplifier stage, a transistor driver and a transistor output s
9、tage. Negative feedback is used for stability and accuracy. 1-2. FEATURES. b. A unique circuit provides complete overload pro- tection for Model 414A without compromising the out- standing features of the MOS FET input. The Picoam- meter .will withstand transient overloads up to 1000 volts without d
10、amage, and overload recovery is almost instantaneous. a. The time and temperature stability of the 414A Picoammeter is unmatched by any other inexpensive pico- ammeter. The Model 414A will operate for days without requiring rezeroing. Zero drift with time is less than 0.5% of full scale per week. Te
11、mperature drift is equally small - a change of 10C affects the reading less than 0.5% of full scale on the most sensitive range. c. Fast warm-up is an inherent characteristic of the Picoammeter. It can be used almost immediately on its most sensitive range. For maximum stability, how- ever, about 10
12、 minutes warm-up time should be allowed. d. One volt or 1 milliampere at full-scale deflec- tion on all ranges is provided to drive chart recorders. The 1 milliampere recorder output will drive the Keith- ley 370 Recorder directly. When potentiometric record- ers are used, the output voltage can be
13、conveniently divided by a shunt resistor and adjusted for full scale with the rear panel Calibration potentiometer. 0671 1 GENERAL DESCRIPTION TABLE 1-1 MODEL 414.4 Front Panel Controls Control Functional Description Paragraph RANGE Switch Selects full-scale current range instrument is to measure. 2
14、-2 METER Switch Turns instrument on; selects meter polarity; checks instrument zero. 2-2,2-4 ZERO ADJ. Control Zeroes meter on any range. 2-2 Pilot Light Glows to indicate instrument is on. - TABLE 1-2 Rear Panel Controls Control Functional Description Paragraph INPUT Receptacle Connects input to so
15、urce. Receptacle is a Teflon-insulated UHF connector. 2-1, 2-2 OUTPUT Receptacle Connects output to monitoring device. 2-4 1 MA - 1 V Switch Selects output of instrument: 1 milliampere or 1 volt. 2-4 1 MA CAL Control Adjusts output from 0.95 to 1.05 mA. 2-4 117-234 V Switch Sets instrument for 117 o
16、r 234 volt ac power line. 2-2, 2-5 I Fuse 3AG Slow-Blow. 117-volt - .125 A; 234-volt - .062 A. 2-2. 2-5 2 0671 GENERAL DESCRIPTION MODEL 414A METER MlOl METER ZERO METER I - I POWER LAMP DS201 ZERO A D J CONTROL R123 METER SWITCH S l O l RANGE s 1 0 2 - SWITCH FIGURE 1. Front Panel Controls. INPUT 5
17、102 t 1 SWITCH S 103 OUTPUT 5103 117-234V U SELECTOR s 2 0 1 F S E IMACAL F 2 0 1 CONTROL R133 - - FIGURE 2. Rear Panel Controls. 0671 3 0 PE RAT ION SECTION 2. OPERATION MODEL 414A PICOAMMETER 2-1. INPUT CONNECTIONS. Use the following precautions when using the Picoammeter on the more sensitive ran
18、ges. a. The INPUT Receptacle of the Model 414A is a Tef- lon-insulated UHF connector. The center terminal is the high impedance terminal, and the outer shield is case ground. b. Carefully shield the input connection and the cur- rent source being measured, since power line frequencies are well withi
19、n the pass band of the Picoammeter on a l l ranges. Unless shielding i s thorough, pickup may cause definite meter disturbances. c. Use high resistance, low-loss materials - such as polyethylene, polystyrene or Teflon - for insula- tion. The insulation resistance of t e s t leads and fix- tures shou
20、ld be several orders of magnitude higher than the source resistance. Excessive leakage w i l l reduce accuracy. Any coaxial cable used should be a low- noise type which employs a graphite coating between the dielectric and the surrounding shield braid. d. Any change i n the capacitance of the measur
21、ing circuit to ground w i l l cause disturbances in the read- ing, especially on the more sensitive ranges. Make the measuring setup as rigid as possible, and t i e down con- necting cables to prevent their movement. If a contin- uous vibration is present, it may appear a t the output as a sinusoida
22、l signal and other precautions may be necessary to isolate the instrument and the connecting cable from the vibration. NOTE Keep the shield cap on the INPUT Receptacle when the Picoarmneter is not in a circuit. 2-2. OPERATING PROCEDURES. a. Check the fuse and the 117-2341 V Switch for the proper lin
23、e voltage. b. Connect the power cord t o the power source. c. Set the RANGE Switch t o ampere, the METER S i t c h t o (+). Within seconds the meter needle should read zero. Zero the meter with the ZERO ADJ. Control. After a few moments increase the current sensitivity b ampere range. Continue zeroi
24、ng with the ZERO A D J Con- trol. The instrument is now ready t o use. advancing the RANGE Switch i n decade steps to the 10- 18 d. I f long term measurements are t o be made, allow the instrument t o warm up for a t least 10 minutes. e. Attach the current source t o the INPUT Receptacle and turn th