1、 Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY Specifications are subject to change without notice. SPEC-2002 Rev. I / November 2009 Page 1 of 15 Model 2002 Multimeter Specifications This document contains the complete specifications for the 2002. Every effort has
2、 been made to make these specifications complete by characterizing its performance under the variety of conditions often encountered in production, engineering, and research. The Model 2002 provides Transfer, 24-hour, 90-day, 1-year, and 2-year specifications, with full specifications for the 90-day
3、, 1-year, and 2-year intervals. This allows the operator to utilize 90-day, 1-year, or 2-year recommended calibration intervals, depending upon the level of accuracy desired. As a general rule, the 2002s 2-year performance exceeds a 6-digit DMMs 90-day, 180-day, or 1-year specifications. ABSOLUTE AC
4、CURACY All DC and AC specifications are given as relative accuracies. To obtain absolute accuracies, the absolute uncertainties of the calibration sources must be added to the relative accuracies. The absolute uncertainties for the calibration sources used during Keithleys factory calibration are gi
5、ven in a table included in the specifications. The uncertainties of the operators sources may be different. TYPICAL ACCURACIES Accuracy can be specified as typical or warranted. All specifications shown are warranted unless specifically noted. Almost 99% of the 2002s specifications are warranted spe
6、cifications. In some cases it is not possible to obtain sources to maintain traceability on the performance of every unit in production on some measurements (e.g., high-voltage, high-frequency signal sources with sufficient accuracy do not exist). These values are listed as typical. 2002 SPECIFIED C
7、ALIBRATION INTERVALS Measurement Function 24 Hour1 90 Day2 1 Year2 2 Year2 DC Volts DC Volts Peak Spikes AC Volts RMS 3 3 3 AC Volts Peak AC Volts Average 3 3 3 AC Volts Crest Factor Ohms DC Current DC In-Circuit Current AC Current Frequency Temperature (Thermocouple) Temperature (RTD) 1 For T CAL 1
8、C. 2 For T CAL 5C. 3 For 2C of last AC self-cal. RadioFans.CN 收音机爱 好者资料库 Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY Specifications are subject to change without notice. Page 2 of 15 SPEC-2002 Rev. I / November 2009 Model 2002 Multimeter Specifications DC VOLTS
9、DCV INPUT CHARACTERISTICS AND ACCURACY Enhanced Accuracy4 10PLC, DFILT 10 Relative Accuracy (ppm of reading + ppm of range) Range Full Scale Resolution Input Resistance Transfer5 24 Hours6 90 Days7 1 Year7 2 Years7 Temperature Coefficient (ppm of reading + ppm of range)/C Outside TCAL 5C 200mV 8 210
10、.0mV 1nV 100G 0.4 + 1.5 3.5 + 3 15 + 8 19 + 9 23 + 10 2 + 1.8 2V 8 2.10V 10nV 100G 0.2 + 0.15 1.2 + 0.3 6 + 0.8 10 + 0.9 14 + 1 0.2 + 0.18 20V 21.0V 100nV 100G 0.1 + 0.05 1.2 + 0.1 6 + 0.15 10 + 0.15 14 + 0.15 0.3 + 0.02 200V 210.0V 1V 10M 1% 0.5 + 0.08 5 + 0.4 14 + 2 22 + 2 30 + 2 1.5 + 0.3 1000V 9
11、 1100.0V 10 V 10 M 1% 1 + 0.05 5 + 0.08 14 + 0.4 22 + 0.4 30 + 0.4 1.5 + 0.06 DC Voltage Uncertainty: = (ppm of reading) x (measured value) + (ppm of range) x (range used) / 1,000,000. % Accuracy: = (ppm accuracy) / 10,000. 1ppm of Range: = 20 counts for ranges up to 200V and 10 counts on 1000V rang
12、e at 7-digits. NORMAL ACCURACY10 1PLC, DFILT OFF Relative Accuracy (ppm of reading + ppm of range) Range Full Scale Resolution Input Resistance 24 Hours6 90 Days7 1 Year7 2 Years7 Temperature Coefficient (ppm of reading + ppm of range)/C Outside TCAL 5C 200mV 8 210.0mV 10nV 100G 3.5 + 6 15 + 11 19 +
13、 12 23 + 13 2 + 1.8 2V 8 2.10V 100nV 100G 1.2 + 0.6 6 + 1.1 10 + 1.2 14 + 1.3 0.2 + 0.18 20V 21.0V 1V 100G 3.2 + 0.35 8 + 0.4 12 + 0.4 16 + 0.4 0.3 + 0.02 200V 210.0V 10V 10M 1% 5 + 1.2 14 + 2.8 22 + 2.8 30 + 2.8 1.5 + 0.3 1000V 9 1100.0V 100V 10M 1% 5 + 0.4 14 + 0.7 22 + 0.7 30 + 0.7 1.5 + 0.06 SPE
14、ED AND ACCURACY Accuracy 4,11 90 DAYS (ppm of reading+ppm of range+ppm of range RMS noise 12) Range 10PLC DFILT On, 10 Readings 10PLC DFILT Off 1PLC DFILT On, 10 Readings 1PLC DFILT Off 0.1PLC DFILT Off 0.01PLC13 DFILT Off 200mV 8 15 + 8 + 0 15 + 8 +0.5 15 + 8 + 0.7 15 + 8 + 1 25 + 10 + 13 100 + 200
15、 + 15 2V 8 6 + 0.8 + 0 6 + 0.8 + 0.05 6 + 0.8 + 0.07 6 + 0.8 + 0.1 7 + 1 + 1.3 130 + 200 + 3 20V 6 + 0.15 + 0 6 + 0.15 + 0.03 7 + 0.15 + 0.05 8 + 0.15 + 0.08 15 + 0.5 + 0.7 130 + 200 + 3 200V 14 + 2 + 0 14 + 2+ 0.1 14 + 2+ 0.15 14 + 2 + 0.25 15 + 2 + 1 130 +200 + 3 1000V 9 14 + 0.4 + 0 14 + 0.4 + 0.
16、05 14 + 0.4 + 0.05 14 + 0.4 + 0.1 15 + 0.5 + 0.5 90 + 200 + 2 PLC = Power Line Cycles. DFILT = Digital Filter. NOISE REJECTION (dB)14 AC and DC CMRR15 AC NMRR Speed (Number of Power Line Cycles) Line Sync On16 Internal Trigger Line Sync On16, 25 Readings DFILT On Line Sync On16 DFILT Off Internal Tr
17、igger DFILT Off PLC 1 140 120 90 80 60 PLC 1 90 60 60 45 0 4 Specifications are for 10 power line cycles, synchronous autozero, 10-reading repeat digital filter, autorange off, except as noted. 5 Specifications apply for 20-reading repeat digital filter, TREF 0.5C (TREF is the initial ambient temper
18、ature), and for measurements within 10% of the initial measurement value and within 10 minutes of the initial measurement time. 6 For T CAL 1C, following 4-hour warm-up. TCAL is ambient temperature at calibration (23C at the factory). Add 0.5 ppm of reading uncertainty if the unit is power cycled du
19、ring this interval. 7 For T CAL 5C, following 4-hour warm-up. 8 Care must be taken to minimize thermal offsets due to operator cables. 9 Add 20ppm (V IN/1000V) 2 additional uncertainty for inputs above 200V, except in transfer accuracy specifications 10 Specifications are for 1 power line cycle, nor
20、mal autozero, digital filter off, autorange off. 11 For T CAL 5C, normal autozero. 1-year or 2-year accuracy can be found by applying the same speed accuracy ppm changes to the 1-year or 2-year base accuracy. 12 Typical values. Peak-to-peak noise equals 6 times RMS noise. 13 In burst mode, display o
21、ff. Burst mode requires autozero refresh (by changing resolution or measurement function) once every 24 hours. 14 For line frequency 0.1%. 15 Applies for 1k imbalance in the LO lead. For 400Hz operation, subtract 10dB. For the 200V and 1000V ranges, subtract 20dB. 16 For noise synchronous to the lin
22、e frequency. RadioFans.CN 收音机爱 好者资料库 Keithley Instruments, Inc. 28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY Specifications are subject to change without notice. SPEC-2002 Rev. I / November 2009 Page 3 of 15 Model 2002 Multimeter Specifications Effective noise is reduced by a factor of 10
23、for every 20dB of noise rejection (140dB reduces effective noise by 10,000,000:1). CMRR is rejection of undesirable AC or DC signal between LO and earth. NMRR is rejection of undesirable power line related AC signal between HI and LO. KEITHLEY FACTORY CALIBRATION UNCERTAINTY Range ppm of reading 200
24、mV 3.2 2V 3.2 20V 2.6 200V 2.6 1000V 2.6 Factory calibration uncertainty represents traceability to NIST. This uncertainty is added to relative accuracy specifications to obtain absolute accuracies. The 200mV and 2V range uncertainties are equal to the uncertainty of the 2V calibration source. The 2